Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method

被引:6
|
作者
Okahara, Shoko [1 ]
Handoh, Tetsuya [2 ]
Wakita, Mitsuru [2 ]
Yamamoto, Takamasa [2 ]
Misawa, Shigeki [2 ]
Miida, Takashi [2 ]
Sumikura, Hiroyuki [1 ]
机构
[1] Juntendo Univ Hosp, Dept Anesthesiol, Bunkyo Ku, 3-1-3 Hongo, Tokyo 1138431, Japan
[2] Juntendo Univ Hosp, Dept Clin Lab, Bunkyo Ku, 3-1-3 Hongo, Tokyo 1138431, Japan
关键词
Clauss method; FibCare; Fibrinogen; Hemorrhage; Point-of-care testing; TRAUMA; PLASMA;
D O I
10.1007/s00540-021-02987-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Timely fibrinogen replacement is key to treating critical hemorrhage. Measuring fibrinogen concentration by conventional laboratory tests requires centrifugation of blood samples and is often time-consuming. A point-of-care testing device (A&T, Yokohama, Japan), CG02N, has been available in Japan since 2011 to measure fibrinogen concentration without centrifugation. However, it has not been widely used as it requires dilution of blood samples using manual micropipetting. To further speed up and simplify the fibrinogen measurement, an improved device called FibCare (Atom Medical, Tokyo, Japan) was developed to avoid diluting blood samples. The purpose of this study is to verify the reliability of FibCare against laboratory measurement using the Clauss method. Fibrinogen concentrations with 60 sodium citrated whole blood samples were measured by both FibCare and Clauss methods in the laboratory. Measured values with the Clauss method were distributed in the 88-300 mg/dL range. By comparing these results, a significant positive correlation was observed between the FibCare and Clauss method (Y = 12.402 + 0.982 X; R = 0.891; P < 0.01). The study indicated that FibCare allows accurate measurement of fibrinogen concentration and shows a possibility to contribute to optimal fibrinogen replacement therapy during critical hemorrhage.
引用
收藏
页码:757 / 760
页数:4
相关论文
共 50 条
  • [41] Clinical concordance of eGFR measurement between laboratory plasma and whole blood point-of-care creatinine methods
    Korpi-Steiner, N. L.
    Wockenfus, A. M.
    Lakin, J. M.
    Koch, C. D.
    Karon, B. S.
    CLINICAL CHEMISTRY, 2009, 55 (06) : A99 - A100
  • [42] INACCURACY OF POINT-OF-CARE (POC) WHOLE BLOOD INR COMPARED TO VENOUS PLASMA INR AT HIGH INR VALUES
    Selby, R.
    Rozmanc, M.
    Bartle
    Huth, B.
    Geerts, W.
    Jaya, R.
    Reis, M.
    AMERICAN JOURNAL OF HEMATOLOGY, 2008, 83 (12) : 944 - 944
  • [43] Comparison of noninvasive and invasive point-of-care testing methods with reference method for hemoglobin measurement
    Avcioglu, Gamze
    Nural, Cemil
    Yilmaz, Fatma Meric
    Baran, Pervin
    Erel, Ozcan
    Yilmaz, Gulsen
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2018, 32 (03)
  • [44] Evaluation of the LABGEO PT10 Point-Of-Care Testing Device: Comparison of Analyte Measurements in Capillary Whole Blood and Lithium Heparin Whole Blood Samples With Those in Central Laboratory
    Lim, Jinsook
    Kim, Hyunjin
    Koo, Sun Hoe
    Kwon, Gye Cheol
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2017, 31 (03)
  • [45] A novel vertical flow assay for point of care measurement of iron from whole blood
    Serhan, Michael
    Jackemeyer, David
    Karam, Karam Abi
    Chakravadhanula, Kasyap
    Sprowls, Mark
    Cay-Durgun, Pinar
    Forzani, Erica
    ANALYST, 2021, 146 (05) : 1633 - 1641
  • [46] Compact Point-of-Care Device for Self-Administered HIV Viral Load Tests from Whole Blood
    Liu, Tianyi
    Politza, Anthony J.
    Kshirsagar, Aneesh
    Zhu, Yusheng
    Guan, Weihua
    ACS SENSORS, 2023, 8 (12) : 4716 - 4727
  • [47] A paper-based point-of-care device for the detection of cysteine using gold nanoparticles from whole blood
    Kumari, Monika
    Kumar, Natish
    Kumar, Sunny
    Gandhi, Shivani
    Zussman, Eyal
    Arun, Ravi Kumar
    ANALYTICAL METHODS, 2024, 16 (19) : 3007 - 3019
  • [48] Effect of Implementation of a Whole Blood Point-of-Care Device on After-Hours Call Backs at a Rural Hospital
    Abdullah, Amid
    Boechler, Carol
    King, Dianne
    Guttridge, Tracy
    Naugler, Christopher
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2012, 138 : A080 - A080
  • [49] IMPLEMENTATION OF A POINT-OF-CARE QUALITY CONTROL FOR BLOOD GLUCOSE MEASUREMENT
    Finderle, P.
    Prezelj, M.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S678 - S678
  • [50] Comparison of a new point-of-care whole blood prothrombin time system to a laboratory reference instrument.
    Spink, BJ
    Schreiner, K
    Buschmann, DJ
    Taborski, U
    BLOOD, 2000, 96 (11) : 102B - 102B