RF-induced heating of interventional devices at 23.66 MHz

被引:5
|
作者
Oezen, Ali Caglar [1 ]
Russe, Maximilian Frederik [2 ]
Lottner, Thomas [1 ]
Reiss, Simon [1 ]
Littin, Sebastian [1 ]
Zaitsev, Maxim [1 ]
Bock, Michael [1 ]
机构
[1] Univ Freiburg, Univ Med Ctr Freiburg, Fac Med, Dept Diag & Intervent Radiol,Div Med Phys, Freiburg, Germany
[2] Univ Freiburg, Univ Med Ctr Freiburg, Fac Med, Dept Diag & Intervent Radiol, Freiburg, Germany
关键词
Magnetic resonance imaging; RF-induced heating; MR safety; Interventional MRI; MR-guided intervention; Cardiovascular catheterization; Hepatic artery catheterization; Low field MRI; GUIDED CORONARY CATHETERIZATION; MAGNETIC-RESONANCE-GUIDANCE; ENDOVASCULAR STENTS; MRI; FEASIBILITY; DESIGN; GUIDEWIRES; EXCITATION; TRACKING; ARTERY;
D O I
10.1007/s10334-023-01099-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveLow-field MRI systems are expected to cause less RF heating in conventional interventional devices due to lower Larmor frequency. We systematically evaluate RF-induced heating of commonly used intravascular devices at the Larmor frequency of a 0.55 T system (23.66 MHz) with a focus on the effect of patient size, target organ, and device position on maximum temperature rise.Materials and methodsTo assess RF-induced heating, high-resolution measurements of the electric field, temperature, and transfer function were combined. Realistic device trajectories were derived from vascular models to evaluate the variation of the temperature increase as a function of the device trajectory. At a low-field RF test bench, the effects of patient size and positioning, target organ (liver and heart) and body coil type were measured for six commonly used interventional devices (two guidewires, two catheters, an applicator and a biopsy needle).ResultsElectric field mapping shows that the hotspots are not necessarily localized at the device tip. Of all procedures, the liver catheterizations showed the lowest heating, and a modification of the transmit body coil could further reduce the temperature increase. For common commercial needles no significant heating was measured at the needle tip. Comparable local SAR values were found in the temperature measurements and the TF-based calculations.ConclusionAt low fields, interventions with shorter insertion lengths such as hepatic catheterizations result in less RF-induced heating than coronary interventions. The maximum temperature increase depends on body coil design.
引用
收藏
页码:439 / 449
页数:11
相关论文
共 50 条
  • [41] Thermo-Acoustic Ultrasound for Detection of RF-Induced Device Lead Heating in MRI
    Dixit, Neerav
    Stang, Pascal P.
    Pauly, John M.
    Scott, Greig C.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (02) : 536 - 546
  • [42] Variability in RF-induced heating of a deep brain stimulation implant across MR systems
    Baker, Kenneth B.
    Tkach, Jean A.
    Phillips, Micheal D.
    Rezai, Ali R.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 24 (06) : 1236 - 1242
  • [43] Impact of RF Shimming on RF-Induced Heating Near Implantable Medical Electrodes in a 3T MRI Coil
    Zeng, Qi
    Wang, Qingyan
    Kainz, Wolfgang
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2020, 62 (01) : 52 - 64
  • [44] RF-INDUCED NEGATIVE RESISTANCE IN JUNCTION DIODES
    MCDADE, JC
    PROCEEDINGS OF THE INSTITUTE OF RADIO ENGINEERS, 1961, 49 (05): : 957 - &
  • [45] Numerical Analysis of RF-Induced Heating While Wearing Face Mask at Magnetic Resonance Imaging
    Cho, Youngdae
    Yoo, Hyoungsuk
    IEEE ACCESS, 2022, 10 : 60946 - 60954
  • [46] RF-Induced Heating of Spinal Cord Stimulator under Different Implantations and Body Loading Orientations
    Chang, Jiajun
    Guo, Ran
    Wang, Zhichao
    Zheng, Jianfeng
    Chen, Ji
    Hoegh, Thomas
    2020 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION AND NORTH AMERICAN RADIO SCIENCE MEETING, 2020, : 1393 - 1394
  • [47] Electromagnetic Modeling of the Implantable Electrode for Transfer Function Calibration in MRI RF-Induced Heating Assessment
    Long, Tiangang
    Jiang, Changqing
    Li, Luming
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2024, 8 (03): : 259 - 264
  • [48] Impacts of Tip Structure on RF-induced Heating of an Implantable Neurostimulator under 1.5 T MRI
    Yang, Rui
    Zheng, Jianfeng
    Chen, Ji
    2018 IEEE ANTENNAS AND PROPAGATION SOCIETY INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2018, : 2383 - 2384
  • [49] Fast prediction of MRI RF-induced heating for a generic stent with arbitrary orientation using ANN
    Ji, Xiaohe
    Zheng, Jianfeng
    Chen, Ji
    2019 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION AND USNC-URSI RADIO SCIENCE MEETING, 2019, : 527 - 528
  • [50] A Cascaded Heterogeneous Equivalent Network for Evaluating RF-Induced Hazards on Active Implantable Medical Devices
    Liu, Jingshen
    Wang, Yu
    Guo, Ran
    Wang, Qingyan
    Zheng, Jianfeng
    Kurpad, Krishna
    Kainz, Wolfgang
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2022, 64 (02) : 286 - 294