Variability in RF-induced heating of a deep brain stimulation implant across MR systems

被引:70
|
作者
Baker, Kenneth B.
Tkach, Jean A.
Phillips, Micheal D.
Rezai, Ali R.
机构
[1] Cleveland Clin Fdn, Ctr Neurol Restorat, Cleveland, OH 44195 USA
[2] Cleveland Clin Fdn, Dept Radiol, Cleveland, OH 44195 USA
[3] Case Western Reserve Univ, Dept Radiol, Cleveland, OH 44106 USA
关键词
magnetic resonance imaging; bioeffects; safety; neurostimulation systems; implants; deep brain stimulation; specific absorption rate;
D O I
10.1002/jmri.20769
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the MRI-related heating per unit of specific absorption rate (SAR) profile of a conductive implant between two 1.5-Tesla/64 MHz MR systems using a transmit/receive (t/r) head coil configuration. Materials and Methods: Deep brain stimulation (DBS) leads were configured within a gel-filled phantom of the human head and torso. Temperature variation at each of four contacts of the bilaterally-placed leads was monitored using fluoroptic thermometry. MRI was performed using the t/r head coils of two different-generation 1.5-Tesla MR systems from the same manufacturer. Temperature changes were normalized to SAR values for the head (Delta T/SAR-H), and the slope of this Delta T/SAR-H by time relationship was compared between the two scanners. Results: The Delta T/SAR-H for the implant ranged from 3.5 to 5.5 times higher on one MR system as compared to the other (P < 0.01) depending on the measurement site. Conclusion: The findings support previous observations that console-reported SAR does not constitute a reliable index of heating for elongated, conductive implants, such as the DBS hardware system tested. In contrast to our previous findings using a t/r body coil, the data presented here reveal marked differences between two MR systems using t/r head coils (the coil configuration was consistent with the implant manufacturer's imaging guidelines).
引用
收藏
页码:1236 / 1242
页数:7
相关论文
共 50 条
  • [21] Study on Search Strategies for Assessing the Worst Case RF-Induced Heating for Multi-Configuration Implant System Under MRI
    Zheng, Jianfeng
    Ji, Xiaohe
    Kainz, Wolfgang
    Chen, Ji
    [J]. IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2020, 62 (01) : 43 - 51
  • [23] Safety of active catheters in MRI: Termination impedance versus RF-induced heating
    Oezen, Ali Caglar
    Lottner, Thomas
    Bock, Michael
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (02) : 1412 - 1423
  • [24] RF-induced heating for Cardiac Rhythm Management (CRM) in patients with different postures
    Yang, Xiaolin
    Guo, Ran
    Zheng, Jianfeng
    Chen, Ji
    [J]. 2022 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY & SIGNAL/POWER INTEGRITY, EMCSI, 2022, : 606 - 609
  • [25] Impacts of MRI frequency on RF-induced Heating for External Fixation with Insulating Material
    Yang, Rui
    Zheng, Jianfeng
    Wang, Yu
    Guo, Ran
    Chen, Ji
    [J]. 2019 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION AND USNC-URSI RADIO SCIENCE MEETING, 2019, : 529 - 530
  • [26] Novel Design of Implantable Esophageal Stent to Reduce the MRI RF-Induced Heating
    Li, Dawei
    Ji, Xiaohe
    Zheng, Jianfeng
    Pan, Changwang
    Chen, Ji
    Kainz, Wolfgang
    [J]. IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2017, 59 (03) : 805 - 812
  • [27] MRI thermometry: Fast mapping of RF-induced heating along conductive wires
    Ehses, Philipp
    Fidler, Florian
    Nordbeck, Peter
    Pracht, Eberhard D.
    Warmuth, Marcus
    Jakob, Peter M.
    Bauer, Wolfgang R.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (02) : 457 - 461
  • [28] Reducing MRI RF-induced heating for the external fixation using capacitive structures
    Zheng, Jianfeng
    Yang, Rui
    Wang, Qingyan
    Guo, Ran
    Xu, Jian
    Shou, Xingxian
    Kainz, Wolfgang
    Chen, Ji
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (15):
  • [29] A Novel Design of Implantable Medical Stent for Reducing the MRI RF-Induced Heating
    Li, Dawei
    Ji, Xiaohe
    Zheng, Jianfeng
    Chen, Ji
    Pan, Changwang
    [J]. 2016 IEEE ANTENNAS AND PROPAGATION SOCIETY INTERNATIONAL SYMPOSIUM, 2016, : 1945 - 1946
  • [30] Influence of a Metallic Shield on RF-Induced Heating of a Lead with Straight and Helical Wires
    Kozlov, Mikhail
    Kainz, Wolfgang
    Daniel, Luca
    [J]. 2019 IEEE MTT-S INTERNATIONAL CONFERENCE ON NUMERICAL ELECTROMAGNETIC AND MULTIPHYSICS MODELING AND OPTIMIZATION (NEMO 2019), 2019,