RF-induced heating of interventional devices at 23.66 MHz

被引:5
|
作者
Oezen, Ali Caglar [1 ]
Russe, Maximilian Frederik [2 ]
Lottner, Thomas [1 ]
Reiss, Simon [1 ]
Littin, Sebastian [1 ]
Zaitsev, Maxim [1 ]
Bock, Michael [1 ]
机构
[1] Univ Freiburg, Univ Med Ctr Freiburg, Fac Med, Dept Diag & Intervent Radiol,Div Med Phys, Freiburg, Germany
[2] Univ Freiburg, Univ Med Ctr Freiburg, Fac Med, Dept Diag & Intervent Radiol, Freiburg, Germany
关键词
Magnetic resonance imaging; RF-induced heating; MR safety; Interventional MRI; MR-guided intervention; Cardiovascular catheterization; Hepatic artery catheterization; Low field MRI; GUIDED CORONARY CATHETERIZATION; MAGNETIC-RESONANCE-GUIDANCE; ENDOVASCULAR STENTS; MRI; FEASIBILITY; DESIGN; GUIDEWIRES; EXCITATION; TRACKING; ARTERY;
D O I
10.1007/s10334-023-01099-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveLow-field MRI systems are expected to cause less RF heating in conventional interventional devices due to lower Larmor frequency. We systematically evaluate RF-induced heating of commonly used intravascular devices at the Larmor frequency of a 0.55 T system (23.66 MHz) with a focus on the effect of patient size, target organ, and device position on maximum temperature rise.Materials and methodsTo assess RF-induced heating, high-resolution measurements of the electric field, temperature, and transfer function were combined. Realistic device trajectories were derived from vascular models to evaluate the variation of the temperature increase as a function of the device trajectory. At a low-field RF test bench, the effects of patient size and positioning, target organ (liver and heart) and body coil type were measured for six commonly used interventional devices (two guidewires, two catheters, an applicator and a biopsy needle).ResultsElectric field mapping shows that the hotspots are not necessarily localized at the device tip. Of all procedures, the liver catheterizations showed the lowest heating, and a modification of the transmit body coil could further reduce the temperature increase. For common commercial needles no significant heating was measured at the needle tip. Comparable local SAR values were found in the temperature measurements and the TF-based calculations.ConclusionAt low fields, interventions with shorter insertion lengths such as hepatic catheterizations result in less RF-induced heating than coronary interventions. The maximum temperature increase depends on body coil design.
引用
收藏
页码:439 / 449
页数:11
相关论文
共 50 条
  • [31] Reducing MRI RF-induced heating for the external fixation using capacitive structures
    Zheng, Jianfeng
    Yang, Rui
    Wang, Qingyan
    Guo, Ran
    Xu, Jian
    Shou, Xingxian
    Kainz, Wolfgang
    Chen, Ji
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (15):
  • [32] A Novel Design of Implantable Medical Stent for Reducing the MRI RF-Induced Heating
    Li, Dawei
    Ji, Xiaohe
    Zheng, Jianfeng
    Chen, Ji
    Pan, Changwang
    2016 IEEE ANTENNAS AND PROPAGATION SOCIETY INTERNATIONAL SYMPOSIUM, 2016, : 1945 - 1946
  • [33] Influence of a Metallic Shield on RF-Induced Heating of a Lead with Straight and Helical Wires
    Kozlov, Mikhail
    Kainz, Wolfgang
    Daniel, Luca
    2019 IEEE MTT-S INTERNATIONAL CONFERENCE ON NUMERICAL ELECTROMAGNETIC AND MULTIPHYSICS MODELING AND OPTIMIZATION (NEMO 2019), 2019,
  • [34] Spatial distribution of RF-induced E-fields and implant heating in MRI
    Nordbeck, Peter
    Fidler, Florian
    Weiss, Ingo
    Warmuth, Marcus
    Friedrich, Michael T.
    Ehses, Philipp
    Geistert, Wolfgang
    Ritter, Oliver
    Jakob, Peter M.
    Ladd, Mark E.
    Quick, Harald H.
    Bauer, Wolfgang R.
    MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (02) : 312 - 319
  • [35] RF-induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T
    Islam, Md Zahidul
    Guo, Ran
    Akter, Mir Khadiza
    Zheng, Jianfeng
    Kainz, Wolfgang
    Long, Stuart
    Chen, Ji
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (05) : 2108 - 2122
  • [36] Effect of Strain Relief Loop Position on the RF-Induced Heating of Active Implantable Medical Devices at 1.5-T MRI
    Akter, Mir Khadiza
    Guo, Ran
    Islam, Md Zahidul
    Zheng, Jianfeng
    Kainz, Wolfgang
    Long, Stuart A.
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2024, 66 (04) : 1041 - 1056
  • [37] PROPAGATION OF RF-INDUCED DENSITY DEPRESSIONS
    HANSEN, K
    PIEL, A
    BONHOMME, G
    PHYSICA SCRIPTA, 1995, 52 (03) : 322 - 329
  • [38] RF-INDUCED SIDEBANDS IN MOSSBAUER SPECTRA
    HEIMAN, ND
    PFEIFFER, L
    WALKER, JC
    PHYSICAL REVIEW LETTERS, 1968, 21 (02) : 93 - &
  • [39] ADVANCES IN RF-INDUCED PLASMA SPUTTERING
    KLOSS, F
    HERTE, L
    SEMICONDUCTOR PRODUCTS AND SOLID STATE TECHNOLOGY, 1967, 10 (12): : 45 - &
  • [40] RF-induced Heating Comparison between TEM and Birdcage Coils for Circular External Fixator
    Yang, Rui
    Zheng, Jianfeng
    Huang, Xin
    Chen, Ji
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1013 - 1014