Evaluation of agreement between a noninvasive method for real-time measurement of critical blood values with a standard point-of-care device

被引:0
|
作者
Ettrich, Rudi H. [1 ,2 ]
Caballero, Joshua [3 ]
Sakharkar, Prashant [4 ]
Ahmed, Sultan [1 ]
Hurlston, Traci [1 ]
Parmar, Jayesh [4 ]
Deb, Subrata [4 ]
机构
[1] Larkin Univ, Coll Biomed Sci, Miami, FL 33169 USA
[2] Florida Int Univ, Wertheim Coll Med, Miami, FL 33199 USA
[3] Univ Georgia, Coll Pharm, Athens, GA USA
[4] Larkin Univ, Coll Pharm, Miami, FL 33169 USA
来源
PLOS ONE | 2024年 / 19卷 / 06期
关键词
D O I
10.1371/journal.pone.0304706
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The purpose of this work was to investigate the degree of agreement between two distinct approaches for measuring a set of blood values and to compare comfort levels reported by participants when utilizing these two disparate measurement methods. Radial arterial blood was collected for the comparator analysis using the Abbott i-STAT (R) POCT device. In contrast, the non-invasive proprietary DBC methodology is used to calculate sodium, potassium, chloride, ionized calcium, total carbon dioxide, pH, bicarbonate, and oxygen saturation using four input parameters (temperature, hemoglobin, pO2, and pCO2). Agreement between the measurement for a set of blood values obtained using i-STAT and DBC methodology was compared using intraclass correlation coefficients, Passing and Bablok regression analyses, and Bland Altman plots. A p-value of <0.05 was considered statistically significant. A total of 37 participants were included in this study. The mean age of the participants was 42.4 +/- 13 years, most were male (65%), predominantly Caucasian/White (75%), and of Hispanic ethnicity (40%). The Intraclass Correlation Coefficients (ICC) analyses indicated agreement levels ranging from poor to moderate between i-STAT and the DBC's algorithm for Hb, pCO2, HCO3, TCO2, and Na, and weak agreement for pO2, HSO2, pH, K, Ca, and Cl. The Passing and Bablok regression analyses demonstrated that values for Hb, pO2, pCO2, TCO2, Cl, and Na obtained from the i-STAT did not differ significantly from that of the DBC's algorithm suggesting good agreement. The values for Hb, K, and Na measured by the DBC algorithm were slightly higher than those obtained by the i-STAT, indicating some systematic differences between these two methods on Bland Altman Plots. The non-invasive DBC methodology was found to be reliable and robust for most of the measured blood values compared to invasive POCT i-STAT device in healthy participants. These findings need further validation in larger samples and among individuals afflicted with various medical conditions.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Novel noninvasive point-of-care device for real time hemoglobin monitoring
    Timm, Ulrich
    Gewiss, Helge
    Kraitl, Jens
    Stuepmann, Kirstin
    Hinz, Michael
    Koball, Sebastian
    Ewald, Hartmut
    OPTICAL DIAGNOSTICS AND SENSING XIV: TOWARD POINT-OF-CARE DIAGNOSTICS, 2014, 8951
  • [2] Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method
    Shoko Okahara
    Tetsuya Handoh
    Mitsuru Wakita
    Takamasa Yamamoto
    Shigeki Misawa
    Takashi Miida
    Hiroyuki Sumikura
    Journal of Anesthesia, 2021, 35 : 757 - 760
  • [3] Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method
    Okahara, Shoko
    Handoh, Tetsuya
    Wakita, Mitsuru
    Yamamoto, Takamasa
    Misawa, Shigeki
    Miida, Takashi
    Sumikura, Hiroyuki
    JOURNAL OF ANESTHESIA, 2021, 35 (05) : 757 - 760
  • [4] Point-of-care measurement of fetal blood lactate - Time to trust a new device
    Wang, Mandy
    Chua, Seng C.
    Bouhadir, Lilain
    Treadwell, Erin L.
    Gibbs, Emma
    McGee, Therese M.
    AUSTRALIAN & NEW ZEALAND JOURNAL OF OBSTETRICS & GYNAECOLOGY, 2018, 58 (01): : 72 - 78
  • [5] A Rapid and Point-of-Care Device for Real-Time Monitoring of Pulmonary Hypertension Biomarkers
    Noushin, Tanzila
    Tabassum, Shawana
    CIRCULATION, 2021, 144
  • [6] A facile and real-time spectroscopic method for biofluid analysis in point-of-care diagnostics
    Kumar, Rajesh
    Singh, Gajendra P.
    Barman, Ishan
    Dingari, Narahara Chari
    Nabi, Ghulam
    BIOANALYSIS, 2013, 5 (15) : 1853 - 1861
  • [7] A Real-time Realization of an Automatic Dynamic Range Adjustment Method on a Smart Mobile Device for Point-of-Care Ultrasound Imaging
    Kim, Jee Hoo
    Lee, YeonHwa
    Kang, Jinbum
    Yoo, Yangmo
    2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 1229 - 1232
  • [8] A portable, high-throughput real-time quantitative PCR device for point-of-care testing
    Yin, Juxin
    Tong, Jizhi
    Li, Jiale
    Shao, Guangye
    Xie, Bo
    Zhuang, Jianjian
    Bi, Gang
    Mu, Ying
    ANALYTICAL BIOCHEMISTRY, 2023, 674
  • [9] Accuracy of a point-of-care blood lactate measurement device in a prehospital setting
    Walther, Louise Houlberg
    Zegers, Floor
    Nybo, Mads
    Mogensen, Christian Backer
    Christensen, Erika Frischknecht
    Lassen, Annmarie Touborg
    Mikkelsen, Soren
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2022, 36 (06) : 1679 - 1687
  • [10] Optoelectronic microfluidic device for point-of-care blood plasma viscosity measurement
    Bakhtiaridoost, Somayyeh
    Musuroi, Cristian
    Volmer, Marius
    Florescu, Monica
    LAB ON A CHIP, 2024, 24 (13) : 3305 - 3314