Variability in RF-induced heating of a deep brain stimulation implant across MR systems

被引:70
|
作者
Baker, Kenneth B.
Tkach, Jean A.
Phillips, Micheal D.
Rezai, Ali R.
机构
[1] Cleveland Clin Fdn, Ctr Neurol Restorat, Cleveland, OH 44195 USA
[2] Cleveland Clin Fdn, Dept Radiol, Cleveland, OH 44195 USA
[3] Case Western Reserve Univ, Dept Radiol, Cleveland, OH 44106 USA
关键词
magnetic resonance imaging; bioeffects; safety; neurostimulation systems; implants; deep brain stimulation; specific absorption rate;
D O I
10.1002/jmri.20769
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the MRI-related heating per unit of specific absorption rate (SAR) profile of a conductive implant between two 1.5-Tesla/64 MHz MR systems using a transmit/receive (t/r) head coil configuration. Materials and Methods: Deep brain stimulation (DBS) leads were configured within a gel-filled phantom of the human head and torso. Temperature variation at each of four contacts of the bilaterally-placed leads was monitored using fluoroptic thermometry. MRI was performed using the t/r head coils of two different-generation 1.5-Tesla MR systems from the same manufacturer. Temperature changes were normalized to SAR values for the head (Delta T/SAR-H), and the slope of this Delta T/SAR-H by time relationship was compared between the two scanners. Results: The Delta T/SAR-H for the implant ranged from 3.5 to 5.5 times higher on one MR system as compared to the other (P < 0.01) depending on the measurement site. Conclusion: The findings support previous observations that console-reported SAR does not constitute a reliable index of heating for elongated, conductive implants, such as the DBS hardware system tested. In contrast to our previous findings using a t/r body coil, the data presented here reveal marked differences between two MR systems using t/r head coils (the coil configuration was consistent with the implant manufacturer's imaging guidelines).
引用
收藏
页码:1236 / 1242
页数:7
相关论文
共 50 条
  • [31] Influence of a Metallic Shield on RF-Induced Heating of a Lead with Straight and Helical Wires
    Kozlov, Mikhail
    Kainz, Wolfgang
    Daniel, Luca
    [J]. 2019 IEEE MTT-S INTERNATIONAL CONFERENCE ON NUMERICAL ELECTROMAGNETIC AND MULTIPHYSICS MODELING AND OPTIMIZATION (NEMO 2019), 2019,
  • [32] MR imaging-related heating of deep brain stimulation electrodes: In vitro study
    Finelli, DA
    Rezai, AR
    Ruggieri, PM
    Tkach, JA
    Nyenhuis, JA
    Hrdlicka, G
    Sharan, A
    Gonzalez-Martinez, J
    Stypulkowski, PH
    Shellock, FG
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2002, 23 (10) : 1795 - 1802
  • [33] MR safety in patients with implanted deep brain stimulation systems (DBS)
    Sharan, A
    Rezai, AR
    Nyenhuis, JA
    Hrdlicka, G
    Tkach, J
    Baker, K
    Turbay, M
    Rugieri, P
    Phillips, M
    Shellock, FG
    [J]. NEUROSURGICAL RE-ENGINEERING OF THE DAMAGED BRAIN AND SPINAL CORD, 2003, 87 : 141 - 145
  • [34] A technique for the reduction of RF-induced heating of active implantable medical devices during MRI
    Wang, Yu
    Zheng, Jianfeng
    Guo, Ran
    Wang, Qingyan
    Kainz, Wolfgang
    Long, Stuart
    Chen, Ji
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2022, 87 (01) : 349 - 364
  • [35] RF-induced Heating Comparison between TEM and Birdcage Coils for Circular External Fixator
    Yang, Rui
    Zheng, Jianfeng
    Huang, Xin
    Chen, Ji
    [J]. 2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1013 - 1014
  • [36] Thermo-Acoustic Ultrasound for Detection of RF-Induced Device Lead Heating in MRI
    Dixit, Neerav
    Stang, Pascal P.
    Pauly, John M.
    Scott, Greig C.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (02) : 536 - 546
  • [37] Impact of RF Shimming on RF-Induced Heating Near Implantable Medical Electrodes in a 3T MRI Coil
    Zeng, Qi
    Wang, Qingyan
    Kainz, Wolfgang
    Chen, Ji
    [J]. IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2020, 62 (01) : 52 - 64
  • [38] RF-Induced Heating of Spinal Cord Stimulator under Different Implantations and Body Loading Orientations
    Chang, Jiajun
    Guo, Ran
    Wang, Zhichao
    Zheng, Jianfeng
    Chen, Ji
    Hoegh, Thomas
    [J]. 2020 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION AND NORTH AMERICAN RADIO SCIENCE MEETING, 2020, : 1393 - 1394
  • [39] Fast Prediction of MRI RF-induced Heating For Implantable Plate Devices Using Neural Network
    Zheng, Jianfeng
    Yang, Rui
    Chen, Ji
    [J]. 2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1011 - 1012
  • [40] Numerical Analysis of RF-Induced Heating While Wearing Face Mask at Magnetic Resonance Imaging
    Cho, Youngdae
    Yoo, Hyoungsuk
    [J]. IEEE ACCESS, 2022, 10 : 60946 - 60954