Reducing MRI RF-induced heating for the external fixation using capacitive structures

被引:2
|
作者
Zheng, Jianfeng [1 ]
Yang, Rui [1 ]
Wang, Qingyan [1 ]
Guo, Ran [1 ]
Xu, Jian [2 ]
Shou, Xingxian [2 ]
Kainz, Wolfgang [3 ]
Chen, Ji [1 ]
机构
[1] Univ Houston, Dept Elect & Comp Engn, Houston, TX 77204 USA
[2] UIH Amer Inc, Houston, TX 77054 USA
[3] US FDA, Div Biomed Phys, Off Sci & Engn Labs, Ctr Devices & Radiol Hlth, Silver Spring, MD 20993 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2020年 / 65卷 / 15期
关键词
RF-induced heating; magnetic resonance imaging; external fixation device; insulating material; DEEP BRAIN-STIMULATION; IMPLANT; SAFETY;
D O I
10.1088/1361-6560/ab9706
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents a generic method to reduce the radiofrequency (RF) induced heating of external fixation devices during the magnetic resonance imaging (MRI) procedure. A simplified equivalent circuit model was developed to illustrate the interaction between the external fixation device and the MRI RF field. Carefully designed mechanical structures, which utilize capacitive reactance from the circuit model, were applied to the external fixation device to mitigate the coupling between the external fixation device and the MRI RF field for RF-induced heating reduction. Both numerical and experimental studies were performed to demonstrate the validity of the circuit model and the effectiveness of the proposed method. By adding capacitive structures in both the clamp-pin and rod-clamp joints, the peak specific absorption rate averaged in 1 gram (SAR(1g)) near the pin tips were reduced from 760.4 W kg(-1)to 12.0 W kg(-1)at 1.5 T and 391.5 W kg(-1)to 25.2 W kg(-1)at 3 T from numerical simulations. Experimental results showed that RF-induced heating was reduced from 7.85 degrees C to 1.01 degrees C at 1.5 T and from 16.70 degrees C to 0.32 degrees C at 3 T for the external fixation device studied here. The carefully designed capacitive structures can be used to detune the coupling between the external fixation device and the MRI fields to reduce the RF-induced heating in the human body for both 1.5 T and 3 T MRI systems. However, as RF-induced heating is very device and design specific all devices must be thoroughly tested based on its final design.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] RF-induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T
    Islam, Md Zahidul
    Guo, Ran
    Akter, Mir Khadiza
    Zheng, Jianfeng
    Kainz, Wolfgang
    Long, Stuart
    Chen, Ji
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (05) : 2108 - 2122
  • [22] Thermo-Acoustic Ultrasound for Detection of RF-Induced Device Lead Heating in MRI
    Dixit, Neerav
    Stang, Pascal P.
    Pauly, John M.
    Scott, Greig C.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (02) : 536 - 546
  • [23] Computational Study of External Fixation Devices Surface Heating in MRI RF Environment
    Liu, Yan
    Shen, Jianxiang
    Kainz, Wolfgang
    Qian, Songsong
    Wu, Wen
    Chen, Ji
    2012 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY (EMC), 2012, : 602 - 606
  • [24] rf-induced heating dynamics of noncrystallized trapped ions
    van Mourik, Martin W.
    Hrmo, Pavel
    Gerster, Lukas
    Wilhelm, Benjamin
    Blatt, Rainer
    Schindler, Philipp
    Monz, Thomas
    PHYSICAL REVIEW A, 2022, 105 (03)
  • [25] RF-induced heating of interventional devices at 23.66 MHz
    Ali Caglar Özen
    Maximilian Frederik Russe
    Thomas Lottner
    Simon Reiss
    Sebastian Littin
    Maxim Zaitsev
    Michael Bock
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2023, 36 : 439 - 449
  • [26] RF-induced heating of interventional devices at 23.66 MHz
    Oezen, Ali Caglar
    Russe, Maximilian Frederik
    Lottner, Thomas
    Reiss, Simon
    Littin, Sebastian
    Zaitsev, Maxim
    Bock, Michael
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2023, 36 (03) : 439 - 449
  • [27] Preliminary Study of RF-Induced Heating of DBS and RNS Devices under 1.5 T MRI
    Akter, Mir Khadiza
    Guo, Ran
    Islam, Md Zahidul
    Zheng, Jianfeng
    Chen, Ji
    2024 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION AND INC/USNCURSI RADIO SCIENCE MEETING, AP-S/INC-USNC-URSI 2024, 2024, : 2655 - 2656
  • [28] Electromagnetic Modeling of the Implantable Electrode for Transfer Function Calibration in MRI RF-Induced Heating Assessment
    Long, Tiangang
    Jiang, Changqing
    Li, Luming
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2024, 8 (03): : 259 - 264
  • [29] Impacts of Tip Structure on RF-induced Heating of an Implantable Neurostimulator under 1.5 T MRI
    Yang, Rui
    Zheng, Jianfeng
    Chen, Ji
    2018 IEEE ANTENNAS AND PROPAGATION SOCIETY INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2018, : 2383 - 2384
  • [30] RF-induced heating effects of metallic biomaterial designs at 1.5T MRI exposure
    Akdogan, Gulsen
    Istanbullu, O. Burak
    INTERNATIONAL JOURNAL OF BIOMEDICAL ENGINEERING AND TECHNOLOGY, 2023, 41 (04) : 376 - 393