An investigation of using log-file analysis for automated patient-specific quality assurance in MRgRT

被引:8
|
作者
Lim, Seng Boh [1 ]
Godoy Scripes, Paola [1 ]
Napolitano, Mary [2 ]
Subashi, Ergys [1 ]
Tyagi, Neelam [1 ]
Cervino Arriba, Laura [1 ]
Lovelock, Dale Michael [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, 1275 York Ave, New York, NY 10021 USA
[2] Standard Imaging Inc, Middleton, WI USA
来源
关键词
elekta unity; linacview; log-file analysis; MRI guided adaptive radiotherapy; MRgRT; patient specific quality assurance; MLC PERFORMANCE; IMRT; QA;
D O I
10.1002/acm2.13361
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective Adaptive radiation therapy (ART) is an integral part of MR-guided RT (MRgRT), requiring a new RT plan for each treatment fraction and resulting in a significant increase in patient-specific quality assurance (PSQA). This study investigates the possibility of using treatment log-file for automated PSQA. Method All treatment plans were delivered in 1.5T Unity MR-Linac (Elekta). A Unity compatible version of LinacView (Standard Imaging) was commissioned to automatically monitor and analyze the log-files. A total of 220 fields were delivered and measured by ArcCheck-MR (Sun Nuclear) and LinacView. Thirty incorrectly matched fields were also delivered to check for error detection sensitivity. The gamma analysis, gamma, with 3%, 3 mm criteria was used in both ArcCheck-MR and LinacView. Additionally, the gantry angle, jaws, and multileaf collimators (MLC) positions reported in the log-file were compared with plan positions using TG-142 criteria. Result The gamma (3%, 3 mm) for the 190 plans were found to be between the range of 72.5%-100.0% and 95.4%-100.0% for ArcCheck-MR and LinacVeiw, respectively. All the delivered gantry angle and jaws were found to be within 0.2 degrees and 2 mm. MLCs that were outside the guard leaves or under the diaphragms were found to have more than 1.0 mm discrepancy. This was attributed to the linac internal override for these MLCs and had no dosimetric impact. Excluding these discrepancies, all MLC positions were found to be within 1.0 mm. The gamma (3%, 3 mm) for the 30 incorrectly matched fields were found to be 3.9%-84.8% and 0.1%-64.4% for ArcCheck-MR and LinacVeiw, respectively. Conclusion Significant ranked correlation demonstrates the automated log-file analysis can be used for PSQA and expedite the ART workflow. Ongoing PSQA will be compared with log-file analysis to investigate the longer term reproducibility and correlation.
引用
收藏
页码:183 / 188
页数:6
相关论文
共 50 条
  • [41] Patient-specific quality assurance method for VMAT treatment delivery
    Schreibmann, Eduard
    Dhabaan, Anees
    Elder, Eric
    Fox, Tim
    MEDICAL PHYSICS, 2009, 36 (10) : 4530 - 4535
  • [42] Evaluation of Patient-Specific Quality Assurance Based On Different Methods
    Yang, T.
    Qu, B.
    Dai, X.
    Xie, C.
    Xiaoshen, W.
    Cong, X.
    Wang, J.
    Xu, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 2896 - 2896
  • [43] Evaluation of a Phantom for Patient-specific Quality Assurance in the MR Linac
    Baumgartl, M.
    Safarkhanlo, Y.
    Ehrbar, S.
    Wilke, L.
    Krayenbuhl, J.
    Chamberlain, M.
    Zamburlini, M.
    Guckenberger, M.
    Andratschke, N.
    Tanadini-Lang, S.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2021, 197 (SUPPL 1) : S33 - S34
  • [44] Survey of patient-specific quality assurance practice for IMRT and VMAT
    Chan, Gordon H.
    Chin, Lee C. L.
    Abdellatif, Ady
    Bissonnette, Jean-Pierre
    Buckley, Lesley
    Comsa, Daria
    Granville, Dal
    King, Jenna
    Rapley, Patrick L.
    Vandermeer, Aaron
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (07): : 155 - 164
  • [45] Patient-specific quality assurance for IMRT delivery: A multicentre study
    Hizam, Diyana Afrina
    Jong, Wei Loong
    Zin, Hafiz Mohd
    Ng, Kwan Hoong
    Ung, Ngie Min
    RADIATION PHYSICS AND CHEMISTRY, 2023, 209
  • [46] Evaluation of Patient-Specific Quality Assurance Based On Different Methods
    Lin, X.
    Sun, T.
    Liu, X.
    MEDICAL PHYSICS, 2018, 45 (06) : E211 - E211
  • [47] Inverse Verification of the Dose Distribution for Intensity Modulated Radiation Therapy Patient-specific Quality Assurance Using Dynamic MLC Log Files
    Lee, Jeong-Woo
    Park, Jeong-Hoon
    Chung, Jin-Beom
    Park, Ji-Yeon
    Choe, Bo-Young
    Suh, Tae-Suk
    Lee, Doo-Hyun
    Hong, Semie
    Kang, Min-Young
    Choi, Kyoung-Sik
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2009, 55 (04) : 1649 - 1656
  • [48] Enhancing Patient-specific Quality Assurance in Carbon-ion Radiation Therapy: Recalculating Delivered Dose Distribution Using Log Data
    Yun, Yongdo
    Han, Min cheol
    Kim, Changhwan
    Hong, Chae-Seon
    Kim, Dong wook
    Lee, Ho
    Kim, Hojin
    Han, Soorim
    Souda, Hikaru
    Lee, Sung hyun
    Miyasaka, Yuya
    Iwai, Takeo
    Kim, Jin sung
    IN VIVO, 2025, 39 (02): : 1086 - 1093
  • [49] Analysis of the Effect of Tomotherapy Plan Parameters on Patient-Specific Delivery Quality Assurance (DQA)
    Kim, Jina
    Kay, Chul-Seung
    Jang, Hong-Seok
    Kang, Young-nam
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2019, 75 (12) : 1043 - 1047
  • [50] Analysis of the Effect of Tomotherapy Plan Parameters on Patient-Specific Delivery Quality Assurance (DQA)
    Jina Kim
    Chul-Seung Kay
    Hong-Seok Jang
    Young-nam Kang
    Journal of the Korean Physical Society, 2019, 75 : 1043 - 1047