摘要
We aimed to investigate the effects of mammary gland density and average glandular dose (AGD) on contrast-to-noise ratio (CNR) of breast-equivalent phantoms with different mammary gland/fat tissue ratios. Full-field digital-mammography breast X-rays were performed on breast-equivalent phantoms with three different mammary gland/fat tissue ratios (Phantom A [30/70], Phantom B [50/50], and Phantom C [70/30]) and seven thicknesses ranging from 10 mm to 70 mm. The prediction formula for the CNR was calculated by multivariate analysis and the effects of the various parameters on CNR were evaluated using a multiple regression analysis model. Higher CNR values were obtained with lower mammary gland/fat tissue ratios and lower phantom thicknesses. Variation in CNR among the three breast models was low (coefficient of variation, 3.4-8.7%) at lower phantom thicknesses (10-30 mm) and high (coefficient of variation, 10.5-16.8%) at higher phantom thickness (50-70 mm). CNR showed a strong negative correlation (r = -0.8989) with AGD across all three mammary gland ratios. A predictive formula for CNR using AGD and mammary gland density was developed. CNR can be predicted with high precision using AGD and mammary gland density. The predicted CNR could be used to measure the diagnostic reliability of mammography in breast cancer.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Average glandular dose with full-field digital mammography - Clinical results
    Hermann, KP
    Kehbel, S
    Obenauer, S
    Fischer, U
    Grabbe, E
    [J]. DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2003, : 502 - 504
  • [2] Dose reduction in full-field digital mammography: an anthropomorphic breast phantom study
    Obenauer, S
    Hermann, KP
    Grabbe, E
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (907): : 478 - 482
  • [3] Dose reduction in full-field digital mammography - An anthropomorphic breast phantom study.
    Obenauer, S
    Hermann, KP
    Grabbe, E
    [J]. DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2003, : 145 - 148
  • [4] Average glandular dose with amorphous silicon full-field digital mammography - Clinical results
    Hermann, KP
    Obenauer, S
    Marten, K
    Kehbel, S
    Fischer, U
    Grabbe, E
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (06): : 696 - 699
  • [5] Comparison of Mean Glandular Dose between Full-Field Digital Mammography and Digital Breast Tomosynthesis
    Teoh, Kar Choon
    Manan, Hanani Abdul
    Norsuddin, Norhashimah Mohd
    Rizuana, Iqbal Hussain
    [J]. HEALTHCARE, 2021, 9 (12)
  • [6] Indices for the evaluation of glandular dose heterogeneity in full-field digital mammography
    Shinohara, Sae
    Araki, Fujio
    Maeda, Megumi
    Okamoto, Rumi
    Nakamura, Mai
    Higashida, Yoshiharu
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2020, 40 (04) : 1429 - 1443
  • [7] In vivo average glandular dose evaluation: one-to-one comparison between digital breast tomosynthesis and full-field digital mammography
    Cavagnetto, Francesca
    Taccini, Gianni
    Rosasco, Raffaella
    Bampi, Rossana
    Calabrese, Massimo
    Tagliafico, Alberto
    [J]. RADIATION PROTECTION DOSIMETRY, 2013, 157 (01) : 53 - 61
  • [8] Dose reduction in Full-Field Digital Electromagnetic Radiation Mammography using Noise reduction Method: Phantom Study
    Choi, Seokyoon
    Seo, Jeongmin
    [J]. JOURNAL OF MAGNETICS, 2019, 24 (04) : 744 - 751
  • [9] Average glandular dose in digital mammography and digital breast tomosynthesis: comparison of phantom and patient data
    Bouwman, R. W.
    van Engen, R. E.
    Young, K. C.
    den Heeten, G. J.
    Broeders, M. J. M.
    Schopphoven, S.
    Jeukens, C. R. L. P. N.
    Veldkamp, W. J. H.
    Dance, D. R.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (20): : 7893 - 7907
  • [10] The effect of the antiscatter grid on full-field digital mammography phantom images
    Chakraborty, DP
    [J]. JOURNAL OF DIGITAL IMAGING, 1999, 12 (01) : 12 - 22