An investigation of using log-file analysis for automated patient-specific quality assurance in MRgRT

被引:8
|
作者
Lim, Seng Boh [1 ]
Godoy Scripes, Paola [1 ]
Napolitano, Mary [2 ]
Subashi, Ergys [1 ]
Tyagi, Neelam [1 ]
Cervino Arriba, Laura [1 ]
Lovelock, Dale Michael [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, 1275 York Ave, New York, NY 10021 USA
[2] Standard Imaging Inc, Middleton, WI USA
来源
关键词
elekta unity; linacview; log-file analysis; MRI guided adaptive radiotherapy; MRgRT; patient specific quality assurance; MLC PERFORMANCE; IMRT; QA;
D O I
10.1002/acm2.13361
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective Adaptive radiation therapy (ART) is an integral part of MR-guided RT (MRgRT), requiring a new RT plan for each treatment fraction and resulting in a significant increase in patient-specific quality assurance (PSQA). This study investigates the possibility of using treatment log-file for automated PSQA. Method All treatment plans were delivered in 1.5T Unity MR-Linac (Elekta). A Unity compatible version of LinacView (Standard Imaging) was commissioned to automatically monitor and analyze the log-files. A total of 220 fields were delivered and measured by ArcCheck-MR (Sun Nuclear) and LinacView. Thirty incorrectly matched fields were also delivered to check for error detection sensitivity. The gamma analysis, gamma, with 3%, 3 mm criteria was used in both ArcCheck-MR and LinacView. Additionally, the gantry angle, jaws, and multileaf collimators (MLC) positions reported in the log-file were compared with plan positions using TG-142 criteria. Result The gamma (3%, 3 mm) for the 190 plans were found to be between the range of 72.5%-100.0% and 95.4%-100.0% for ArcCheck-MR and LinacVeiw, respectively. All the delivered gantry angle and jaws were found to be within 0.2 degrees and 2 mm. MLCs that were outside the guard leaves or under the diaphragms were found to have more than 1.0 mm discrepancy. This was attributed to the linac internal override for these MLCs and had no dosimetric impact. Excluding these discrepancies, all MLC positions were found to be within 1.0 mm. The gamma (3%, 3 mm) for the 30 incorrectly matched fields were found to be 3.9%-84.8% and 0.1%-64.4% for ArcCheck-MR and LinacVeiw, respectively. Conclusion Significant ranked correlation demonstrates the automated log-file analysis can be used for PSQA and expedite the ART workflow. Ongoing PSQA will be compared with log-file analysis to investigate the longer term reproducibility and correlation.
引用
收藏
页码:183 / 188
页数:6
相关论文
共 50 条
  • [21] Quick, efficient and effective patient-specific intensity-modulated radiation therapy quality assurance using log file and electronic portal imaging device
    Kumar, Rajesh
    Amols, H. I.
    Lovelock, M.
    Sharma, S. D.
    Datta, D.
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2017, 13 (02) : 297 - 303
  • [22] Postoperative analysis of surgical instrument movement using a navigation log-file
    Woerdeman, P. A.
    Willems, P. W. A.
    Noordmans, H. J.
    van der Sprenkel, J. W. Berkelbach
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2006, 1 : 304 - 307
  • [23] Machine Log File and Calibration Errors-based Patient-specific Quality Assurance (QA) for Volumetric Modulated Arc Therapy (VMAT)
    Ma, Yangguang
    Mou, Xuanqin
    Beeraka, Narasimha M.
    Guo, Yuexin
    Liu, Junqi
    Dai, Jianrong
    Fan, Ruitai
    CURRENT PHARMACEUTICAL DESIGN, 2023, 29 (34) : 2738 - 2751
  • [24] Analysis of ion beam teletherapy patient-specific quality assurance
    Liu, Xiaoli
    Deng, Yu
    Schlegel, Nicki
    Huang, Zhijie
    Moyers, Michael F.
    MEDICAL DOSIMETRY, 2019, 44 (01) : 43 - 50
  • [25] Automated Patient-specific Quality Assurance for Automated Segmentation of Organs at Risk in Nasopharyngeal Carcinoma Radiotherapy
    Wang, Yixuan
    Hu, Jiang
    Chen, Lixin
    Zhang, Dandan
    Zhu, Jinhan
    CANCER CONTROL, 2025, 32
  • [26] Investigation On the Portability of Patient-Specific Quality Assurance Between Matched Linear Accelerators
    Barraclough, B.
    Frigo, S.
    Labby, Z.
    MEDICAL PHYSICS, 2020, 47 (06) : E806 - E806
  • [27] A Method for Log File QA On An IBA Proteus System for Patient Specific Spot Scanning Quality Assurance
    Tang, S.
    Ho, M.
    Rice, I.
    Doan, D.
    Chen, C.
    Mac Rae, B.
    Mah, D.
    MEDICAL PHYSICS, 2016, 43 (06) : 3501 - 3501
  • [28] Development of a Log File Based Patient-Specific Quality Assurance for PBS Proton Therapy Using Open-Source Software Within the RayStation Treatment Planning System
    Chiang, B.
    Meyer, J.
    Wong, T.
    Maes, D.
    MEDICAL PHYSICS, 2022, 49 (06) : E375 - E375
  • [29] Statistical analysis of EPID-based VMAT patient-specific quality assurance and investigation of contributing factors
    Szwec, Stuart
    Greer, Peter
    Lee, Danny
    Barnes, Michael
    Hernandez, Victor
    Saez, Jordi
    Lam, Dao
    Sun, Baozhou
    ASIA-PACIFIC JOURNAL OF CLINICAL ONCOLOGY, 2020, 16 : 39 - 39
  • [30] Automated 3D Gamma and Spot Regression Analysis as a Component of Proton Patient-Specific Quality Assurance
    Morales, D. Hernandez
    Liu, W.
    Shan, J.
    Augustine, K.
    Davis, M.
    Bues, M.
    Stoker, J.
    MEDICAL PHYSICS, 2017, 44 (06) : 2994 - 2994