COMPARISON OF GLUCOSE POINT-OF-CARE VALUES WITH LABORATORY VALUES IN CRITICALLY ILL PATIENTS

被引:25
|
作者
Shearer, Anna [1 ]
Boehmer, Marilyn [1 ]
Closs, Melanie [1 ]
Dela Rosa, Rosalina [1 ]
Hamilton, Jean [1 ]
Horton, Karen [1 ]
McGrath, Rose [1 ]
Schulman, Christine
机构
[1] Providence St Vincents Med Ctr, Intens Care Unit, Portland, OR 97225 USA
关键词
INTENSIVE INSULIN THERAPY; ARTERIAL-BLOOD; METERS; HEMATOCRIT; CAPILLARY; ACCURACY; REFLECTANCE; AGREEMENT; MONITORS; PH;
D O I
10.4037/ajcc2009448
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background Blood from central venous or arterial catheters as well as from fingersticks is often used for point-of-care glucose testing. Objectives To compare glucose values obtained with a point-of-care device for catheter and fingerstick blood samples with values obtained by clinical laboratory analysis of a catheter blood sample. Methods A method-comparison design was used. Point-of-care values for fingerstick and catheter blood samples were compared with laboratory values for blood from catheters in a convenience sample of 63 critically ill patients. Device bias and precision were calculated and graphed according to the Bland-Altman method, and a t test was used to determine differences in glucose values for the 2 methods. Results Laboratory glucose values for blood from a catheter sample differed significantly from point-of-care values for blood from a fingerstick (t(1,61) = 5.01; P < .001) and from a catheter (t(1,61) = 3.91; P < .001). Bias and precision for the point-of-care device was large (fingerstick, 8.7 and 13.7; catheter, 7.0 and 14.0); 20% of the values differed from the laboratory glucose values by more than 20 mg/dL. Point-of-care glucose values for fingerstick and catheter samples did not differ (P = .41). Conclusions Glucose values obtained with a point-of-care device differ significantly from those obtained by laboratory analysis. The magnitude of these differences calls into question the widespread practice of using point-of-care glucose testing to guide insulin titration for tight glucose control. Errors in dosing could easily be made because of the large bias and precision associated with a point-of-care device. (American Journal of Critical Care. 2009; 18: 224-230)
引用
收藏
页码:224 / 230
页数:7
相关论文
共 50 条
  • [1] DIFFERENCES IN GLUCOSE VALUES OBTAINED FROM POINT-OF-CARE GLUCOSE METERS AND LABORATORY ANALYSIS IN CRITICALLY ILL PATIENTS
    Cook, Anjannette
    Laughlin, Delyla
    Moore, Margery
    North, Doreen
    Wilkins, Kathleen
    Wong, Gay
    Wallace-Scroggs, Allyson
    Halvorsen, Lisa
    [J]. AMERICAN JOURNAL OF CRITICAL CARE, 2009, 18 (01) : 65 - 71
  • [2] Comparison of point-of-care and laboratory glucose analysis in critically ill patients
    Lacara, Teresita
    Domagtoy, Caroline
    Lickliter, Donna
    Quattrocchi, Kathy
    Snipes, Lydia
    Kuszaj, Joanne
    Prasnikar, MaryClare
    [J]. AMERICAN JOURNAL OF CRITICAL CARE, 2007, 16 (04) : 336 - 346
  • [3] GLUCOSE LEVELS IN CRITICALLY ILL PATIENTS: POINT-OF-CARE (POC) TESTING VERSUS CENTRAL LABORATORY VALUES
    Arce-Delgado, S.
    Bernat-Adell, A.
    Blanch-Eixarch, F.
    Ferrandiz-Selles, A.
    Guillamon-Navarro, S.
    Mateu-Campos, L.
    DeLeon-Belmar, J.
    [J]. INTENSIVE CARE MEDICINE, 2010, 36 : S219 - S219
  • [4] POINT-OF-CARE GLUCOSE TESTING IN CRITICALLY ILL PATIENTS
    Scimio, Kylie
    Carter, Chris
    El-Oshar, Seraj
    [J]. CRITICAL CARE MEDICINE, 2018, 46 (01) : 202 - 202
  • [5] BLOOD GLUCOSE MEASUREMENT IN PATIENTS WITH SUSPECTED DIABETIC KETOACIDOSIS: A COMPARISON OF ABBOTT MEDISENSE PCX POINT-OF-CARE METER VALUES TO REFERENCE LABORATORY VALUES
    Blank, Fidela S. J.
    Miller, Moses
    Nichols, James
    Smithline, Howard
    Crabb, Gillian
    Pekow, Penelope
    [J]. JOURNAL OF EMERGENCY NURSING, 2009, 35 (02) : 93 - 96
  • [6] Clinical Reliability of Point-of-Care Glucose Testing in Critically Ill Patients
    Pilackas, Kylie
    El-Oshar, Seraj
    Carter, Chris
    [J]. JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2020, 14 (01): : 65 - 69
  • [7] Comparison of lactate values between point-of-care and central laboratory analyzers
    Karon, Brad S.
    Scott, Renee
    Burritt, Mary F.
    Santrach, Paula J.
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2007, 128 (01) : 168 - 171
  • [8] Point-of-Care Testing in Critically Ill Patients
    Fries, Dietmar
    Streif, Werner
    [J]. SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2015, 41 (01): : 75 - 83
  • [9] Comparison of blood glucose values and variability using three glucose protocols in critically ill patients
    Hopkins, Ramona O.
    Orme, James F., Jr.
    Suchyta, Mary R.
    Jephson, Al
    Schoenfeld, David A.
    Morris, Alan H.
    [J]. CHEST, 2006, 130 (04) : 205S - 205S
  • [10] Point-of-Care Lung Ultrasound in Critically ill Patients
    Shrestha, Gentle S.
    Weeratunga, Dameera
    Baker, Kylie
    [J]. REVIEWS ON RECENT CLINICAL TRIALS, 2018, 13 (01) : 15 - 26