Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method

被引:6
|
作者
Okahara, Shoko [1 ]
Handoh, Tetsuya [2 ]
Wakita, Mitsuru [2 ]
Yamamoto, Takamasa [2 ]
Misawa, Shigeki [2 ]
Miida, Takashi [2 ]
Sumikura, Hiroyuki [1 ]
机构
[1] Juntendo Univ Hosp, Dept Anesthesiol, Bunkyo Ku, 3-1-3 Hongo, Tokyo 1138431, Japan
[2] Juntendo Univ Hosp, Dept Clin Lab, Bunkyo Ku, 3-1-3 Hongo, Tokyo 1138431, Japan
关键词
Clauss method; FibCare; Fibrinogen; Hemorrhage; Point-of-care testing; TRAUMA; PLASMA;
D O I
10.1007/s00540-021-02987-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Timely fibrinogen replacement is key to treating critical hemorrhage. Measuring fibrinogen concentration by conventional laboratory tests requires centrifugation of blood samples and is often time-consuming. A point-of-care testing device (A&T, Yokohama, Japan), CG02N, has been available in Japan since 2011 to measure fibrinogen concentration without centrifugation. However, it has not been widely used as it requires dilution of blood samples using manual micropipetting. To further speed up and simplify the fibrinogen measurement, an improved device called FibCare (Atom Medical, Tokyo, Japan) was developed to avoid diluting blood samples. The purpose of this study is to verify the reliability of FibCare against laboratory measurement using the Clauss method. Fibrinogen concentrations with 60 sodium citrated whole blood samples were measured by both FibCare and Clauss methods in the laboratory. Measured values with the Clauss method were distributed in the 88-300 mg/dL range. By comparing these results, a significant positive correlation was observed between the FibCare and Clauss method (Y = 12.402 + 0.982 X; R = 0.891; P < 0.01). The study indicated that FibCare allows accurate measurement of fibrinogen concentration and shows a possibility to contribute to optimal fibrinogen replacement therapy during critical hemorrhage.
引用
收藏
页码:757 / 760
页数:4
相关论文
共 50 条
  • [1] Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method
    Shoko Okahara
    Tetsuya Handoh
    Mitsuru Wakita
    Takamasa Yamamoto
    Shigeki Misawa
    Takashi Miida
    Hiroyuki Sumikura
    Journal of Anesthesia, 2021, 35 : 757 - 760
  • [2] A novel point-of-care device for blood potassium detection of patients on dialysis: Comparison with a reference method
    Novell, Marta
    Rico, Nayra
    Blondeau, Pascal
    Blasco, Miquel
    Maceira, Adria
    Bedini, Jose Luis
    Javier Andrade, Francisco
    Maduell, Francesc
    NEFROLOGIA, 2020, 40 (03): : 363 - 364
  • [3] A point-of-care microfluidic channel-based device for rapid and direct detection of fibrinogen in whole blood
    Ban, Qinan
    Zhang, Yulong
    Li, Yuxuan
    Cao, Daye
    Ye, Weifeng
    Zhan, Linsheng
    Wang, Daming
    Wang, Xiaohui
    LAB ON A CHIP, 2022, 22 (14) : 2714 - 2725
  • [4] A novel point-of-care device accurately measures thyrotropin in whole blood, capillary blood and serum
    Kahaly, George J.
    Lotz, Johannes
    Walder, Sara
    Hammad, Cara
    Kraemer, Rebecca
    Frommer, Lara
    Koenig, Jochem
    Wolf, Jan
    Gottwald-Hostalek, Ulrike
    Urgatz, Bogumila
    Lackner, Karl J.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2022, 60 (10) : 1607 - 1616
  • [5] Novel point of care testing device for rapid measurement of clozapine in whole blood
    Sumanth, A.
    Kozo, D.
    Courtney, J.
    Siegfried, N.
    Gill, R.
    Hilaire, M. R.
    Venkateswaran, A.
    Harney, R.
    Jarrah, J.
    Distefano, K.
    Salamone, S.
    Baburina, I.
    CLINICA CHIMICA ACTA, 2019, 493 : S767 - S768
  • [6] Novel Point-of-Care Whole Blood Thrombin Generation Assay
    Stephens, Daniel
    Butenas, Saulius
    Mann, Kenneth G.
    Ferrara, Micheal J.
    Joseph, Immermann M.
    Spears, Grant M.
    Bailey, Kent R.
    Jenkins, Donald
    Kozar, Rosemary
    Ashrani, Aneel A.
    Park, Myung S.
    TRANSFUSION, 2019, 59 : 114A - 114A
  • [7] Evaluation of agreement between a noninvasive method for real-time measurement of critical blood values with a standard point-of-care device
    Ettrich, Rudi H.
    Caballero, Joshua
    Sakharkar, Prashant
    Ahmed, Sultan
    Hurlston, Traci
    Parmar, Jayesh
    Deb, Subrata
    PLOS ONE, 2024, 19 (06):
  • [8] Accuracy of a point-of-care blood lactate measurement device in a prehospital setting
    Walther, Louise Houlberg
    Zegers, Floor
    Nybo, Mads
    Mogensen, Christian Backer
    Christensen, Erika Frischknecht
    Lassen, Annmarie Touborg
    Mikkelsen, Soren
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2022, 36 (06) : 1679 - 1687
  • [9] Optoelectronic microfluidic device for point-of-care blood plasma viscosity measurement
    Bakhtiaridoost, Somayyeh
    Musuroi, Cristian
    Volmer, Marius
    Florescu, Monica
    LAB ON A CHIP, 2024, 24 (13) : 3305 - 3314
  • [10] Accuracy of a point-of-care blood lactate measurement device in a prehospital setting
    Louise Houlberg Walther
    Floor Zegers
    Mads Nybo
    Christian Backer Mogensen
    Erika Frischknecht Christensen
    Annmarie Touborg Lassen
    Søren Mikkelsen
    Journal of Clinical Monitoring and Computing, 2022, 36 : 1679 - 1687