Advanced Design of Birdcage RF Coil for Various Absorption Regions at 3T MRI System

被引:0
|
作者
Lee, Jung-Woo [1 ]
Choe, Bo-Young [1 ]
Yoon, Sung-Ik [1 ]
Choi, Chi-Bong [1 ]
Huh, Soon-Nyoung [2 ]
机构
[1] Catholic Univ Korea, Kangnam St Marys Hosp, Coll Med, Dept Biomed Engn, 505 Banpo Dong, Seoul 137040, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Radiat Oncol, Seoul 110744, South Korea
来源
关键词
Birdcage Resonator; 3; T; Signal-to-Noise Ratio (SNR);
D O I
暂无
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Purpose: The purpose of this study was to design and build an optimized birdcage resonator configuration with a low pass filter, which would facilitate the acquisition of high-resolution 3D-images of small animals at 3T MRI system. Methods and Materials: The birdcage resonator with 12-element structures was built, in order to ensure B1 homogeneity over the image volume and maximum filling factor, and hence to maximize the signal to noise ratio (SNR) and resolution of the 3-dimensional images. The diameter and length of each element of a birdcage resonator were as follows: (1) diameter 13 cm, length 22 cm, (2) diameter 15 cm, length 22 cm, (3) diameter 17 cm, length 25 cm. Spin echo pulse sequence and fast spin echo pulse sequence were employed in obtaining MR images. The quality of the manufactured birdcage resonators was evaluated on the basis of the return loss following matching and tuning process. Results: The experimental MR images of phantoms by the various manufactured birdcage resonators were obtained to compare the SNR in accordance with the size of objects. The size of an object to that of coil was identified by parameters that were estimated from the images of a phantom. First, the diameter of the birdcage resonator was 15cm, and the ratio of the tangerine to the birdcage resonator accounted for approximately 27%. The Q factor was 53.2 and the SNR was 150.7. Second, at the same birdcage resonator, the ratio of the orange was approximately 53%. The SNR and the Q parameter was 212.8 and 91.2, respectively. Conclusion: The present study demonstrated that if birdcage resonators have the same forms, SNR could be different depending on the size of an object, especially when the size of an object to that of coil is approximately 40 similar to 80%, the former is bigger than the latter. Therefore, when the size of an object to be observed is smaller than that of coil, the coil should be manufactured in accordance with the size of an object in order to obtain much more excellent images.
引用
收藏
页码:48 / 60
页数:13
相关论文
共 50 条
  • [31] Design and Analysis of Microstrip-Based RF Birdcage Coil for 1.5 T Magnetic Resonance Imaging
    Thiyagarajan, Krishnan
    Kesavamurthy, Thangavelu
    Bharathkumar, Gattu
    APPLIED MAGNETIC RESONANCE, 2014, 45 (03) : 255 - 268
  • [32] Magnetic Field Homogeneity of Birdcage Coil for 4 T MRI System with No Lumped Circuit Elements
    Suga, Ryotaro
    Saito, Kazuyuki
    Takahashi, Masaharu
    Ito, Koichi
    IEICE TRANSACTIONS ON COMMUNICATIONS, 2014, E97B (04): : 791 - 797
  • [33] Birdcage Coil with Inductively Coupled RF Coil Array for Improving |B1| -Field Sensitivity In 7-T MRI
    Byun, Jong-Deok
    Seo, Jeung-Hoon
    Kang, Taewon
    Ryu, Yeunchul
    Kim, Kyoung-Nam
    JOURNAL OF MAGNETICS, 2017, 22 (03) : 378 - 381
  • [34] RF Transmit/Receive Quadrature Fed Body Birdcage Coil for a 1.5T MR System
    Yadav, Niraj S.
    Kochar, Inderkumar
    Bhuiya, Tapas K.
    Harsh, Rajesh
    2018 2ND INTERNATIONAL CONFERENCE ON ELECTRONICS, MATERIALS ENGINEERING & NANO-TECHNOLOGY (IEMENTECH), 2018, : 469 - 475
  • [35] Dynamic contrast enhanced imaging of mice kidney metabolism using high-temperature superconducting RF coil on a 3T MRI system
    Chen, Kai-Yuan
    Kuo, Li-Wei
    Lin, Wei-Ting
    Su, Chia-Hao
    Chen, Jyh-Horng
    2007 JOINT MEETING OF THE 6TH INTERNATIONAL SYMPOSIUM ON NONINVASIVE FUNCTIONAL SOURCE IMAGING OF THE BRAIN AND HEART AND THE INTERNATIONAL CONFERENCE ON FUNCTIONAL BIOMEDICAL IMAGING, 2007, : 180 - 182
  • [36] Size-Adaptive Occipital 18-Channel Receive-Only RF Coil for 3T MRI
    Mathieu, William
    Popovic, Milica
    Farivar, Reza
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2024,
  • [37] Influence of reference tube location on the measured sodium concentrations in calf muscles using a birdcage coil at 3T
    Baron, Paul
    Potze, Jan Hendrik
    Sijens, Paul E.
    MAGNETIC RESONANCE IN MEDICINE, 2023, 90 (02) : 624 - 632
  • [38] Implementation of high-temperature superconducting tapes RF coils for 3T MRI system
    Lin, In-Tsang
    Hsieh, Chang-Wei
    Kuo, Li-Wei
    Yang, Hong-Chang
    Yao, Ching
    Chang, Wei-Hao
    Chen, Jyh-Horng
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 7413 - 7416
  • [39] Implementation of bi-2223 HTS tapes as RF coils for 3T MRI system
    Lin, In-Tsang
    Hsieh, Chang-Wei
    Kuo, Li-Wei
    Yang, Hong-Chang
    Horng, Herng-Er
    Chang, Wei-Hao
    Chen, Jyh-Homg
    TENCON 2005 - 2005 IEEE REGION 10 CONFERENCE, VOLS 1-5, 2006, : 2426 - +
  • [40] Simulation Design of Incremental Leg Tapered Birdcage Coil for Head Imaging at 4.7T MRI
    Song, Hyunwoo
    Kim, Kyoung-Nam
    Hernandez, Daniel
    APPLIED SCIENCES-BASEL, 2021, 11 (05): : 1 - 7