RF-induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T

被引:0
|
作者
Islam, Md Zahidul [1 ]
Guo, Ran [1 ]
Akter, Mir Khadiza [1 ]
Zheng, Jianfeng [1 ]
Kainz, Wolfgang [2 ]
Long, Stuart [1 ]
Chen, Ji [1 ]
机构
[1] Univ Houston, Dept Elect & Comp Engn, Houston, TX 77204 USA
[2] High Performance Comp MRI Safety, Jasper, GA USA
基金
美国国家科学基金会;
关键词
magnetic resonance imaging (MRI); MR safety; partially in and partially out (PIPO) medical implants; radio frequency (RF) heating mitigation; trajectory modification; transfer function (TF); SAFETY; LEAD; TRANSMISSION; ELECTRODES; FIELDS; MODEL;
D O I
10.1002/mrm.30389
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo address the issue of RF-induced heating for partially in and partially out (PIPO) medical devices during 1.5 T MRI scans by proposing a method of minimizing the external portion.MethodsA method of tightly winding the external segment of the PIPO device is proposed to minimize the overall device effective reception length during MRI scans to mitigate the RF-induced heating. Two commercially available PIPO medical devices and simplified solid wires were used to demonstrate the concept. RF heating results are compared between typical and minimized-length trajectories under the American Society for Testing and Materials (ASTM) testing procedure. In addition, 16 scaled and validated device models were used in conjuncture with human body numerical simulations within three virtual human models to estimate clinically relevant heating.ResultsThe wound segments in PIPO devices functioned as a lumped element rather than a receiving antenna, reducing induced energy/heating as compared to the original PIPO devices under typical straight or loop configurations. Minimizing the lead's external portion can reduce the RF-induced heating by significant factors for all studied cases during ASTM phantom measurements and in human body simulations.ConclusionOur findings show a significant reduction in RF heating by minimizing the external segment, thereby enhancing patient safety during 1.5 T MRI procedures. Although limited to four devices at 1.5 T across two applications, the extent of heating reduction may vary for others. Nonetheless, tightly winding the external segment of PIPO electrodes holds promise for improving device safety under MRI.
引用
收藏
页码:2108 / 2122
页数:15
相关论文
共 50 条
  • [31] RF-induced heating in tissue near bilateral DBS implants during MRI at 1.5 T and 3T: The role of surgical lead management
    Golestanirad, Laleh
    Kirsch, John
    Bonmassar, Giorgio
    Downs, Sean
    Elahi, Behzad
    Martin, Alastair
    Iacono, Maria-Ida
    Angelone, Leonardo M.
    Keil, Boris
    Wald, Lawrence L.
    Pilitsis, Julie
    NEUROIMAGE, 2019, 184 : 566 - 576
  • [32] Numerical evaluation of RF-induced heating for gap and pitch variation of helical stent under MRI
    Ji, Xiaohe
    Zheng, Jianfeng
    Chen, Ji
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1017 - 1018
  • [33] Numerical investigations of MRI RF field induced heating for external fixation devices
    Yan Liu
    Jianxiang Shen
    Wolfgang Kainz
    Songsong Qian
    Wen Wu
    Ji Chen
    BioMedical Engineering OnLine, 12
  • [34] Numerical investigations of MRI RF field induced heating for external fixation devices
    Liu, Yan
    Shen, Jianxiang
    Kainz, Wolfgang
    Qian, Songsong
    Wu, Wen
    Chen, Ji
    BIOMEDICAL ENGINEERING ONLINE, 2013, 12
  • [35] RF-induced Heating Evaluation for Passive Device in Tissue-Reduced Virtual Family Models at 1.5 T
    Xia, Meiqi
    Guo, Ran
    Zheng, Jianfeng
    Chen, Ji
    2022 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY & SIGNAL/POWER INTEGRITY, EMCSI, 2022, : 610 - 613
  • [36] The impact of shimming strategies and scan regions on RF-induced heating near a bone screw under 3T MRI
    Yang, Xiaolin
    Zheng, Jianfeng
    Chen, Ji
    2021 JOINT IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY, SIGNAL & POWER INTEGRITY, AND EMC EUROPE (EMC+SIPI AND EMC EUROPE), 2021, : 703 - 706
  • [37] Evaluation of Patient Positioning to Mitigate RF-induced Heating of Cardiac Implantable Electronic Devices for Pediatric MRI Exams
    Martinez, Jessica A.
    Cork, Tyler E.
    Chubb, Henry
    Vasanawala, Shreyas
    Ennis, Daniel B.
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 5027 - 5030
  • [38] MRI RF-Induced Heating Prediction of Complex-Shaped Passive Implantable Medical Devices Using Mesh-Based Convolutional Neural Network
    Chang, Jiajun
    Lan, Qianlong
    Guo, Ran
    Zheng, Jianfeng
    Romero, Rebeca
    Kainz, Wolfgang
    Long, Stuart A. A.
    Chen, Ji
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2023, 71 (05) : 2207 - 2214
  • [39] RF Heating Study of a New Medical Implant Lead for 1.5 T, 3 T, and 7 T MRI Systems
    Das, Rupam
    Yoo, Hyoungsuk
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2017, 59 (02) : 360 - 366
  • [40] MRI Induced Heating for Fully Implanted, Partially Implanted and Minimum Implanted Medical Electrode Leads
    Zeng, Qi
    Zheng, Jianfeng
    Chen, Ji
    2015 31st International Review of Progress in Applied Computational Electromagnetics (ACES) Vol 31, 2015,