BLOOD GLUCOSE MEASUREMENT IN PATIENTS WITH SUSPECTED DIABETIC KETOACIDOSIS: A COMPARISON OF ABBOTT MEDISENSE PCX POINT-OF-CARE METER VALUES TO REFERENCE LABORATORY VALUES

被引:6
|
作者
Blank, Fidela S. J. [1 ]
Miller, Moses [2 ]
Nichols, James
Smithline, Howard [2 ]
Crabb, Gillian [2 ]
Pekow, Penelope [3 ,4 ]
机构
[1] Tufts Univ, Sch Med, Boston, MA 02111 USA
[2] Baystate Med Ctr, Dept Emergency Med, Springfield, MA USA
[3] Baystate Med Ctr, Div Healthcare Qual, Springfield, MA USA
[4] Univ Massachusetts, Sch Publ Hlth & Hlth Sci, Amherst, MA 01003 USA
关键词
Diabetic; Ketoacidosis; Blood glucose measurement; Emergency department;
D O I
10.1016/j.jen.2008.01.008
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: The purpose of this study is to compare blood glucose levels measured by a point of care (POC) device to laboratory measurement using the same sample venous blood from patients with suspected diabetic ketoacidosis (DKA). Methods: A descriptive correlational design was used for this IRB-approved quality assurance project. The study site was the 50-bed BMC emergency department (ED) which has an annual census of over 100, 000 patient visits. The convenience sample consisted of 54 blood samples from suspected DKA patients with orders for hourly blood draws for glucose measurement. Spearman correlations of the glucose POC values, reference lab values, and differences between the two, were evaluated. A chi-square test was used to evaluate the association between the acidosis status and FDA acceptability of POC values. Results: Patient age range was 10-86 years; 63% were females; 46% had a final diagnosis of DKA. POC values underestimated glucose levels 93% of the time. There was a high correlation between the lab value and the magnitude of the difference, (lab minus POC value) indicating that the higher the true glucose value, the greater the difference between the lab and the POC value. A chi-square test showed no overall association between acidosis and FDA-acceptability. Discussion: The POC values underestimated lab reported glucose levels in 50 of 54 cases even with the use of same venous sample sent to the lab, which make it highly unreliable for use in monitoring suspected DKA patients.
引用
收藏
页码:93 / 96
页数:4
相关论文
共 50 条
  • [1] COMPARISON OF GLUCOSE POINT-OF-CARE VALUES WITH LABORATORY VALUES IN CRITICALLY ILL PATIENTS
    Shearer, Anna
    Boehmer, Marilyn
    Closs, Melanie
    Dela Rosa, Rosalina
    Hamilton, Jean
    Horton, Karen
    McGrath, Rose
    Schulman, Christine
    [J]. AMERICAN JOURNAL OF CRITICAL CARE, 2009, 18 (03) : 224 - 230
  • [2] Comparison of a Point-of-care Cholesterol Meter With a Reference Laboratory Analyzer in Companion Psittaciformes
    Barboza, Trinita
    Beaufrere, Hugues
    [J]. JOURNAL OF AVIAN MEDICINE AND SURGERY, 2019, 33 (01) : 7 - 14
  • [3] Comparison of lactate values between point-of-care and central laboratory analyzers
    Karon, Brad S.
    Scott, Renee
    Burritt, Mary F.
    Santrach, Paula J.
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2007, 128 (01) : 168 - 171
  • [4] DIFFERENCES IN GLUCOSE VALUES OBTAINED FROM POINT-OF-CARE GLUCOSE METERS AND LABORATORY ANALYSIS IN CRITICALLY ILL PATIENTS
    Cook, Anjannette
    Laughlin, Delyla
    Moore, Margery
    North, Doreen
    Wilkins, Kathleen
    Wong, Gay
    Wallace-Scroggs, Allyson
    Halvorsen, Lisa
    [J]. AMERICAN JOURNAL OF CRITICAL CARE, 2009, 18 (01) : 65 - 71
  • [5] GLUCOSE LEVELS IN CRITICALLY ILL PATIENTS: POINT-OF-CARE (POC) TESTING VERSUS CENTRAL LABORATORY VALUES
    Arce-Delgado, S.
    Bernat-Adell, A.
    Blanch-Eixarch, F.
    Ferrandiz-Selles, A.
    Guillamon-Navarro, S.
    Mateu-Campos, L.
    DeLeon-Belmar, J.
    [J]. INTENSIVE CARE MEDICINE, 2010, 36 : S219 - S219
  • [6] Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method
    Shoko Okahara
    Tetsuya Handoh
    Mitsuru Wakita
    Takamasa Yamamoto
    Shigeki Misawa
    Takashi Miida
    Hiroyuki Sumikura
    [J]. Journal of Anesthesia, 2021, 35 : 757 - 760
  • [7] Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method
    Okahara, Shoko
    Handoh, Tetsuya
    Wakita, Mitsuru
    Yamamoto, Takamasa
    Misawa, Shigeki
    Miida, Takashi
    Sumikura, Hiroyuki
    [J]. JOURNAL OF ANESTHESIA, 2021, 35 (05) : 757 - 760
  • [8] Comparison of prothrombin time values of standard laboratory test and two point-of-care devices
    Urwyler, N.
    Staub, E.
    Staub, L.
    Daetwyler, K.
    Fluri, S.
    Sollberger, T.
    Deplaces, M.
    Theiler, L.
    Greif, R.
    [J]. SWISS MEDICAL WEEKLY, 2008, 138 (41-42) : 5S - 5S
  • [9] Use of point-of-care blood ketone testing to diagnose and monitor patients with diabetic ketoacidosis
    Armer, Jane
    Hunt, Natalie
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 2015, 52 (03) : 413 - 414
  • [10] Comparison of point-of-care and laboratory glucose analysis in critically ill patients
    Lacara, Teresita
    Domagtoy, Caroline
    Lickliter, Donna
    Quattrocchi, Kathy
    Snipes, Lydia
    Kuszaj, Joanne
    Prasnikar, MaryClare
    [J]. AMERICAN JOURNAL OF CRITICAL CARE, 2007, 16 (04) : 336 - 346