A Novel Design of Implantable Medical Stent for Reducing the MRI RF-Induced Heating

被引:0
|
作者
Li, Dawei [1 ]
Ji, Xiaohe [1 ]
Zheng, Jianfeng [1 ]
Chen, Ji [1 ]
Pan, Changwang [2 ]
机构
[1] Univ Houston, Dept Elect & Comp Engn, Houston, TX 77204 USA
[2] Micro Tech Co Ltd, Nanjing 210061, Jiangsu, Peoples R China
关键词
Magnetic resonance imaging (MRI); Specific absorption rate (SAR); Medical implants; Stent; MRI safety; RFinduced heating;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this work, a novel implantable medical stent is designed to reduce the radio frequency (RF)-induced heating from implantable medical stent under magnetic resonance imaging (MRI) procedure. By using segmented structure, the induced specific absorption rate (SAR) for a 100 mm length stent is reduced from 54.4 mW/g to 13.4 mW/g. Numerical simulation demonstrates the effectiveness of the designed structure.
引用
收藏
页码:1945 / 1946
页数:2
相关论文
共 50 条
  • [1] Novel Design of Implantable Esophageal Stent to Reduce the MRI RF-Induced Heating
    Li, Dawei
    Ji, Xiaohe
    Zheng, Jianfeng
    Pan, Changwang
    Chen, Ji
    Kainz, Wolfgang
    [J]. IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2017, 59 (03) : 805 - 812
  • [2] A technique for the reduction of RF-induced heating of active implantable medical devices during MRI
    Wang, Yu
    Zheng, Jianfeng
    Guo, Ran
    Wang, Qingyan
    Kainz, Wolfgang
    Long, Stuart
    Chen, Ji
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2022, 87 (01) : 349 - 364
  • [3] Effects of Strain Relief Loop on MRI RF-Induced Heating of Active Implantable Medical Devices
    Akter, Mir Khadiza
    Guo, Ran
    Islam, Md Zahidul
    Zheng, Jianfeng
    Chen, Ji
    [J]. PROCEEDINGS OF THE 2022 IEEE TEXAS SYMPOSIUM ON WIRELESS AND MICROWAVE CIRCUITS AND SYSTEMS (WMCS), 2022,
  • [4] Impact of RF Shimming on RF-Induced Heating Near Implantable Medical Electrodes in a 3T MRI Coil
    Zeng, Qi
    Wang, Qingyan
    Kainz, Wolfgang
    Chen, Ji
    [J]. IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2020, 62 (01) : 52 - 64
  • [5] Prediction of MRI RF-induced Heating for Passive Implantable Medical Devices Using Convolutional Neural Network
    Lan, Qianlong
    Zheng, Jianfeng
    Chen, Ji
    Zhang, Mingchi
    [J]. 2020 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY AND SIGNAL & POWER INTEGRITY VIRTUAL SYMPOSIUM(IEEE EMC+SIPI), 2020, : 270 - 275
  • [6] RF-induced heating for active implantable medical devices in dual parallel leads configurations at 1.5 T MRI
    Hu, Wei
    Guo, Ran
    Wang, Qingyan
    Zheng, Jianfeng
    Tsang, Jeffrey
    Kainz, Wolfgang
    Long, Stuart
    Chen, Ji
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2023, 90 (02) : 686 - 698
  • [7] Reducing MRI RF-induced heating for the external fixation using capacitive structures
    Zheng, Jianfeng
    Yang, Rui
    Wang, Qingyan
    Guo, Ran
    Xu, Jian
    Shou, Xingxian
    Kainz, Wolfgang
    Chen, Ji
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (15):
  • [8] Impact of Patient Postures on the RF-Induced Heating for the Active Implantable Medical Device (AIMD) at 1.5T MRI
    Yang, Xiaolin
    Guo, Ran
    Zheng, Jianfeng
    Chen, Xuemin
    Chen, Ji
    [J]. IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2024, 66 (04) : 1029 - 1040
  • [9] A counterpoise design for RF-induced heating reduction
    Wang, Yu
    Zheng, Jianfeng
    Wang, Qingyan
    Chen, Ji
    [J]. 42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20, 2020, : 4200 - 4203
  • [10] Fast Prediction of MRI RF-induced Heating For Implantable Plate Devices Using Neural Network
    Zheng, Jianfeng
    Yang, Rui
    Chen, Ji
    [J]. 2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1011 - 1012