Anterior chamber angle measurement with optical coherence tomography:: Intraobserver and interobserver variability

被引:78
|
作者
Mueller, Maya
Dahmen, Gerlinde
Poerksen, Erk
Geerling, Gerd
Laqua, Horst
Ziegler, Andreas
Hoerauf, Hans
机构
[1] Univ Eye Clin, D-23538 Lubeck, Germany
[2] Inst Med Biometry & Stat, Lubeck, Germany
[3] Eye Clin, Kiel, Germany
[4] Univ Wurzburg, Eye Clin, Wurzburg, Germany
来源
关键词
D O I
10.1016/j.jcrs.2006.07.014
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To assess intraobserver and interobserver variability of anterior segment optical coherence tomography (AS-OCT) as an objective diagnostic tool to quantify the anterior chamber angle and opening width. SETTING: University Eye Clinic, Lubeck, Germany. METHODS: The anterior chamber angle and opening width were assessed in 18 eyes of 9 healthy volunteers by 2 observers. Intraobserver reproducibility was evaluated by calculating an intraclass correlation coefficient (ICC) in a mixed model. Each observer had a separate model using information from 5 scans. Interobserver variability was determined by Bland-Altman analysis. The ICC was calculated in a mixed model using a residual maximum likelihood method. The results of 3 and 5 repeated scans were evaluated to indicate the change to 1 measurement application. RESULTS: The mean anterior chamber angle measurement was 35.9 degrees +/-5.7 (SD) for observer A and 36.2 +/- 5.7 degrees for observer B. The ICC was 0.94 and 0.91, respectively. The mean opening width was 315 +/- 62 mu m for observer A and 317 +/- 60 mu m for observer B. The ICC was 0.97 and 0.93, respectively. Interobserver comparisons showed a mean difference between anterior chamber angle measurements of -0.27 +/- 1.6 degrees, a limit of agreement (LOA) interval from -3.52 to 2.98 degrees, and an ICC estimate of 0.96. The mean difference in opening width measurements was 2.40 +/- 12.40 mu m, the LOA from -27.20 to 22.40 mu m, and the estimated ICC 0.96. Using I instead of the mean of 5 measurements, the LOA range increased by 3.46 degrees for the anterior chamber angle and 30.0 mu m for the opening width. CONCLUSION: Anterior chamber angle and opening width measurements by OCT showed low intraobserver and interobserver variability, indicating OCT is a valuable technique for quantitative assessment that provides reproducible measurements and objective documentation by different examiners.
引用
收藏
页码:1803 / 1808
页数:6
相关论文
共 50 条
  • [11] Visualization of anterior chamber angle dynamics using optical coherence tomography
    Leung, CKS
    Chan, WM
    Ko, CY
    Chui, SI
    Woo, J
    Tsang, MK
    Tse, RKK
    OPHTHALMOLOGY, 2005, 112 (06) : 980 - 984
  • [12] Anterior chamber angle analysis using optical coherence tomography.
    Mori, K
    Kawasaki, S
    Naruse, S
    Kobayashi, L
    Kinoshita, S
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2000, 41 (04) : S285 - S285
  • [13] Interobserver variability in optical coherence tomography for macular
    Browning, DJ
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 137 (06) : 1116 - 1117
  • [14] Intraobserver and Interobserver Measurement Variability of the Rib-Vertebral Angle Difference
    Tysklind, Robert G.
    Myung, Karen
    Gantsoudes, George
    Kishan, Shyam
    Turner, Chad
    Lee, Stella
    Wanner, Matthew
    Sabatino, Meagan
    Linger, Scott
    Loder, Randall T.
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2018, 38 (09) : 455 - 458
  • [15] Anatomical Changes of the Anterior Chamber Angle With Anterior-Segment Optical Coherence Tomography
    Liu, Lance
    ARCHIVES OF OPHTHALMOLOGY, 2008, 126 (12) : 1682 - 1686
  • [16] Reproducibility of anterior chamber angle measurements obtained with anterior segment optical coherence tomography
    Radhakrishnan, Sunita
    See, Jovina
    Smith, Scott D.
    Nolan, Winifred P.
    Ce, Zheng
    Friedman, David S.
    Huang, David
    Li, Yan
    Aung, Tin
    Chew, Paul T. K.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (08) : 3683 - 3688
  • [17] Effect of Angle of Incidence on Anterior Chamber Angle Metrics From Optical Coherence Tomography
    Pan, Xiaojing
    Maram, Jyotsna
    Marion, Ken
    Dastiridou, Anna
    Zhang, Zhou Yuan
    Dominguez-Velasco, Maria F.
    Sadda, Srinivas R.
    Chopra, Vikas
    JOURNAL OF GLAUCOMA, 2016, 25 (01) : E19 - E23
  • [18] Effect of Hemodialysis on Anterior Chamber Angle Measured by Anterior Segment Optical Coherence Tomography
    Shin, Yong Un
    Kim, Ji Hong
    Cho, Heeyoon
    Kim, Dae Sung
    Yi, Joo-Hark
    Han, Sang-Woong
    Seong, Mincheol
    JOURNAL OF OPHTHALMOLOGY, 2019, 2019
  • [19] Effect of Angle of Incidence on Anterior Chamber Angle Metrics from Optical Coherence Tomography
    Maram, Jyotsna
    Pan, Xiaojing
    Marion, Ken
    Zhang, Zhou Y.
    Sadda, Srinivas R.
    Chopra, Vikas
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [20] Anterior chamber width measurement by high speed optical coherence tomography
    Goldsmith, JA
    Li, Y
    Chalita, MR
    Westphal, V
    Patil, CA
    Rollins, AM
    Izatt, JA
    Huang, D
    OPHTHALMOLOGY, 2005, 112 (02) : 238 - 244