Accuracy of the i-STAT™ bedside blood gas analyser

被引:54
|
作者
Sediame, S
Zerah-Lancner, F
d'Ortho, MP
Adnot, S
Harf, A [1 ]
机构
[1] Hop Henri Mondor, Serv Physiol Explorat Fonctionnelles, F-94010 Creteil, France
[2] Hop Henri Mondor, INSERM, U492, F-94010 Creteil, France
关键词
acidosis; high oxygen tension; permissive hypercapnia; point-of-care testing; thin-film electrode biosensors;
D O I
10.1034/j.1399-3003.1999.14a36.x
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
The performance of the i-STAT portable clinical analyser for measuring blood gases and pH was evaluated,vith reference to a conventional blood gas analyser (ABL520 Radiometer). Ninety-two samples from the routine blood gas analysis laboratory were chosen according to a wide distribution of partial pressure of carbon dioxide (Pa,CO2), partial pressure oxygen (Pa,O-2) and pH and then analysed. All measurements were performed in duplicate by trained technicians from the central hospital laboratory. Differences between duplicate measurements were computed for Pa,CO2: (1.2 versus 0.4%), Pa,O-2 (1.7 versus 1.1%) and pH (0.06 versus 0.02%), for the i-STAT and ABL520, respectively. pH and Pa,CO2 values measured with the i-STAT were very close to those obtained with the ABL520, the difference (mean+/-sD) being 0.006+/-0.018 and -0.13+/-0.17 kPa, respectively. Statistical analysis showed that the differences between analysers did not depend on values of pH or Pa,CO2 The performance of the analysers depended on the level of PO2. Below 15 kPa (n=48), the two systems gave nearly identical values, the mean difference was 0.01+/-0.37 kPa. Between 16 and 55 kPa (n=44), there was a systematic but small (-0.69+0.67 kPa) underestimation of Pa,O-2 measured with the i-STAT (p<10(-8)). In conclusion, this study shows that blood gas analysis using the i-STAT portable device is comparable,vith that performed by a conventional laboratory blood gas analyser.
引用
收藏
页码:214 / 217
页数:4
相关论文
共 50 条
  • [41] Use of the i-STAT portable clinical analyzer in mice
    Tinkey, Peggy
    Lembo, Thomas
    Craig, Suzanne
    West, Cheri
    Van Pelt, Carolyn
    LAB ANIMAL, 2006, 35 (02) : 45 - 50
  • [42] Evaluation of the i-STAT Portable Clinical Analyzer in dogs
    Matburger, C
    Henke, J
    Hirschberger, J
    Matburger, S
    Erhardt, W
    TIERARZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2000, 28 (02): : 132 - 137
  • [43] Evaluation of the IRMA TRUpoint and i-STAT creatinine assays
    Nichols, James H.
    Bartholomew, Cathy
    Bonzagi, Arianna
    Garb, Jane L.
    Jin, Long
    CLINICA CHIMICA ACTA, 2007, 377 (1-2) : 201 - 205
  • [44] Use of the i-STAT portable clinical analyzer in mice
    Peggy Tinkey
    Thomas Lembo
    Suzanne Craig
    Cheri West
    Carolyn Van Pelt
    Lab Animal, 2006, 35 : 45 - 50
  • [45] Evaluation of the blood-saving Kaolin i-STAT ACT technique in paediatric cardiac anaesthesia
    Ulmer, F.
    Bettex, D.
    Baulig, W.
    Buerki, C.
    Weiss, M.
    SWISS MEDICAL WEEKLY, 2008, 138 (41-42) : 11S - 11S
  • [46] Accuracy of capillary whole blood international normalized ratio on the CoaguChek S, CoaguChek XS, and i-STAT 1 point-of-care analyzers
    Karon, Brad S.
    McBane, Robert D., II
    Chaudhry, Rajeev
    Beyer, Lisa K.
    Santrach, Paula J.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 130 (01) : 88 - 92
  • [47] Blood gases in the dog: Comparison of results obtained with the I-STAT portable analyzer and a reference method
    Malet, C
    Verwaerde, P
    Lagente, M
    de La Farge, F
    Braun, JP
    REVUE DE MEDECINE VETERINAIRE, 2000, 151 (07) : 733 - 733
  • [48] Analytical performance of the i-STAT cardiac troponin I assay.
    Apple, FS
    Murakami, MM
    Christenson, RH
    Campbell, JL
    Miller, CJ
    Hock, KG
    Scott, MG
    CLINICAL CHEMISTRY, 2004, 50 (06) : A9 - A9
  • [49] Which point-of-care creatinine analyser for radiology: direct comparison of the i-Stat and StatStrip creatinine methods with different sample types
    Dimeski, Goce
    Tilley, Vera
    Jones, Brock W.
    Brown, Nigel N.
    ANNALS OF CLINICAL BIOCHEMISTRY, 2013, 50 (01) : 47 - 52
  • [50] Evaluation of the i-STAT portable clinical analyzer for use in a helicopter
    Burritt, MF
    Santrach, PJ
    Hankins, DG
    Herr, D
    Newton, NC
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1996, 56 : 121 - 128