Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method

被引:6
|
作者
Okahara, Shoko [1 ]
Handoh, Tetsuya [2 ]
Wakita, Mitsuru [2 ]
Yamamoto, Takamasa [2 ]
Misawa, Shigeki [2 ]
Miida, Takashi [2 ]
Sumikura, Hiroyuki [1 ]
机构
[1] Juntendo Univ Hosp, Dept Anesthesiol, Bunkyo Ku, 3-1-3 Hongo, Tokyo 1138431, Japan
[2] Juntendo Univ Hosp, Dept Clin Lab, Bunkyo Ku, 3-1-3 Hongo, Tokyo 1138431, Japan
关键词
Clauss method; FibCare; Fibrinogen; Hemorrhage; Point-of-care testing; TRAUMA; PLASMA;
D O I
10.1007/s00540-021-02987-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Timely fibrinogen replacement is key to treating critical hemorrhage. Measuring fibrinogen concentration by conventional laboratory tests requires centrifugation of blood samples and is often time-consuming. A point-of-care testing device (A&T, Yokohama, Japan), CG02N, has been available in Japan since 2011 to measure fibrinogen concentration without centrifugation. However, it has not been widely used as it requires dilution of blood samples using manual micropipetting. To further speed up and simplify the fibrinogen measurement, an improved device called FibCare (Atom Medical, Tokyo, Japan) was developed to avoid diluting blood samples. The purpose of this study is to verify the reliability of FibCare against laboratory measurement using the Clauss method. Fibrinogen concentrations with 60 sodium citrated whole blood samples were measured by both FibCare and Clauss methods in the laboratory. Measured values with the Clauss method were distributed in the 88-300 mg/dL range. By comparing these results, a significant positive correlation was observed between the FibCare and Clauss method (Y = 12.402 + 0.982 X; R = 0.891; P < 0.01). The study indicated that FibCare allows accurate measurement of fibrinogen concentration and shows a possibility to contribute to optimal fibrinogen replacement therapy during critical hemorrhage.
引用
收藏
页码:757 / 760
页数:4
相关论文
共 50 条
  • [31] COMPARISON OF GLUCOSE POINT-OF-CARE VALUES WITH LABORATORY VALUES IN CRITICALLY ILL PATIENTS
    Shearer, Anna
    Boehmer, Marilyn
    Closs, Melanie
    Dela Rosa, Rosalina
    Hamilton, Jean
    Horton, Karen
    McGrath, Rose
    Schulman, Christine
    AMERICAN JOURNAL OF CRITICAL CARE, 2009, 18 (03) : 224 - 230
  • [32] Critical evaluation of the use of a whole blood point-of-care (POC) device by home care nurses in for assessing coumadin prophylaxis
    Gosselin, R
    Pollock, M
    Owings, J
    Larkin, E
    THROMBOSIS AND HAEMOSTASIS, 1997, : P3119 - P3119
  • [33] A Novel Whole Blood Point-of-Care Coagulometer to Measure the Effect of Direct Oral Anticoagulants and Heparins
    Ansell, Jack
    Zappe, Stefan
    Jiang, Xuan
    Chen, Lirong
    Steiner, Solomon
    Laulicht, Bryan
    Bakhru, Sasha
    SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2019, 45 (03): : 259 - 263
  • [34] A Novel Point-of-Care Whole Blood Coagulation Assay to Monitor Emicizumab Therapy in Patients with Hemophilia
    Suster, Michael A.
    Maji, Debnath
    Nayak, Lalitha V.
    Jenkins, Christina
    Hunter, Susan
    Schmaier, Alvin H.
    Mohseni, Pedram
    Ahuja, Sanjay
    BLOOD, 2018, 132
  • [35] A cartridge based Point-of-Care device for complete blood count
    Usama Abbasi
    Prasanta Chowdhury
    Sasikala Subramaniam
    Prakhar Jain
    Nitin Muthe
    Faisal Sheikh
    Subham Banerjee
    V. Kumaran
    Scientific Reports, 9
  • [36] Fast blood plasma separation device for point-of-care applications
    Durc, Pavol
    Foret, Frantisek
    Kuban, Petr
    TALANTA, 2018, 183 : 55 - 60
  • [37] A cartridge based Point-of-Care device for complete blood count
    Abbasi, Usama
    Chowdhury, Prasanta
    Subramaniam, Sasikala
    Jain, Prakhar
    Muthe, Nitin
    Sheikh, Faisal
    Banerjee, Subham
    Kumaran, V.
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [38] The qLabs® FIB system, a novel point-of-care technology for a rapid and accurate quantification of functional fibrinogen concentration from a single drop of citrated whole blood
    Sanfilippo, Sandra
    Buisson, Leslie
    Rouabehi, Hafsa
    Dujaric, Marie-Eglantine
    Donnet, Thibault
    de Raucourt, Emmanuelle
    Dumont, Benedicte
    Peynaud-Debayle, Edith
    THROMBOSIS RESEARCH, 2023, 226 : 159 - 164
  • [39] Leukodepletion as a Point-of-Care Method for Monitoring HIV-1 Viral Load in Whole Blood
    Titchmarsh, Logan
    Zeh, Clement
    Verpoort, Thierry
    Allain, Jean-Pierre
    Lee, Helen
    JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (04) : 1080 - 1086
  • [40] Improvement of point of care testing device for accurate whole blood glucose measurement in early neonates
    Koyano, Kosuke
    Arioka, Makoto
    Nakao, Yasuhiro
    Morimoto, Aya
    Sugino, Masashiro
    Morita, Hirosuke
    Nakamura, Shinji
    Kondo, Sonoko
    Konishi, Yukihiko
    Yasuda, Saneyuki
    Kusaka, Takashi
    ANNALS OF CLINICAL BIOCHEMISTRY, 2024, 61 (05) : 386 - 390