Reducing MRI RF-induced heating for the external fixation using capacitive structures

被引:2
|
作者
Zheng, Jianfeng [1 ]
Yang, Rui [1 ]
Wang, Qingyan [1 ]
Guo, Ran [1 ]
Xu, Jian [2 ]
Shou, Xingxian [2 ]
Kainz, Wolfgang [3 ]
Chen, Ji [1 ]
机构
[1] Univ Houston, Dept Elect & Comp Engn, Houston, TX 77204 USA
[2] UIH Amer Inc, Houston, TX 77054 USA
[3] US FDA, Div Biomed Phys, Off Sci & Engn Labs, Ctr Devices & Radiol Hlth, Silver Spring, MD 20993 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2020年 / 65卷 / 15期
关键词
RF-induced heating; magnetic resonance imaging; external fixation device; insulating material; DEEP BRAIN-STIMULATION; IMPLANT; SAFETY;
D O I
10.1088/1361-6560/ab9706
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents a generic method to reduce the radiofrequency (RF) induced heating of external fixation devices during the magnetic resonance imaging (MRI) procedure. A simplified equivalent circuit model was developed to illustrate the interaction between the external fixation device and the MRI RF field. Carefully designed mechanical structures, which utilize capacitive reactance from the circuit model, were applied to the external fixation device to mitigate the coupling between the external fixation device and the MRI RF field for RF-induced heating reduction. Both numerical and experimental studies were performed to demonstrate the validity of the circuit model and the effectiveness of the proposed method. By adding capacitive structures in both the clamp-pin and rod-clamp joints, the peak specific absorption rate averaged in 1 gram (SAR(1g)) near the pin tips were reduced from 760.4 W kg(-1)to 12.0 W kg(-1)at 1.5 T and 391.5 W kg(-1)to 25.2 W kg(-1)at 3 T from numerical simulations. Experimental results showed that RF-induced heating was reduced from 7.85 degrees C to 1.01 degrees C at 1.5 T and from 16.70 degrees C to 0.32 degrees C at 3 T for the external fixation device studied here. The carefully designed capacitive structures can be used to detune the coupling between the external fixation device and the MRI fields to reduce the RF-induced heating in the human body for both 1.5 T and 3 T MRI systems. However, as RF-induced heating is very device and design specific all devices must be thoroughly tested based on its final design.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Impacts of MRI frequency on RF-induced Heating for External Fixation with Insulating Material
    Yang, Rui
    Zheng, Jianfeng
    Wang, Yu
    Guo, Ran
    Chen, Ji
    2019 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION AND USNC-URSI RADIO SCIENCE MEETING, 2019, : 529 - 530
  • [2] Effect of insulating layer material on RF-induced heating for external fixation system in 1.5 T MRI system
    Liu, Yan
    Kainz, Wolfgang
    Qian, Songsong
    Wu, Wen
    Chen, Ji
    ELECTROMAGNETIC BIOLOGY AND MEDICINE, 2014, 33 (03) : 223 - 227
  • [3] A Novel Design of Implantable Medical Stent for Reducing the MRI RF-Induced Heating
    Li, Dawei
    Ji, Xiaohe
    Zheng, Jianfeng
    Chen, Ji
    Pan, Changwang
    2016 IEEE ANTENNAS AND PROPAGATION SOCIETY INTERNATIONAL SYMPOSIUM, 2016, : 1945 - 1946
  • [4] Numerical Investigations of MRI RF-Induced Heating for External Fixation Device in TEM and Birdcage Body Coils at 3 T
    Yang, Rui
    Zheng, Jianfeng
    Kainz, Wolfgang
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2018, 60 (03) : 598 - 604
  • [5] Effects of patient orientations, landmark positions, and device positions on the MRI RF-induced heating for modular external fixation devices
    Xia, Meiqi
    Zheng, Jianfeng
    Yang, Rui
    Song, Shuo
    Xu, Jian
    Liu, Qi
    Kainz, Wolfgang
    Long, Stuart A.
    Chen, Ji
    MAGNETIC RESONANCE IN MEDICINE, 2021, 85 (03) : 1669 - 1680
  • [6] MRI safety: RF-induced heating near straight wires
    Park, SM
    Kamondetdacha, R
    Amjad, A
    Nyenhuis, JA
    IEEE TRANSACTIONS ON MAGNETICS, 2005, 41 (10) : 4197 - 4199
  • [7] Numerical investigations of MRI RF field induced heating for external fixation devices
    Yan Liu
    Jianxiang Shen
    Wolfgang Kainz
    Songsong Qian
    Wen Wu
    Ji Chen
    BioMedical Engineering OnLine, 12
  • [8] Numerical investigations of MRI RF field induced heating for external fixation devices
    Liu, Yan
    Shen, Jianxiang
    Kainz, Wolfgang
    Qian, Songsong
    Wu, Wen
    Chen, Ji
    BIOMEDICAL ENGINEERING ONLINE, 2013, 12
  • [9] Reducing RF-Induced Heating Near Implanted Leads Through High-Dielectric Capacitive Bleeding of Current (CBLOC)
    Golestanirad, Laleh
    Angelone, Leonardo M.
    Kirsch, John
    Downs, Sean
    Keil, Boris
    Bonmassar, Giorgio
    Wald, Lawrence L.
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2019, 67 (03) : 1265 - 1273
  • [10] Fast Prediction of MRI RF-induced Heating For Implantable Plate Devices Using Neural Network
    Zheng, Jianfeng
    Yang, Rui
    Chen, Ji
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1011 - 1012