Which point-of-care creatinine analyser for radiology: direct comparison of the i-Stat and StatStrip creatinine methods with different sample types

被引:11
|
作者
Dimeski, Goce [1 ]
Tilley, Vera [1 ]
Jones, Brock W. [1 ]
Brown, Nigel N. [1 ]
机构
[1] Princess Alexandra Hosp, Pathol Queensland, Brisbane, Qld 4102, Australia
关键词
CONTRAST-INDUCED NEPHROPATHY; INTERFERENCE; PATIENT; KIDNEY; TIME; RISK;
D O I
10.1258/acb.2012.012081
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Availability of whole blood creatinine estimation for patients scheduled to undergo radiological contrast investigations can provide information to aid patient care by reducing adverse effects and improving departmental efficiencies. Methods: We performed imprecision studies, different patient sample type comparison in 40 participants, and a limited interference study with dopamine and dobutamine on the i-Stat and StatStrip point-of-care enzymatic analysers with the Beckman DxC800 Jaffe assay. Results: Imprecision results showed that the i-Stat performed better. Patient comparison data indicated that the i-Stat provided better correlation than the StatStrip for all the different sample types with correlation coefficients (r(2)) being 0.995-0.996 and 0.918-0.995, respectively. The i-Stat results had a small positive bias of 6-9% for the three different sample types, which required different reference intervals. The StatStrip method showed greater scatter and overall small negative bias of -6% for the whole blood samples and a 10% positive bias with the plasma samples. Dopamine caused significant positive interference with the i-Stat only while dobutamine caused a small negative bias with the StatStrip method only. Conclusions: The findings indicated there are differences offered by the two systems. The StatStrip requires a very small finger prick capillary sample, calculates estimation of the glomerular filtration rate and has an adjustment option to improve correlation with the local method. The i-Stat offers better analytical imprecision and patient comparison with the laboratory method with the three sample types but showed significant interference from dopamine. A final consideration was the availability of middleware to capture patient results with the i-Stat. Based on all the study data, the i-Stat was recommended.
引用
收藏
页码:47 / 52
页数:6
相关论文
共 13 条
  • [1] Performance of i-STAT Point-of-Care Analyser Cartridges After Ambient and Subzero Temperature Exposure
    Fortgens, Philip H.
    du Toit, Leon
    Hofmeyr, Ross
    Dyer, Robert A.
    Swanevelder, Justiaan L.
    POINT OF CARE, 2016, 15 (04): : 132 - 136
  • [2] Analysis of bloodgas, electrolytes and glucose from intraosseous samples using an i-STAT® point-of-care analyser
    Veldhoen, Esther S.
    de Vooght, Karen M. K.
    Slieker, Martijn G.
    Versluys, Anne B.
    Turner, Nigel Mcb.
    RESUSCITATION, 2014, 85 (03) : 359 - 363
  • [3] Assessment of using fingerstick blood sample with i-STAT point-of-care device for cardiac troponin I assay
    Loewenstein, Devin
    Stake, Christine
    Cichon, Mark
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2013, 31 (08): : 1236 - 1239
  • [4] Investigation of Falsely Decreased Creatinine Results Observed From the Abbott I-STAT Point-of-Care Device in Use for Testing Specimens From Ambulatory Oncology Patients
    Bahar, Burak
    DeChristopher, Phillip J.
    Holmes, Earle W.
    Kahn, Stephen E.
    POINT OF CARE, 2016, 15 (02): : 72 - 77
  • [5] Investigating Interferences of a Whole-Blood Point-of-Care Creatinine Analyzer: Comparison to Plasma Enzymatic and Definitive Creatinine Methods in an Acute-Care Setting
    Straseski, Joely A.
    Lyon, Martha E.
    Clarke, William
    DuBois, Jeffrey A.
    Phelan, Lois A.
    Lyon, Andrew W.
    CLINICAL CHEMISTRY, 2011, 57 (11) : 1566 - 1573
  • [6] VALIDATION OF SERUM CREATININE (SCR) MEASUREMENT WITH THE POINT-OF-CARE TESTING (POCT) DEVICE I-STAT IN A MEXICAN GOVERNMENT SPONSORED-CHRONIC KIDNEY DISEASE (CKD) SCREENING PROGRAM
    Obrador, Gregorio
    Olvera, Nadia
    Ortiz de la Pe, Daniela
    Villa, Antonio
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2012, 27 : 100 - 100
  • [7] Comparison of point-of-care testing (POCT): i-STAT® international normalized ratio (INR) vs reference laboratory INR in pediatric patients undergoing major surgery
    Spielmann, Nelly
    Mauch, Jacqueline Y.
    Madjdpour, Caveh
    Schmugge, Markus
    Albisetti, Manuela
    Weiss, Markus
    Haas, Thorsten
    PEDIATRIC ANESTHESIA, 2011, 21 (10) : 1041 - 1045
  • [8] Evaluation of the i-STAT point-of-care capillary whole blood prothrombin time and international normalized ratio: Comparison to the Tcoag MDAII coagulation analyzer in the central laboratory
    Pena, J. A.
    Lewandrowski, K. B.
    Lewandrowski, E. L.
    Gregory, K.
    Baron, J. M.
    Van Cott, E. M.
    CLINICA CHIMICA ACTA, 2012, 413 (11-12) : 955 - 959
  • [9] Evaluation of a new point-of-care celite-activated clotting time analyzer in different clinical settings -: The i-STAT celite-activated clotting time test
    Paniccia, R
    Fedi, S
    Carbonetto, F
    Noferí, D
    Conti, P
    Bandinelli, B
    Giusti, B
    Evangelisti, L
    Pretelli, P
    Palmarini, MFG
    Abbate, R
    Prisco, D
    ANESTHESIOLOGY, 2003, 99 (01) : 54 - 59
  • [10] Evaluation of the i-STAT point-of-care capillary whole blood hematocrit and hemoglobin: Comparison to the Siemens RAPIDLab 1200, Sysmex XE5000, and manual spun hematocrit
    Rudolf, Joseph
    Douglass, Joseph
    Baron, Jason
    Lewandrowski, Kent
    CLINICA CHIMICA ACTA, 2015, 446 : 37 - 42