RF-induced heating of interventional devices at 23.66 MHz

被引:5
|
作者
Oezen, Ali Caglar [1 ]
Russe, Maximilian Frederik [2 ]
Lottner, Thomas [1 ]
Reiss, Simon [1 ]
Littin, Sebastian [1 ]
Zaitsev, Maxim [1 ]
Bock, Michael [1 ]
机构
[1] Univ Freiburg, Univ Med Ctr Freiburg, Fac Med, Dept Diag & Intervent Radiol,Div Med Phys, Freiburg, Germany
[2] Univ Freiburg, Univ Med Ctr Freiburg, Fac Med, Dept Diag & Intervent Radiol, Freiburg, Germany
关键词
Magnetic resonance imaging; RF-induced heating; MR safety; Interventional MRI; MR-guided intervention; Cardiovascular catheterization; Hepatic artery catheterization; Low field MRI; GUIDED CORONARY CATHETERIZATION; MAGNETIC-RESONANCE-GUIDANCE; ENDOVASCULAR STENTS; MRI; FEASIBILITY; DESIGN; GUIDEWIRES; EXCITATION; TRACKING; ARTERY;
D O I
10.1007/s10334-023-01099-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveLow-field MRI systems are expected to cause less RF heating in conventional interventional devices due to lower Larmor frequency. We systematically evaluate RF-induced heating of commonly used intravascular devices at the Larmor frequency of a 0.55 T system (23.66 MHz) with a focus on the effect of patient size, target organ, and device position on maximum temperature rise.Materials and methodsTo assess RF-induced heating, high-resolution measurements of the electric field, temperature, and transfer function were combined. Realistic device trajectories were derived from vascular models to evaluate the variation of the temperature increase as a function of the device trajectory. At a low-field RF test bench, the effects of patient size and positioning, target organ (liver and heart) and body coil type were measured for six commonly used interventional devices (two guidewires, two catheters, an applicator and a biopsy needle).ResultsElectric field mapping shows that the hotspots are not necessarily localized at the device tip. Of all procedures, the liver catheterizations showed the lowest heating, and a modification of the transmit body coil could further reduce the temperature increase. For common commercial needles no significant heating was measured at the needle tip. Comparable local SAR values were found in the temperature measurements and the TF-based calculations.ConclusionAt low fields, interventions with shorter insertion lengths such as hepatic catheterizations result in less RF-induced heating than coronary interventions. The maximum temperature increase depends on body coil design.
引用
收藏
页码:439 / 449
页数:11
相关论文
共 50 条
  • [1] RF-induced heating of interventional devices at 23.66 MHz
    Ali Caglar Özen
    Maximilian Frederik Russe
    Thomas Lottner
    Simon Reiss
    Sebastian Littin
    Maxim Zaitsev
    Michael Bock
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2023, 36 : 439 - 449
  • [2] RF-induced ankle heating in a rhesus monkey at 100 MHz
    Olsen, RG
    Van Matre, BJ
    Lords, JL
    ELECTRICITY AND MAGNETISM IN BIOLOGY AND MEDICINE, 1999, : 307 - 310
  • [3] A counterpoise design for RF-induced heating reduction
    Wang, Yu
    Zheng, Jianfeng
    Wang, Qingyan
    Chen, Ji
    42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20, 2020, : 4200 - 4203
  • [4] A technique for the reduction of RF-induced heating of active implantable medical devices during MRI
    Wang, Yu
    Zheng, Jianfeng
    Guo, Ran
    Wang, Qingyan
    Kainz, Wolfgang
    Long, Stuart
    Chen, Ji
    MAGNETIC RESONANCE IN MEDICINE, 2022, 87 (01) : 349 - 364
  • [5] rf-induced heating dynamics of noncrystallized trapped ions
    van Mourik, Martin W.
    Hrmo, Pavel
    Gerster, Lukas
    Wilhelm, Benjamin
    Blatt, Rainer
    Schindler, Philipp
    Monz, Thomas
    PHYSICAL REVIEW A, 2022, 105 (03)
  • [6] Fast Prediction of MRI RF-induced Heating For Implantable Plate Devices Using Neural Network
    Zheng, Jianfeng
    Yang, Rui
    Chen, Ji
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1011 - 1012
  • [7] Preliminary Study of RF-Induced Heating of DBS and RNS Devices under 1.5 T MRI
    Akter, Mir Khadiza
    Guo, Ran
    Islam, Md Zahidul
    Zheng, Jianfeng
    Chen, Ji
    2024 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION AND INC/USNCURSI RADIO SCIENCE MEETING, AP-S/INC-USNC-URSI 2024, 2024, : 2655 - 2656
  • [8] MR safety: Fast T1 thermometry of the RF-induced heating of medical devices
    Gensler, D.
    Fidler, F.
    Ehses, P.
    Warmuth, M.
    Reiter, T.
    Duering, M.
    Ritter, O.
    Ladd, M. E.
    Quick, H. H.
    Jakob, P. M.
    Bauer, W. R.
    Nordbeck, P.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (05) : 1593 - 1599
  • [9] Effects of Strain Relief Loop on MRI RF-Induced Heating of Active Implantable Medical Devices
    Akter, Mir Khadiza
    Guo, Ran
    Islam, Md Zahidul
    Zheng, Jianfeng
    Chen, Ji
    PROCEEDINGS OF THE 2022 IEEE TEXAS SYMPOSIUM ON WIRELESS AND MICROWAVE CIRCUITS AND SYSTEMS (WMCS), 2022,
  • [10] Lead Electromagnetic Model to Evaluate RF-Induced Heating of a Coax Lead: A Numerical Case Study at 128 MHz
    Kozlov, Mikhail
    Kainz, Wolfgang
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2018, 2 (04): : 286 - 293