RF-induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T

被引:0
|
作者
Islam, Md Zahidul [1 ]
Guo, Ran [1 ]
Akter, Mir Khadiza [1 ]
Zheng, Jianfeng [1 ]
Kainz, Wolfgang [2 ]
Long, Stuart [1 ]
Chen, Ji [1 ]
机构
[1] Univ Houston, Dept Elect & Comp Engn, Houston, TX 77204 USA
[2] High Performance Comp MRI Safety, Jasper, GA USA
基金
美国国家科学基金会;
关键词
magnetic resonance imaging (MRI); MR safety; partially in and partially out (PIPO) medical implants; radio frequency (RF) heating mitigation; trajectory modification; transfer function (TF); SAFETY; LEAD; TRANSMISSION; ELECTRODES; FIELDS; MODEL;
D O I
10.1002/mrm.30389
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo address the issue of RF-induced heating for partially in and partially out (PIPO) medical devices during 1.5 T MRI scans by proposing a method of minimizing the external portion.MethodsA method of tightly winding the external segment of the PIPO device is proposed to minimize the overall device effective reception length during MRI scans to mitigate the RF-induced heating. Two commercially available PIPO medical devices and simplified solid wires were used to demonstrate the concept. RF heating results are compared between typical and minimized-length trajectories under the American Society for Testing and Materials (ASTM) testing procedure. In addition, 16 scaled and validated device models were used in conjuncture with human body numerical simulations within three virtual human models to estimate clinically relevant heating.ResultsThe wound segments in PIPO devices functioned as a lumped element rather than a receiving antenna, reducing induced energy/heating as compared to the original PIPO devices under typical straight or loop configurations. Minimizing the lead's external portion can reduce the RF-induced heating by significant factors for all studied cases during ASTM phantom measurements and in human body simulations.ConclusionOur findings show a significant reduction in RF heating by minimizing the external segment, thereby enhancing patient safety during 1.5 T MRI procedures. Although limited to four devices at 1.5 T across two applications, the extent of heating reduction may vary for others. Nonetheless, tightly winding the external segment of PIPO electrodes holds promise for improving device safety under MRI.
引用
收藏
页码:2108 / 2122
页数:15
相关论文
共 50 条
  • [21] Investigation of RF-Induced Heating Near Interventional Catheters at 1.5 T MRI: A Combined Modeling and Experimental Study
    Zeng, Qi
    Liu, Jingshen
    Angelone, Leonardo M.
    Lloyd, Tom
    Wedan, Steve
    Chen, Ji
    Kainz, Wolfgang
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2019, 61 (05) : 1423 - 1431
  • [22] Numerical evaluation of RF-induced heating for various esophageal stent designs under MRI 1.5 Tesla system
    Ji, Xiaohe
    Zheng, Jianfeng
    Chen, Ji
    ELECTROMAGNETIC BIOLOGY AND MEDICINE, 2017, 36 (04) : 379 - 386
  • [23] Impact of RF Shimming on RF-Induced Heating Near Implantable Medical Electrodes in a 3T MRI Coil
    Zeng, Qi
    Wang, Qingyan
    Kainz, Wolfgang
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2020, 62 (01) : 52 - 64
  • [24] A Novel Design of Implantable Medical Stent for Reducing the MRI RF-Induced Heating
    Li, Dawei
    Ji, Xiaohe
    Zheng, Jianfeng
    Chen, Ji
    Pan, Changwang
    2016 IEEE ANTENNAS AND PROPAGATION SOCIETY INTERNATIONAL SYMPOSIUM, 2016, : 1945 - 1946
  • [25] Effects of patient orientations, landmark positions, and device positions on the MRI RF-induced heating for modular external fixation devices
    Xia, Meiqi
    Zheng, Jianfeng
    Yang, Rui
    Song, Shuo
    Xu, Jian
    Liu, Qi
    Kainz, Wolfgang
    Long, Stuart A.
    Chen, Ji
    MAGNETIC RESONANCE IN MEDICINE, 2021, 85 (03) : 1669 - 1680
  • [26] Numerical Investigations of MRI RF-Induced Heating for External Fixation Device in TEM and Birdcage Body Coils at 3 T
    Yang, Rui
    Zheng, Jianfeng
    Kainz, Wolfgang
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2018, 60 (03) : 598 - 604
  • [27] Fast Prediction of MRI RF-induced Heating For Implantable Plate Devices Using Neural Network
    Zheng, Jianfeng
    Yang, Rui
    Chen, Ji
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1011 - 1012
  • [28] A Transmission Line Model for the Evaluation of MRI RF-Induced Fields on Active Implantable Medical Devices
    Liu, Jingshen
    Zheng, Jianfeng
    Wang, Qingyan
    Kainz, Wolfgang
    Chen, Ji
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2018, 66 (09) : 4271 - 4281
  • [29] Impact of Electrode Structure on RF-Induced Heating for an AIMD Implanted Lead in a 1.5-Tesla MRI System
    Yang, Rui
    Zheng, Jianfeng
    Wang, Yu
    Guo, Ran
    Kainz, Wolfgang
    Chen, Ji
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2019, 3 (04): : 247 - 253
  • [30] Calculating RF-Induced Voltages for Implanted Medical Devices in MRI Using Computational Human Models
    Brown, James E.
    Qiang, Rui
    Stadnik, Paul J.
    Stotts, Larry J.
    Von Arx, Jeffrey A.
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 3866 - 3869