NEMA NU-4 2008 performance evaluation of Xtrim-PET: A prototype SiPM-based preclinical scanner

被引:24
|
作者
Amirrashedi, Mahsa [1 ,2 ]
Sarkar, Saeed [1 ,2 ]
Ghafarian, Pardis [3 ,4 ]
Shahraki, Reza Hashemi [1 ,2 ]
Geramifar, Parham [5 ]
Zaidi, Habib [6 ,7 ,8 ,9 ]
Ay, Mohammad Reza [1 ,2 ]
机构
[1] Univ Tehran Med Sci, Dept Med Phys & Biomed Engn, Tehran, Iran
[2] Univ Tehran Med Sci, Res Ctr Mol & Cellular Imaging, Tehran, Iran
[3] Shahid Beheshti Univ Med Sci, Chron Resp Dis Res Ctr, NRITLD, Tehran, Iran
[4] Shahid Beheshti Univ Med Sci, Masih Daneshvari Hosp, PET CT & Cyclotron Ctr, Tehran, Iran
[5] Univ Tehran Med Sci, Shariati Hosp, Res Ctr Nucl Med, Tehran, Iran
[6] Geneva Univ Hosp, Div Nucl Med & Mol Imaging, CH-1211 Geneva, Switzerland
[7] Univ Geneva, Neuroctr, CH-1205 Geneva, Switzerland
[8] Univ Groningen, Univ Med Ctr Groningen, Dept Nucl Med & Mol Imaging, NL-9700 RB Groningen, Netherlands
[9] Univ Southern Denmark, Dept Nucl Med, DK-500 Odense, Denmark
关键词
detectors; NEMA standard; performance evaluation; PET; small animal imaging; PET/MRI; INSERT; ALGORITHMS; SYSTEM; MRI;
D O I
10.1002/mp.13785
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Xtrim-PET is a newly designed Silicon Photomultipliers (SiPMs)-based prototype PET scanner dedicated for small laboratory animal imaging. We present the performance evaluation of the Xtrim-PET scanner following NEMA NU-4 2008 standards to help optimizing scanning protocols which can be achieved through standard and reliable system performance characterization. Methods The performance assessment was conducted according to the National Electrical Manufacturers Association (NEMA) NU-4 2008 standards in terms of spatial resolution, sensitivity, counting rate performance, scatter fraction and image quality. The in vivo imaging capability of the scanner is also showcased through scanning a normal mouse injected with F-18-FDG. Furthermore, the performance characteristics of the developed scanner are compared with commercially available systems and current prototypes. Results The volumetric spatial resolution at 5 mm radial offset from the central axis of the scanner is 6.81 mu l, whereas a peak absolute sensitivity of 2.99% was achieved using a 250-650 keV energy window and a 10 ns timing window. The peak noise-equivalent count rate (NECR) using a mouse-like phantom is 113.18 kcps at 0.34 KBq/cc with 12.5% scatter fraction, whereas the NECR peaked at 82.76 kcps for an activity concentration level of 0.048 KBq/cc with a scatter fraction of 25.8% for rat-like phantom. An excellent uniformity (3.8%) was obtained using NEMA image quality phantom. Recovery coefficients of 90%, 86%, 68%, 40% and 12% were calculated for rod diameters of 5, 4, 3, 2 and 1 mm, respectively. Spill-over ratios for air-filled and water-filled chambers were 35% and 25% without applying any correction for attenuation and Compton scattering effects. Conclusion Our findings revealed that beyond compactness, lightweight, easy installation and good energy resolution, the Xtrim-PET prototype presents a reasonable performance making it suitable for preclinical molecular imaging-based research.
引用
收藏
页码:4816 / 4825
页数:10
相关论文
共 50 条
  • [1] Performance Evaluation of a Newly Designed SiPM-based Preclinical Scanner Based on NEMA-NU4 2008
    Amirrashedi, M.
    Sarkar, S.
    Ghafarian, P.
    Shahraki, R. Hashemi
    Geramifar, P.
    Zaidi, H.
    Ay, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S813 - S814
  • [2] NEMA NU-4 Performance Evaluation of the IRIS PET/CT Preclinical Scanner
    Belcari N.
    Camarlinghi N.
    Ferretti S.
    Iozzo P.
    Panetta D.
    Salvadori P.A.
    Sportelli G.
    Del Guerra A.
    IEEE Transactions on Radiation and Plasma Medical Sciences, 2017, 1 (04): : 301 - 309
  • [3] Performance evaluation of the Inveon PET Scanner using GATE based on the NEMA NU-4 Standards
    Lu, Lijun
    Karakatsanis, Nicolas
    Ma, Jianhua
    Bian, Zhaoying
    Han, Yanjiang
    Tang, Jing
    Rahmim, Arman
    Chen, Wufan
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [4] Performance Evaluation of the Inveon Dedicated PET Preclinical Tomograph Based on the NEMA NU-4 Standards
    Bao, Qinan
    Newport, Danny
    Chen, Mu
    Stout, David B.
    Chatziioannou, Arion F.
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (03) : 401 - 408
  • [5] NEMA NU-4 performance evaluation of PETbox4, a high sensitivity dedicated PET preclinical tomograph
    Gu, Z.
    Taschereau, R.
    Vu, N. T.
    Wang, H.
    Prout, D. L.
    Silverman, R. W.
    Bai, B.
    Stout, D. B.
    Phelps, M. E.
    Chatziioannou, A. F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (11): : 3791 - 3814
  • [6] NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner
    Strugari, Matthew E. E.
    DeBay, Drew R. R.
    Beyea, Steven D. D.
    Brewer, Kimberly D. D.
    EJNMMI PHYSICS, 2023, 10 (01)
  • [7] PERFORMANCE EVALUATION OF PRECLINICAL PET SCANNERS WITH THE NEMA NU-4 IMAGE QUALITY PHANTOM USING A COLLECTION OF RADIOISOTOPES
    Nezich, Rik
    Reilhac, Anthonin
    Zahra, David
    Price, Roger
    INTERNAL MEDICINE JOURNAL, 2013, 43 : 23 - 23
  • [8] NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner
    Matthew E. Strugari
    Drew R. DeBay
    Steven D. Beyea
    Kimberly D. Brewer
    EJNMMI Physics, 10
  • [9] NEMA NU-4 performance evaluation of a non-human primate animal PET
    Chai Pei
    Feng Baotong
    Zhang Zhiming
    Tang Haohui
    Liu Shuanquan
    Sun Xiaoli
    Wang Peilin
    Wang Xiaoming
    Zhao Xudong
    Wei Long
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (10):
  • [10] Performance Evaluation of a New Dedicated Breast PET Scanner Using NEMA NU4-2008 Standards
    Miyake, Kanae K.
    Matsumoto, Keiichi
    Inoue, Mika
    Nakamoto, Yuji
    Kanao, Shotaro
    Oishi, Tae
    Kawase, Shigeto
    Kitamura, Keishi
    Yamakawa, Yoshiyuki
    Akazawa, Ayako
    Kobayashi, Tetsuya
    Ohi, Junichi
    Togashi, Kaori
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55 (07) : 1198 - 1203