NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner

被引:2
|
作者
Strugari, Matthew E. E. [1 ,2 ]
DeBay, Drew R. R. [1 ,3 ]
Beyea, Steven D. D. [1 ,2 ,4 ,5 ]
Brewer, Kimberly D. D. [1 ,2 ,4 ,5 ,6 ]
机构
[1] Biomed Translat Imaging Ctr, Halifax, NS, Canada
[2] Dalhousie Univ, Dept Phys & Atmospher Sci, Halifax, NS, Canada
[3] Cubresa Inc, Winnipeg, MB, Canada
[4] Dalhousie Univ, Dept Diagnost Radiol, Halifax, NS, Canada
[5] Dalhousie Univ, Sch Biomed Engn, Halifax, NS, Canada
[6] Dalhousie Univ, Dept Microbiol & Immunol, Halifax, NS, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Molecular imaging; Nuclear medicine; SPECT; Animal imaging instrumentation; Monte Carlo method; Computer-assisted image processing; Imaging phantoms; MULTI-PINHOLE SPECT; SMALL-ANIMAL SPECT; HIGH-RESOLUTION; OPTIMIZATION; CAPABILITIES;
D O I
10.1186/s40658-023-00555-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The Cubresa Spark is a novel benchtop silicon-photomultiplier (SiPM)-based preclinical SPECT system. SiPMs in SPECT significantly improve resolution and reduce detector size compared to preclinical cameras with photomultiplier tubes requiring highly magnifying collimators. The NEMA NU 1 Standard for Performance Measurements of Gamma Cameras provides methods that can be readily applied or extended to characterize preclinical cameras with minor modifications. The primary objective of this study is to characterize the Spark according to the NEMA NU 1-2018 standard to gain insight into its nuclear medicine imaging capabilities. The secondary objective is to validate a GATE Monte Carlo simulation model of the Spark for use in preclinical SPECT studies. Methods: NEMA NU 1-2018 guidelines were applied to characterize the Spark's intrinsic, system, and tomographic performance with single- and multi-pinhole collimators. Phantoms were fabricated according to NEMA specifications with deviations involving high-resolution modifications. GATE was utilized to model the detector head with the single-pinhole collimator, and NEMA measurements were employed to tune and validate the model. Single-pinhole and multi-pinhole SPECT data were reconstructed with the Software for Tomographic Image Reconstruction and HiSPECT, respectively. Results: The limiting intrinsic resolution was measured as 0.85 mm owing to a high-resolution SiPM array combined with a 3 mm-thick scintillation crystal. The average limiting tomographic resolution was 1.37 mm and 1.19 mm for the single- and multi-pinhole collimators, respectively, which have magnification factors near unity at the center of rotation. The maximum observed count rate was 15,400 cps, and planar sensitivities of 34 cps/MBq and 150 cps/MBq were measured at the center of rotation for the single- and multi-pinhole collimators, respectively. All simulated tests agreed well with measurement, where the most considerable deviations were below 7%. Conclusions: NEMA NU 1-2018 standards determined that a SiPM detector mitigates the need for highly magnifying pinhole collimators while preserving detailed information in projection images. Measured and simulated NEMA results were highly comparable with differences on the order of a few percent, confirming simulation accuracy and validating the GATE model. Of the collimators initially provided with the Spark, the multi-pinhole collimator offers high resolution and sensitivity for organ-specific imaging of small animals, and the single-pinhole collimator enables high-resolution whole-body imaging of small animals.
引用
收藏
页数:25
相关论文
共 8 条
  • [1] NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner
    Matthew E. Strugari
    Drew R. DeBay
    Steven D. Beyea
    Kimberly D. Brewer
    EJNMMI Physics, 10
  • [2] Development and validation of a Monte Carlo model for a SiPM-based PET scanner
    Merlet, A.
    Presles, B.
    Su, K. H.
    Cochet, A.
    Vrigneaud, J. M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S345 - S345
  • [3] Performance Evaluation of a Newly Designed SiPM-based Preclinical Scanner Based on NEMA-NU4 2008
    Amirrashedi, M.
    Sarkar, S.
    Ghafarian, P.
    Shahraki, R. Hashemi
    Geramifar, P.
    Zaidi, H.
    Ay, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S813 - S814
  • [4] NEMA NU-4 2008 performance evaluation of Xtrim-PET: A prototype SiPM-based preclinical scanner
    Amirrashedi, Mahsa
    Sarkar, Saeed
    Ghafarian, Pardis
    Shahraki, Reza Hashemi
    Geramifar, Parham
    Zaidi, Habib
    Ay, Mohammad Reza
    MEDICAL PHYSICS, 2019, 46 (11) : 4816 - 4825
  • [5] Validation and Performance Assessment of a Preclinical SiPM-Based SPECT/MRI Insert
    Carminati, M.
    Baratelli, F. M.
    Occhipinti, M.
    Erlandsson, K.
    Nagy, K.
    Nyitrai, Z.
    Czeller, M.
    Kuehne, A.
    Niendorf, T.
    Valtorta, S.
    Belloli, S.
    Moresco, R. M.
    Savi, A.
    Iadanza, A.
    Falini, A.
    Politi, L. S.
    Cadioli, M.
    Hutton, B. E.
    Fiorini, C.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2019, 3 (04) : 483 - 490
  • [6] NEMA NU 4-2008 validation and applications of the PET-SORTEO Monte Carlo simulations platform for the geometry of the Inveon PET preclinical scanner
    Boisson, F.
    Wimberley, C. J.
    Lehnert, W.
    Zahra, D.
    Pham, T.
    Perkins, G.
    Hamze, H.
    Gregoire, M-C
    Reilhac, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (19): : 6749 - 6763
  • [7] Performance evaluation of a high-resolution, dualhead animal SPECT system (HiReSPECT II) based on the NEMA NU1-2018 standard
    Samizadeh, M.
    Yousefzadeh, F.
    Teimourianfard, B.
    Ay, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S726 - S726
  • [8] Validation of a discovery MI 4-ring model according to the NEMA NU 2-2018 standards: from Monte Carlo simulations to clinical-like reconstructions
    Merlet, Antoine
    Presles, Benoit
    Su, Kuan-Hao
    Salvadori, Julien
    Sayah, Farzam
    Jozi, Hanieh
    Cochet, Alexandre
    Vrigneaud, Jean-Marc
    EJNMMI PHYSICS, 2024, 11 (01)