RF-induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T

被引:0
|
作者
Islam, Md Zahidul [1 ]
Guo, Ran [1 ]
Akter, Mir Khadiza [1 ]
Zheng, Jianfeng [1 ]
Kainz, Wolfgang [2 ]
Long, Stuart [1 ]
Chen, Ji [1 ]
机构
[1] Univ Houston, Dept Elect & Comp Engn, Houston, TX 77204 USA
[2] High Performance Comp MRI Safety, Jasper, GA USA
基金
美国国家科学基金会;
关键词
magnetic resonance imaging (MRI); MR safety; partially in and partially out (PIPO) medical implants; radio frequency (RF) heating mitigation; trajectory modification; transfer function (TF); SAFETY; LEAD; TRANSMISSION; ELECTRODES; FIELDS; MODEL;
D O I
10.1002/mrm.30389
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo address the issue of RF-induced heating for partially in and partially out (PIPO) medical devices during 1.5 T MRI scans by proposing a method of minimizing the external portion.MethodsA method of tightly winding the external segment of the PIPO device is proposed to minimize the overall device effective reception length during MRI scans to mitigate the RF-induced heating. Two commercially available PIPO medical devices and simplified solid wires were used to demonstrate the concept. RF heating results are compared between typical and minimized-length trajectories under the American Society for Testing and Materials (ASTM) testing procedure. In addition, 16 scaled and validated device models were used in conjuncture with human body numerical simulations within three virtual human models to estimate clinically relevant heating.ResultsThe wound segments in PIPO devices functioned as a lumped element rather than a receiving antenna, reducing induced energy/heating as compared to the original PIPO devices under typical straight or loop configurations. Minimizing the lead's external portion can reduce the RF-induced heating by significant factors for all studied cases during ASTM phantom measurements and in human body simulations.ConclusionOur findings show a significant reduction in RF heating by minimizing the external segment, thereby enhancing patient safety during 1.5 T MRI procedures. Although limited to four devices at 1.5 T across two applications, the extent of heating reduction may vary for others. Nonetheless, tightly winding the external segment of PIPO electrodes holds promise for improving device safety under MRI.
引用
收藏
页码:2108 / 2122
页数:15
相关论文
共 50 条
  • [1] Preliminary Study of RF-Induced Heating of DBS and RNS Devices under 1.5 T MRI
    Akter, Mir Khadiza
    Guo, Ran
    Islam, Md Zahidul
    Zheng, Jianfeng
    Chen, Ji
    2024 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION AND INC/USNCURSI RADIO SCIENCE MEETING, AP-S/INC-USNC-URSI 2024, 2024, : 2655 - 2656
  • [2] RF-induced heating for active implantable medical devices in dual parallel leads configurations at 1.5 T MRI
    Hu, Wei
    Guo, Ran
    Wang, Qingyan
    Zheng, Jianfeng
    Tsang, Jeffrey
    Kainz, Wolfgang
    Long, Stuart
    Chen, Ji
    MAGNETIC RESONANCE IN MEDICINE, 2023, 90 (02) : 686 - 698
  • [3] A technique for the reduction of RF-induced heating of active implantable medical devices during MRI
    Wang, Yu
    Zheng, Jianfeng
    Guo, Ran
    Wang, Qingyan
    Kainz, Wolfgang
    Long, Stuart
    Chen, Ji
    MAGNETIC RESONANCE IN MEDICINE, 2022, 87 (01) : 349 - 364
  • [4] Evaluations of the MRI RF-Induced Heating for Helical Stents Under a 1.5T MRI System
    Ji, Xiaohe
    Zheng, Jianfeng
    Yang, Rui
    Kainz, Wolfgang
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2019, 61 (01) : 57 - 64
  • [5] Impacts of Tip Structure on RF-induced Heating of an Implantable Neurostimulator under 1.5 T MRI
    Yang, Rui
    Zheng, Jianfeng
    Chen, Ji
    2018 IEEE ANTENNAS AND PROPAGATION SOCIETY INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2018, : 2383 - 2384
  • [6] Effect of Strain Relief Loop Position on the RF-Induced Heating of Active Implantable Medical Devices at 1.5-T MRI
    Akter, Mir Khadiza
    Guo, Ran
    Islam, Md Zahidul
    Zheng, Jianfeng
    Kainz, Wolfgang
    Long, Stuart A.
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2024, 66 (04) : 1041 - 1056
  • [7] Effect of insulating layer material on RF-induced heating for external fixation system in 1.5 T MRI system
    Liu, Yan
    Kainz, Wolfgang
    Qian, Songsong
    Wu, Wen
    Chen, Ji
    ELECTROMAGNETIC BIOLOGY AND MEDICINE, 2014, 33 (03) : 223 - 227
  • [8] A CP-ANN-Based RF-Induced Heating Estimation Method for Passive Orthopaedic Implantable Devices Under 1.5 T and 3.0 T MRI
    Yao, Aiping
    Pei, Yunfeng
    He, Junchen
    Wang, Jing
    Yang, Pengfei
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2024, 66 (02) : 405 - 416
  • [9] Impact of Patient Postures on the RF-Induced Heating for the Active Implantable Medical Device (AIMD) at 1.5T MRI
    Yang, Xiaolin
    Guo, Ran
    Zheng, Jianfeng
    Chen, Xuemin
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2024, 66 (04) : 1029 - 1040
  • [10] RF-induced heating effects of metallic biomaterial designs at 1.5T MRI exposure
    Akdogan, Gulsen
    Istanbullu, O. Burak
    INTERNATIONAL JOURNAL OF BIOMEDICAL ENGINEERING AND TECHNOLOGY, 2023, 41 (04) : 376 - 393