An investigation of using log-file analysis for automated patient-specific quality assurance in MRgRT

被引:8
|
作者
Lim, Seng Boh [1 ]
Godoy Scripes, Paola [1 ]
Napolitano, Mary [2 ]
Subashi, Ergys [1 ]
Tyagi, Neelam [1 ]
Cervino Arriba, Laura [1 ]
Lovelock, Dale Michael [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, 1275 York Ave, New York, NY 10021 USA
[2] Standard Imaging Inc, Middleton, WI USA
来源
关键词
elekta unity; linacview; log-file analysis; MRI guided adaptive radiotherapy; MRgRT; patient specific quality assurance; MLC PERFORMANCE; IMRT; QA;
D O I
10.1002/acm2.13361
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective Adaptive radiation therapy (ART) is an integral part of MR-guided RT (MRgRT), requiring a new RT plan for each treatment fraction and resulting in a significant increase in patient-specific quality assurance (PSQA). This study investigates the possibility of using treatment log-file for automated PSQA. Method All treatment plans were delivered in 1.5T Unity MR-Linac (Elekta). A Unity compatible version of LinacView (Standard Imaging) was commissioned to automatically monitor and analyze the log-files. A total of 220 fields were delivered and measured by ArcCheck-MR (Sun Nuclear) and LinacView. Thirty incorrectly matched fields were also delivered to check for error detection sensitivity. The gamma analysis, gamma, with 3%, 3 mm criteria was used in both ArcCheck-MR and LinacView. Additionally, the gantry angle, jaws, and multileaf collimators (MLC) positions reported in the log-file were compared with plan positions using TG-142 criteria. Result The gamma (3%, 3 mm) for the 190 plans were found to be between the range of 72.5%-100.0% and 95.4%-100.0% for ArcCheck-MR and LinacVeiw, respectively. All the delivered gantry angle and jaws were found to be within 0.2 degrees and 2 mm. MLCs that were outside the guard leaves or under the diaphragms were found to have more than 1.0 mm discrepancy. This was attributed to the linac internal override for these MLCs and had no dosimetric impact. Excluding these discrepancies, all MLC positions were found to be within 1.0 mm. The gamma (3%, 3 mm) for the 30 incorrectly matched fields were found to be 3.9%-84.8% and 0.1%-64.4% for ArcCheck-MR and LinacVeiw, respectively. Conclusion Significant ranked correlation demonstrates the automated log-file analysis can be used for PSQA and expedite the ART workflow. Ongoing PSQA will be compared with log-file analysis to investigate the longer term reproducibility and correlation.
引用
收藏
页码:183 / 188
页数:6
相关论文
共 50 条
  • [31] Patient-Specific Quality Assurance Techniques for RapidArc Radiotherapy
    O'Daniel, J.
    McMahon, R.
    Chang, Z.
    Yan, H.
    Sakhalkar, H.
    Das, S.
    Wu, Q.
    Yin, F.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [32] Use of treatment log files in spot scanning proton therapy as part of patient-specific quality assurance
    Li, Heng
    Sahoo, Narayan
    Poenisch, Falk
    Suzuki, Kazumichi
    Li, Yupeng
    Li, Xiaoqiang
    Zhang, Xiaodong
    Lee, Andrew K.
    Gillin, Michael T.
    Zhu, X. Ronald
    MEDICAL PHYSICS, 2013, 40 (02)
  • [33] Improvement in SRS Commission and Patient-Specific Quality Assurance
    Ding, M.
    Heard, D.
    Ma, C.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [34] Dosimetric Effect on AI Based Virtual Log File Patient-Specific QA
    Chuang, K. C.
    Wu, Y.
    Lay, L. M.
    Cui, G.
    O'Daniel, J. C.
    Adamson, J.
    MEDICAL PHYSICS, 2024, 51 (09) : 6602 - 6602
  • [35] An Investigation for the Use of OSLDs for in Vivo Patient-Specific Quality Assurance for Patients Requiring Bolus
    Savacool, M.
    Yaparpalvi, R.
    Mynampati, D.
    Chen, Y.
    Tome, W.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [36] Quantitative Evaluation of Patient-Specific Quality Assurance Using Online Dosimetry System
    Jung, Jae-Yong
    Shin, Young-Ju
    Sohn, Seung-Chang
    Min, Jung-Whan
    Kim, Yon-Lae
    Kim, Dong-Su
    Choe, Bo-Young
    Suh, Tae-Suk
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 72 (02) : 312 - 319
  • [37] Ethos Patient-Specific Quality Assurance Outcome Prediction Using Machine Learning
    Witztum, A.
    Hink, J.
    Wall, P. D. H.
    Neylon, J.
    Solberg, T. D.
    Lamb, J. M.
    Valdes, G.
    MEDICAL PHYSICS, 2024, 51 (10) : 7706 - 7707
  • [38] Evaluation of DVH Based Patient-Specific Quality-Assurance Using COMPASS
    Sharma, S.
    Subramani, V.
    Chander, S.
    Bhaskar, S.
    Pathy, S.
    Kumar, P.
    Gopishankar, N.
    Thulkar, S.
    Singh, M.
    Binjola, A.
    MEDICAL PHYSICS, 2019, 46 (06) : E601 - E602
  • [39] Quantitative evaluation of patient-specific quality assurance using online dosimetry system
    Jae-Yong Jung
    Young-Ju Shin
    Seung-Chang Sohn
    Jung-Whan Min
    Yon-Lae Kim
    Dong-Su Kim
    Bo-Young Choe
    Tae-Suk Suh
    Journal of the Korean Physical Society, 2018, 72 : 312 - 319
  • [40] Analysis of dose comparison techniques for patient-specific quality assurance in radiation therapy
    Yu, Liting
    Tang, Timothy L. S.
    Cassim, Naasiha
    Livingstone, Alexander
    Cassidy, Darren
    Kairn, Tanya
    Crowe, Scott B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (11): : 189 - 198