Patient dose in full-field digital mammography: an Italian survey

被引:0
|
作者
Gisella Gennaro
Paola Baldelli
Angelo Taibi
Cosimo Di Maggio
Mauro Gambaccini
机构
[1] Padua University,Department of Oncological Sciences
[2] Ferrara University,Department of Physics
来源
European Radiology | 2004年 / 14卷
关键词
Digital mammography; Patient dose; Breast imaging;
D O I
暂无
中图分类号
学科分类号
摘要
The aim of this study was to compare performance and patient dose of full-field digital mammography units for clinical use. Measurements of linearity and automatic exposure control stability were performed on four units installed in as many Italian sites. The tube output was also obtained by the same ionization chamber, permitting to evaluate ratios mGy/mAs for each available spectrum. The entrance air-kerma was calculated over a sample of 800 cranio-caudal mammograms and the average glandular dose obtained, assuming two mean glandular compositions of 50 and 30%, respectively. Digital systems showed very good linearity and comparable responses. The stability of the automatic exposure control was better than 5% for all systems. Regarding doses, the two mammography units that work mainly in contrast mode deliver, respectively, 17 and 28% more dose compared with those working in standard mode. For the latter mode, the mean average glandular dose was in the range 1.25–1.37 mGy and 1.37–1.49 mGy for the 50 and 30% glandular composition, respectively. Results of this study were compared with those of other surveys, showing that full-field digital mammography allows a significant clinical dose reduction compared with screen/film mammography.
引用
收藏
页码:645 / 652
页数:7
相关论文
共 50 条
  • [21] Current status of full-field digital mammography
    Pisano, ED
    [J]. RADIOLOGY, 2000, 214 (01) : 26 - 28
  • [22] Mammographic Artifacts on Full-Field Digital Mammography
    Jae Jeong Choi
    Sung Hun Kim
    Bong Joo Kang
    Byung Gil Choi
    ByungJoo Song
    Haijo Jung
    [J]. Journal of Digital Imaging, 2014, 27 : 231 - 236
  • [23] Mammographic Artifacts on Full-Field Digital Mammography
    Choi, Jae Jeong
    Kim, Sung Hun
    Kang, Bong Joo
    Choi, Byung Gil
    Song, ByungJoo
    Jung, Haijo
    [J]. JOURNAL OF DIGITAL IMAGING, 2014, 27 (02) : 231 - 236
  • [24] Advantages of gridless full-field digital mammography
    Nykänen, K
    Siltanen, S
    [J]. MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 137 - 146
  • [25] Current status of full-field digital mammography
    Pisano, ED
    Yaffe, MJ
    Hemminger, BM
    Hendrick, RE
    Niklason, LT
    Maidment, ADA
    Kimme-Smith, CM
    Feig, SA
    Sickles, EA
    Braeuning, MP
    [J]. ACADEMIC RADIOLOGY, 2000, 7 (04) : 266 - 280
  • [26] Potential of dose reduction after marker placement with full-field digital mammography
    Riedl, CC
    Jaromi, S
    Floery, D
    Pfarl, G
    Fuchsjaeger, MH
    Helbich, TH
    [J]. INVESTIGATIVE RADIOLOGY, 2005, 40 (06) : 343 - 348
  • [27] Dose reduction in full-field digital mammography: an anthropomorphic breast phantom study
    Obenauer, S
    Hermann, KP
    Grabbe, E
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (907): : 478 - 482
  • [28] Comparison of Mean Glandular Dose between Full-Field Digital Mammography and Digital Breast Tomosynthesis
    Teoh, Kar Choon
    Manan, Hanani Abdul
    Norsuddin, Norhashimah Mohd
    Rizuana, Iqbal Hussain
    [J]. HEALTHCARE, 2021, 9 (12)
  • [29] Performance of full-field digital mammography versus digital breast
    Wang, Mengru
    Zhuang, Shan
    Sheng, Liuli
    Zhao, Yu Nian
    Shen, Wenrong
    [J]. PRECISION MEDICAL SCIENCES, 2022, 11 (02): : 56 - 61
  • [30] Implementation of telemanagement for full-field digital mammography (FFDM)
    Fan, Y
    Lou, SL
    Wong, A
    Zhang, H
    Sickles, EA
    [J]. MEDICAL IMAGING 2000: PACS DESIGN AND EVALUATION - ENGINEERING AND CLINICAL ISSUES, 2000, 3980 : 400 - 407