Influence of a Metallic Shield on RF-Induced Heating of a Lead with Straight and Helical Wires

被引:0
|
作者
Kozlov, Mikhail [1 ]
Kainz, Wolfgang [2 ]
Daniel, Luca [3 ]
机构
[1] Max Plank Inst Human Cognit & Brain Sci, Leipzig, Germany
[2] US FDA, Silver Spring, MD USA
[3] MIT, 77 Massachusetts Ave, Cambridge, MA 02139 USA
关键词
electromagnetics; RF safety; heating; active implantable medical devices; transfer functions;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We present results including a lead electromagnetic model (LEM), power deposition, and RF-induced heating for a set of single electrode leads with straight and helical wires, as regular unshielded lead, and leads with a metallic shield. Our results were obtained numerically, employing 3D electromagnetic and thermal co-simulations at 128 MHz. We observe that the metallic shield raises the LEM calibration factor and thereby it increases the RF-induced heating of the lead with helical wires, a circumstance that previously had mostly been neglected.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] MRI safety: RF-induced heating near straight wires
    Park, SM
    Kamondetdacha, R
    Amjad, A
    Nyenhuis, JA
    [J]. IEEE TRANSACTIONS ON MAGNETICS, 2005, 41 (10) : 4197 - 4199
  • [2] Influence of Metallic Shielding on Radio Frequency Energy-Induced Heating of Leads With Straight and Helical Wires: A Numerical Case Study
    Kozlov, Mikhail
    Kainz, Wolfgang
    Daniel, Luca
    [J]. IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2020, 68 (02) : 509 - 515
  • [3] RF-induced heating of metallic implants simulated as PEC: Is there something missing?
    Zilberti, Luca
    Zanovello, Umberto
    Arduino, Alessandro
    Bottauscio, Oriano
    Chiampi, Mario
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2021, 85 (02) : 583 - 586
  • [4] MRI thermometry: Fast mapping of RF-induced heating along conductive wires
    Ehses, Philipp
    Fidler, Florian
    Nordbeck, Peter
    Pracht, Eberhard D.
    Warmuth, Marcus
    Jakob, Peter M.
    Bauer, Wolfgang R.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (02) : 457 - 461
  • [5] Modeling RF-Induced Power Deposition and Temperature Rise of Coaxial Leads with Helical Wires
    Kozlov, Mikhail
    Horner, Marc
    Kainz, Wolfgang
    [J]. 2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 1895 - 1898
  • [6] Effect of electrode geometry on the lead electromagnetic model and RF-induced heating
    Kozlov, Mikhail
    Kainz, W.
    [J]. 2018 IEEE ANTENNAS AND PROPAGATION SOCIETY INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2018, : 899 - 900
  • [7] RF-induced temperature elevation along metallic wires in clinical magnetic resonance imaging: Influence of diameter and length
    Armenean, C
    Perrin, E
    Armenean, M
    Beuf, O
    Pilleul, F
    Saint-Jalmes, H
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (05) : 1200 - 1206
  • [8] A counterpoise design for RF-induced heating reduction
    Wang, Yu
    Zheng, Jianfeng
    Wang, Qingyan
    Chen, Ji
    [J]. 42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20, 2020, : 4200 - 4203
  • [9] Numerical evaluation of RF-induced heating for gap and pitch variation of helical stent under MRI
    Ji, Xiaohe
    Zheng, Jianfeng
    Chen, Ji
    [J]. 2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1017 - 1018
  • [10] Thermo-Acoustic Ultrasound for Detection of RF-Induced Device Lead Heating in MRI
    Dixit, Neerav
    Stang, Pascal P.
    Pauly, John M.
    Scott, Greig C.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (02) : 536 - 546