RF-induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T

被引:0
|
作者
Islam, Md Zahidul [1 ]
Guo, Ran [1 ]
Akter, Mir Khadiza [1 ]
Zheng, Jianfeng [1 ]
Kainz, Wolfgang [2 ]
Long, Stuart [1 ]
Chen, Ji [1 ]
机构
[1] Univ Houston, Dept Elect & Comp Engn, Houston, TX 77204 USA
[2] High Performance Comp MRI Safety, Jasper, GA USA
基金
美国国家科学基金会;
关键词
magnetic resonance imaging (MRI); MR safety; partially in and partially out (PIPO) medical implants; radio frequency (RF) heating mitigation; trajectory modification; transfer function (TF); SAFETY; LEAD; TRANSMISSION; ELECTRODES; FIELDS; MODEL;
D O I
10.1002/mrm.30389
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo address the issue of RF-induced heating for partially in and partially out (PIPO) medical devices during 1.5 T MRI scans by proposing a method of minimizing the external portion.MethodsA method of tightly winding the external segment of the PIPO device is proposed to minimize the overall device effective reception length during MRI scans to mitigate the RF-induced heating. Two commercially available PIPO medical devices and simplified solid wires were used to demonstrate the concept. RF heating results are compared between typical and minimized-length trajectories under the American Society for Testing and Materials (ASTM) testing procedure. In addition, 16 scaled and validated device models were used in conjuncture with human body numerical simulations within three virtual human models to estimate clinically relevant heating.ResultsThe wound segments in PIPO devices functioned as a lumped element rather than a receiving antenna, reducing induced energy/heating as compared to the original PIPO devices under typical straight or loop configurations. Minimizing the lead's external portion can reduce the RF-induced heating by significant factors for all studied cases during ASTM phantom measurements and in human body simulations.ConclusionOur findings show a significant reduction in RF heating by minimizing the external segment, thereby enhancing patient safety during 1.5 T MRI procedures. Although limited to four devices at 1.5 T across two applications, the extent of heating reduction may vary for others. Nonetheless, tightly winding the external segment of PIPO electrodes holds promise for improving device safety under MRI.
引用
收藏
页码:2108 / 2122
页数:15
相关论文
共 50 条
  • [41] Evaluation of the RF-induced lead-tip heating of AIMDs using a Volume-Weighed Tissue-Cluster Model for 1.5T MRI
    Guo, Ran
    Zheng, Jianfeng
    Xia, Meiqi
    Jiang, Guangqiang
    Shrivastava, Devashish
    Kainz, Wolfgang
    Chen, Ji
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 1527 - 1530
  • [42] MRI Induced Heating for Fully Implanted, Partially Implanted and Minimally Implanted Medical Electrode Leads
    Zeng, Qi
    Wang, Qingyan
    Zheng, Jianfeng
    Chen, Ji
    PROCEEDINGS OF THE 2015 INTERNATIONAL CONFERENCE ON ELECTROMAGNETICS IN ADVANCED APPLICATIONS (ICEAA), 2015, : 1590 - 1591
  • [43] Comparison Study of RF-induced Heating in Leg Phantom with Circular External Fixator for TEM and Birdcage Coils at 3 T
    Yang, Rui
    Zheng, Jianfeng
    Chen, Ji
    Kainz, Wolfgang
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY & SIGNAL/POWER INTEGRITY (EMCSI), 2017,
  • [44] A comparative study of numerical and experimental evaluation of RF-induced heating for an endovascular stent-graft at 1.5T and 3T
    Camps-Raga, Bruno
    Goertz, Wolfgang
    Schaefers, Gregor
    Mezape, Yoav
    Shalev, Alon
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2012, 57 : 27 - 27
  • [45] RF Injection Network Development for Testing of Active Implantable Medical Devices Exposed to RF Fields in 1.5 T MRI Systems
    Attaran, Ali
    Handler, William B.
    Chronik, Blaine A.
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2020, 4 (01): : 2 - 9
  • [46] Study on Search Strategies for Assessing the Worst Case RF-Induced Heating for Multi-Configuration Implant System Under MRI
    Zheng, Jianfeng
    Ji, Xiaohe
    Kainz, Wolfgang
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2020, 62 (01) : 43 - 51
  • [47] Modifying surgical implantation of deep brain stimulation leads significantly reduces RF-induced heating during 3 T MRI
    Vu, Jasmine
    Bhusal, Bhumi
    Rosenow, Joshua
    Pilitsis, Julie
    Golestanirad, Laleh
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 4978 - 4981
  • [48] Fast Prediction of RF-induced Heating for Sacral Neuromodulation System Exposed to Multi-Channel 2 RF Field at 3T MRI
    Lan, Qianlong
    Guo, Ran
    Chang, Jiajun
    Zheng, Jianfeng
    Yu, Kyle
    Chen, Ji
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 4159 - 4162
  • [49] Impact of MRI RF coil design on the RF-induced heating of medical implants: fixed B1+ rms exposure versus normal operating mode
    Yao, Aiping
    Li, Zihan
    Ma, Mingjuan
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (05):
  • [50] A simple apparatus for safety assessment of magnetically induced torque on active implantable medical devices (AIMDs) under 1.5 T and 3.0 T MRI
    Hyo-Chang Seo
    Yongkwan Lee
    Segyeong Joo
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2021, 34 : 767 - 774