Accuracy of the i-STAT™ bedside blood gas analyser

被引:54
|
作者
Sediame, S
Zerah-Lancner, F
d'Ortho, MP
Adnot, S
Harf, A [1 ]
机构
[1] Hop Henri Mondor, Serv Physiol Explorat Fonctionnelles, F-94010 Creteil, France
[2] Hop Henri Mondor, INSERM, U492, F-94010 Creteil, France
关键词
acidosis; high oxygen tension; permissive hypercapnia; point-of-care testing; thin-film electrode biosensors;
D O I
10.1034/j.1399-3003.1999.14a36.x
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
The performance of the i-STAT portable clinical analyser for measuring blood gases and pH was evaluated,vith reference to a conventional blood gas analyser (ABL520 Radiometer). Ninety-two samples from the routine blood gas analysis laboratory were chosen according to a wide distribution of partial pressure of carbon dioxide (Pa,CO2), partial pressure oxygen (Pa,O-2) and pH and then analysed. All measurements were performed in duplicate by trained technicians from the central hospital laboratory. Differences between duplicate measurements were computed for Pa,CO2: (1.2 versus 0.4%), Pa,O-2 (1.7 versus 1.1%) and pH (0.06 versus 0.02%), for the i-STAT and ABL520, respectively. pH and Pa,CO2 values measured with the i-STAT were very close to those obtained with the ABL520, the difference (mean+/-sD) being 0.006+/-0.018 and -0.13+/-0.17 kPa, respectively. Statistical analysis showed that the differences between analysers did not depend on values of pH or Pa,CO2 The performance of the analysers depended on the level of PO2. Below 15 kPa (n=48), the two systems gave nearly identical values, the mean difference was 0.01+/-0.37 kPa. Between 16 and 55 kPa (n=44), there was a systematic but small (-0.69+0.67 kPa) underestimation of Pa,O-2 measured with the i-STAT (p<10(-8)). In conclusion, this study shows that blood gas analysis using the i-STAT portable device is comparable,vith that performed by a conventional laboratory blood gas analyser.
引用
收藏
页码:214 / 217
页数:4
相关论文
共 50 条
  • [1] Evaluation of the portable blood gas analyser i-STAT
    Tschudi, PR
    SCHWEIZER ARCHIV FUR TIERHEILKUNDE, 1998, 140 (12): : 507 - 512
  • [2] The reliability of the i-STAT clinical portable analyser
    Dascombe, B. J.
    Reaburn, P. R. J.
    Sirotic, A. C.
    Coutts, A. J.
    JOURNAL OF SCIENCE AND MEDICINE IN SPORT, 2007, 10 (03) : 135 - 140
  • [3] The accuracy of the i-STAT portable analyser for measuring blood gases and pH in whole-blood samples from dogs
    Verwaerde, P
    Malet, C
    Lagente, M
    De la Farge, F
    Braun, JP
    RESEARCH IN VETERINARY SCIENCE, 2002, 73 (01) : 71 - 75
  • [4] The i-STAT analyser, a new, hand-held whole blood analyser for measurement of haematocrit, electrolytes, and blood gases
    Schneider, J
    Dudziak, K
    Westphal, K
    Vettermann, J
    ANAESTHESIST, 1997, 46 (08): : 704 - 714
  • [5] CARTRIDGE TEMPERATURE SIGNIFICANTLY EFFECTS PAO2 RESULTS USING THE I-STAT PORTABLE CLINICAL ANALYSER (I-STAT)
    Harnett, Lauren A.
    Johnston, Sonya D.
    Thornton, Andrew T.
    Yeo, Aeneas
    RESPIROLOGY, 2012, 17 : 6 - 6
  • [6] AN EVALUATION OF BLOOD-GAS AND IONIZED CALCIUM MEASUREMENT OF THE I-STAT ANALYZER
    MOCK, T
    YATSCOFF, RW
    CLINICAL CHEMISTRY, 1995, 41 (06) : S182 - S182
  • [7] Comparison of the EPOC and i-STAT analysers for canine blood gas and electrolyte analysis
    West, E.
    Bardell, D.
    Senior, J. M.
    JOURNAL OF SMALL ANIMAL PRACTICE, 2014, 55 (03) : 139 - 144
  • [8] The portable laboratory:: an evaluation of the accuracy and reproducibility of i-STAT©
    Bingham, D
    Kendall, J
    Clancy, M
    ANNALS OF CLINICAL BIOCHEMISTRY, 1999, 36 : 66 - 71
  • [9] The i-STAT analyser, a new, hand-held whole blood analyser for measurement of haematocrit, electrolytes, and blood gases [Der i-STAT Analyzer. Ein neues, tragbares Gerat zur Bedsidebestimmung des Hamatokrits, der Blutgase und Elektrolyte]
    Schneider J.
    Dudziak R.
    Westphal K.
    Vettermann J.
    Der Anaesthesist, 1997, 46 (8) : 704 - 714
  • [10] Accuracy of the Glucose Test in the i-STAT® G Cartridge Using the i-STAT 1 System for Venous, Arterial and Capillary Whole Blood Specimens in Clinical Settings
    Kornfeld, S. F.
    Aguirre, M. Gonzalez
    Zhang, H.
    Hahn, S.
    CLINICAL CHEMISTRY, 2024, 70