BLOOD GLUCOSE MEASUREMENT IN PATIENTS WITH SUSPECTED DIABETIC KETOACIDOSIS: A COMPARISON OF ABBOTT MEDISENSE PCX POINT-OF-CARE METER VALUES TO REFERENCE LABORATORY VALUES

被引:6
|
作者
Blank, Fidela S. J. [1 ]
Miller, Moses [2 ]
Nichols, James
Smithline, Howard [2 ]
Crabb, Gillian [2 ]
Pekow, Penelope [3 ,4 ]
机构
[1] Tufts Univ, Sch Med, Boston, MA 02111 USA
[2] Baystate Med Ctr, Dept Emergency Med, Springfield, MA USA
[3] Baystate Med Ctr, Div Healthcare Qual, Springfield, MA USA
[4] Univ Massachusetts, Sch Publ Hlth & Hlth Sci, Amherst, MA 01003 USA
关键词
Diabetic; Ketoacidosis; Blood glucose measurement; Emergency department;
D O I
10.1016/j.jen.2008.01.008
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: The purpose of this study is to compare blood glucose levels measured by a point of care (POC) device to laboratory measurement using the same sample venous blood from patients with suspected diabetic ketoacidosis (DKA). Methods: A descriptive correlational design was used for this IRB-approved quality assurance project. The study site was the 50-bed BMC emergency department (ED) which has an annual census of over 100, 000 patient visits. The convenience sample consisted of 54 blood samples from suspected DKA patients with orders for hourly blood draws for glucose measurement. Spearman correlations of the glucose POC values, reference lab values, and differences between the two, were evaluated. A chi-square test was used to evaluate the association between the acidosis status and FDA acceptability of POC values. Results: Patient age range was 10-86 years; 63% were females; 46% had a final diagnosis of DKA. POC values underestimated glucose levels 93% of the time. There was a high correlation between the lab value and the magnitude of the difference, (lab minus POC value) indicating that the higher the true glucose value, the greater the difference between the lab and the POC value. A chi-square test showed no overall association between acidosis and FDA-acceptability. Discussion: The POC values underestimated lab reported glucose levels in 50 of 54 cases even with the use of same venous sample sent to the lab, which make it highly unreliable for use in monitoring suspected DKA patients.
引用
收藏
页码:93 / 96
页数:4
相关论文
共 50 条
  • [41] Comparing the accuracy of point-of-care with laboratory (capillary, venous, and arterial) blood glucose levels in critically ill patients with and without shock
    Alshaer, Abdulaziz
    Badgheish, Basma A.
    Alsadah, Zahra Hashim
    Sewify, Khalid
    Alghazal, Sarah
    Alzahrani, Sarah
    Qadi, Abeer
    Alqahtani, Reham
    Farsani, Ghadeer Abdullah
    Shilash, Amal
    [J]. BMC RESEARCH NOTES, 2022, 15 (01)
  • [42] Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: a cross-sectional method comparison study
    Chutipongtanate, Arpa
    Yasaeng, Churairat
    Virankabutra, Tanit
    Chutipongtanate, Somchai
    [J]. BMC ANESTHESIOLOGY, 2020, 20 (01)
  • [43] Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: a cross-sectional method comparison study
    Arpa Chutipongtanate
    Churairat Yasaeng
    Tanit Virankabutra
    Somchai Chutipongtanate
    [J]. BMC Anesthesiology, 20
  • [44] Comparison of arterial and venous blood-gas values in anesthetized Dorset cross-bred lambs (Ovis aries) using a point-of-care analyzer
    Izer, Jenelle
    Mattern, Erin
    Ellwanger, Joy
    Wilson, Ronald
    [J]. VETERINARY ANAESTHESIA AND ANALGESIA, 2019, 46 (02) : 209 - 213
  • [45] Point-of-care Hb measurement in pooled capillary blood by a portable autoanalyser: comparison with venous blood Hb measured by reference methods in cross-sectional and longitudinal studies
    Dasi, Teena
    Palika, Ravindranadh
    Pullakhandam, Raghu
    Augustine, Little Flower
    Boiroju, Naveen Kumar
    Prasannanavar, Devraj J.
    Pradhan, Anju Sinha
    Kurpad, Anura, V
    Sachdev, Harshpal Singh
    Kulkarni, Bharati
    [J]. BRITISH JOURNAL OF NUTRITION, 2022, 128 (06) : 1108 - 1117
  • [46] Comparison of Accu Chek Inform II point-of-care test blood glucose meter with Hexokinase Plasma method for a diabetes mellitus population during surgery under general anesthesia
    Lascaris, B.
    Freling, H. G.
    Edens, M. A.
    Fokkert, M. J.
    Olthof, C. G.
    Slingerland, R. J.
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2022, 36 (02) : 355 - 361
  • [47] Comparison of Accu Chek Inform II point-of-care test blood glucose meter with Hexokinase Plasma method for a diabetes mellitus population during surgery under general anesthesia
    B. Lascaris
    H. G. Freling
    M. A. Edens
    M. J. Fokkert
    C. G. Olthof
    R. J. Slingerland
    [J]. Journal of Clinical Monitoring and Computing, 2022, 36 : 355 - 361
  • [48] Comparison of point-of-care testing (POCT): i-STAT® international normalized ratio (INR) vs reference laboratory INR in pediatric patients undergoing major surgery
    Spielmann, Nelly
    Mauch, Jacqueline Y.
    Madjdpour, Caveh
    Schmugge, Markus
    Albisetti, Manuela
    Weiss, Markus
    Haas, Thorsten
    [J]. PEDIATRIC ANESTHESIA, 2011, 21 (10) : 1041 - 1045
  • [49] A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease
    Chothia, Mogamat-Yazied
    Kassum, Patricia
    Zemlin, Annalise
    [J]. BIOCHEMIA MEDICA, 2020, 30 (03) : 1 - 7
  • [50] Concordance between point-of-care blood gas analysis and laboratory autoanalyzer in measurement of hemoglobin and electrolytes in critically ill patients (vol 32, e22425, 2018)
    Prakash, Shivesh
    Bihari, Shailesh
    Lim, Zhan Y.
    Verghese, Santosh
    Kulkarni, Hemant
    Bersten, Andrew D.
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2019, 33 (03)