BLOOD GLUCOSE MEASUREMENT IN PATIENTS WITH SUSPECTED DIABETIC KETOACIDOSIS: A COMPARISON OF ABBOTT MEDISENSE PCX POINT-OF-CARE METER VALUES TO REFERENCE LABORATORY VALUES

被引:6
|
作者
Blank, Fidela S. J. [1 ]
Miller, Moses [2 ]
Nichols, James
Smithline, Howard [2 ]
Crabb, Gillian [2 ]
Pekow, Penelope [3 ,4 ]
机构
[1] Tufts Univ, Sch Med, Boston, MA 02111 USA
[2] Baystate Med Ctr, Dept Emergency Med, Springfield, MA USA
[3] Baystate Med Ctr, Div Healthcare Qual, Springfield, MA USA
[4] Univ Massachusetts, Sch Publ Hlth & Hlth Sci, Amherst, MA 01003 USA
关键词
Diabetic; Ketoacidosis; Blood glucose measurement; Emergency department;
D O I
10.1016/j.jen.2008.01.008
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: The purpose of this study is to compare blood glucose levels measured by a point of care (POC) device to laboratory measurement using the same sample venous blood from patients with suspected diabetic ketoacidosis (DKA). Methods: A descriptive correlational design was used for this IRB-approved quality assurance project. The study site was the 50-bed BMC emergency department (ED) which has an annual census of over 100, 000 patient visits. The convenience sample consisted of 54 blood samples from suspected DKA patients with orders for hourly blood draws for glucose measurement. Spearman correlations of the glucose POC values, reference lab values, and differences between the two, were evaluated. A chi-square test was used to evaluate the association between the acidosis status and FDA acceptability of POC values. Results: Patient age range was 10-86 years; 63% were females; 46% had a final diagnosis of DKA. POC values underestimated glucose levels 93% of the time. There was a high correlation between the lab value and the magnitude of the difference, (lab minus POC value) indicating that the higher the true glucose value, the greater the difference between the lab and the POC value. A chi-square test showed no overall association between acidosis and FDA-acceptability. Discussion: The POC values underestimated lab reported glucose levels in 50 of 54 cases even with the use of same venous sample sent to the lab, which make it highly unreliable for use in monitoring suspected DKA patients.
引用
收藏
页码:93 / 96
页数:4
相关论文
共 50 条
  • [21] Comparison of Disparities between FGM/CGM Sensor Glucose Values and Blood Glucose Values in Type 1 Diabetic Patients on SAP Therapy
    Tanizawa, Mika
    Takahashi, Hiroshi
    Mori, Yutaka
    Utsunomiya, Kazunori
    [J]. DIABETES, 2018, 67
  • [22] Precision and Accuracy Comparison of Point-of-Care and Laboratory Glucose Concentrations in Cardiothoracic Surgery Patients
    Denfeld, Quin E.
    Goodell, Teresa T.
    Stafford, Kelly N.
    Kazmierczak, Steve
    [J]. JOURNAL OF CARDIOVASCULAR NURSING, 2011, 26 (06) : 512 - 518
  • [23] A Comparison Study Between Point-of-Care Testing Systems and Central Laboratory for Determining Blood Glucose in Venous Blood
    Wei, Huiping
    Lan, Fang
    He, Qitian
    Li, Haiwei
    Zhang, Fuyong
    Qin, Xue
    Li, Shan
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2017, 31 (03)
  • [24] Comparison of Disparities between FGM/CGM Sensor Glucose Values and Blood Glucose Values in Type 2 Diabetic Patients Admitted for Glycemic Control
    Tanizawa, Mika
    Takahashi, Hiroshi
    Mori, Yutaka
    Utsunomiya, Kazunori
    [J]. DIABETES, 2018, 67
  • [25] Comparison of a portage blood glucose meter analyzer with a benchtop point-of-care chemistry analyzer for measurement of blood glucose concentration in client-owned ferrets (Mustela putorius furo)
    Proulx, Marie-Pier
    Vergneau-Grosset, Claire
    Hebert, Julie
    Bedard, Christian
    Maccolini, Edouard
    [J]. JOURNAL OF EXOTIC PET MEDICINE, 2022, 43 : 22 - 28
  • [26] Comparison of a new point-of-care whole blood prothrombin time system to a laboratory reference instrument.
    Spink, BJ
    Schreiner, K
    Buschmann, DJ
    Taborski, U
    [J]. BLOOD, 2000, 96 (11) : 102B - 102B
  • [27] Comparison of a Point-of-Care Whole Blood Beta-Hydroxybutyrate System with a Clinical Laboratory Reference Method
    Chin, A. C.
    Koch, D. D.
    [J]. CLINICAL CHEMISTRY, 2009, 55 (06) : A104 - A104
  • [28] Point-of-care blood glucose measurement errors overestimate hypoglycaemia rates in critically ill patients
    Nya-Ngatchou, Jean-Jacques
    Corl, Dawn
    Onstad, Susan
    Yin, Tom
    Tylee, Tracy
    Suhr, Louise
    Thompson, Rachel E.
    Wisse, Brent E.
    [J]. DIABETES-METABOLISM RESEARCH AND REVIEWS, 2015, 31 (02) : 147 - 154
  • [29] Comparison of the point-of-care blood gas analyzer versus the laboratory auto-analyzer for the measurement of electrolytes
    Jain A.
    Subhan I.
    Joshi M.
    [J]. International Journal of Emergency Medicine, 2009, 2 (2) : 117 - 120
  • [30] A novel point-of-care device for blood potassium detection of patients on dialysis: Comparison with a reference method
    Novell, Marta
    Rico, Nayra
    Blondeau, Pascal
    Blasco, Miquel
    Maceira, Adria
    Bedini, Jose Luis
    Javier Andrade, Francisco
    Maduell, Francesc
    [J]. NEFROLOGIA, 2020, 40 (03): : 363 - 364