Birdcage Coil with Inductively Coupled RF Coil Array for Improving |B1| -Field Sensitivity In 7-T MRI

被引:6
|
作者
Byun, Jong-Deok [1 ]
Seo, Jeung-Hoon [2 ]
Kang, Taewon [1 ]
Ryu, Yeunchul [3 ]
Kim, Kyoung-Nam [4 ]
机构
[1] Kangwon Natl Univ, Dept Mech & Biomed Engn, Kangwon Do 24341, South Korea
[2] Gachon Univ, Neurosci Res Inst, Incheon 21565, South Korea
[3] Gachon Univ, Dept Radiol Sci, Incheon 13120, South Korea
[4] Gachon Univ, Dept Biomed Engn, Incheon 13120, South Korea
基金
新加坡国家研究基金会;
关键词
Magnetic Resonance Imaging (MRI); Radio-Frequency (RF) coil; Inductively Coupled Coil; 7-Tesla (T); VOLUME COIL; HOMOGENEITY; DESIGN; MODE;
D O I
10.4283/JMAG.2017.22.3.378
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A birdcage coil integrated with a 16-channel, inductively coupled, radio-frequency (RF) array was designed to improve the magnetic-flux distribution for human brain magnetic resonance imaging at 7-T. A numerical calculation was performed using the finite-difference time-domain method for the birdcage coil with an inductively coupled array. The electromagnetic calculation results for the birdcage coil with the inductively coupled array were compared to those for the birdcage coil without an inductively coupled array in terms of their magnetic-flux, electric-field, and specific-absorption-rate distributions in a cylindrical phantom and human model. The proposed birdcage coil with an inductively coupled array offers a superior magnetic-flux sensitivity without sacrificing the RF power deposition at 7-T. The modifications of the coil geometry accompanying the inductively coupled RF array could be applied to the generally used transmit/receive volume coils and extended to parallel RF transmission array in ultra-high-field magnetic resonance imaging.
引用
收藏
页码:378 / 381
页数:4
相关论文
共 50 条
  • [1] Comparison of birdcage and phase array coil using FDTD for the B1 homogeneity in high field MRI
    Seo, J. H.
    Heo, H. Y.
    Han, B. H.
    Hee, S. Y.
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 3902 - 3905
  • [2] B1 field homogeneity and SAR calculations for the birdcage coil
    Ibrahim, TS
    Lee, R
    Baertlein, BA
    Robitaille, PML
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02): : 609 - 619
  • [3] Investigation of the B1 field distribution and RF power deposition in a birdcage coil as functions of the number of coil legs at 4.7 T, 7.0 T, and 11.7 T
    Seo, Jeung-Hoon
    Han, Sang-Doc
    Kim, Kyoung-Nam
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 66 (12) : 1822 - 1826
  • [4] Investigation of the B1 field distribution and RF power deposition in a birdcage coil as functions of the number of coil legs at 4.7 T, 7.0 T, and 11.7 T
    Jeung-Hoon Seo
    Sang-Doc Han
    Kyoung-Nam Kim
    Journal of the Korean Physical Society, 2015, 66 : 1822 - 1826
  • [5] Improving SNR and B1 transmit field for an endorectal coil in 7 T MRI and MRS of prostate cancer
    de Castro, C. S. Arteaga
    van den Bergen, B.
    Luijten, P. R.
    van der Heide, U. A.
    van Vulpen, M.
    Klomp, D. W. J.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (01) : 311 - 318
  • [6] Surface Coil with an Inductively Coupled Wireless Surface and Volume Coil for Improving the Magnetic Field Sensitivity at 400-MHz MRI
    Seo, Jeung-Hoon
    Lee, Jae Jun
    Kim, Kyoung-Nam
    JOURNAL OF MAGNETICS, 2018, 23 (02) : 192 - 195
  • [7] Simulations of a birdcage coil B1+ field on a human body model for designing a 3T multichannel TMS/MRI head coil array
    de Lara, Lucia I. Navarro
    Rad, Laleh Golestani
    Makarov, Sergey N.
    Stockmann, Jason
    Wald, Lawrence L.
    Nummenmaa, Aapo
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 4752 - 4755
  • [8] Helmholtz transceiver array for improving the |B1|-field homogeneity at 7-T magnetic resonance imaging
    J.-H. Seo
    H. Song
    H. J. Kim
    S.-D. Han
    P. Heo
    D. Kim
    Y. Ryu
    K.-N. Kim
    Physics of Wave Phenomena, 2017, 25 : 147 - 150
  • [9] Helmholtz transceiver array for improving the |B1|-field homogeneity at 7-T magnetic resonance imaging
    Seo, J. -H.
    Song, H.
    Kim, H. J.
    Han, S. -D.
    Heo, P.
    Kim, D.
    Ryu, Y.
    Kim, K. -N.
    PHYSICS OF WAVE PHENOMENA, 2017, 25 (02) : 147 - 150
  • [10] Computing the B1 field of the toroidal MRI coil
    Butterworth, EJ
    Gore, JC
    JOURNAL OF MAGNETIC RESONANCE, 2005, 175 (01) : 114 - 123