Evaluations of a flat-panel based compact daily quality assurance device for proton pencil beam scanning (PBS) system

被引:10
|
作者
Su, Zhong [1 ,2 ]
Hsi, Wen [1 ,2 ]
Forthomme, Julien [3 ]
Rossomme, Severine [3 ]
机构
[1] Univ Florida, Dept Radiat Oncol, Gainesville, FL USA
[2] Univ Florida, Hlth Proton Therapy Inst, 2015 N Jefferson St, Jacksonville, FL 32206 USA
[3] IBA Dosimetry, Schwarzenbruck, Germany
关键词
Proton therapy; Pencil beam scanning (PBS); Daily quality assurance (QA); Flat-panel detector; LINEAR-ENERGY-TRANSFER; DAILY QA; THERAPY; DOSIMETER; IMRT;
D O I
10.1016/j.ejmp.2020.10.026
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the flat-panel detector quenching effect and clinical usability of a flat-panel based compact QA device for PBS daily constancy measurements. Materials & Method: The QA device, named Sphinx Compact, is composed of a 20x20 cm(2) flat-panel imager mounted on a portable frame with removable plastic modules for constancy checks of proton energy (100 MeV, 150 MeV, 200 MeV), Spread-Out-Bragg-Peak (SOBP) profile, and machine output. The potential quenching effect of the flat-panel detector was evaluated. Daily PBS QA tests of X-ray/proton isocenter coincidence, the constancy of proton spot position and sigma as well as the energy of pristine proton beam, and the flatness of SOBP proton beam through the 'transformed' profile were performed and analyzed. Furthermore, the sensitivity of detecting energy changes of pristine proton beam was also evaluated. Results: The quenching effect was observed at depths near the pristine peak regions. The flat-panel measured range of the distal 80% is within 0.9 mm to the defined ranges of the delivered proton beams. X-ray/proton isocenter coincidence tests demonstrated maximum mismatch of 0.3 mm between the two isocenters. The device can detect 0.1 mm change of spot position and 0.1 MeV energy changes of pristine proton beams. The measured transformed SOBP beam profile through the wedge module rendered as flat. Conclusions: Even though the flat-panel detector exhibited quenching effect at the Bragg peak region, the proton range can still be accurately measured. The device can fulfill the requirements of the daily QA tests recommended by the AAPM TG224 Report.
引用
收藏
页码:243 / 250
页数:8
相关论文
共 50 条
  • [21] Commissioning of the world's first compact pencil-beam scanning proton therapy system
    Pidikiti, Rajesh
    Patel, Bijal C.
    Maynard, Matthew R.
    Dugas, Joseph P.
    Syh, Joseph
    Sahoo, Narayan
    Wu, Hsinshun Terry
    Rosen, Lane R.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 94 - 105
  • [22] Development of an Assemblable, Daily, Quality-Assurance Device for Line-Scanning Proton Therapy
    Chung, Kwangzoo
    Park, Jihyun
    Ahn, Sung Hwan
    Han, Youngyih
    Lim, Do Hoon
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2019, 75 (04) : 344 - 349
  • [23] Development of an Assemblable, Daily, Quality-Assurance Device for Line-Scanning Proton Therapy
    Kwangzoo Chung
    Jihyun Park
    Sung Hwan Ahn
    Youngyih Han
    Do Hoon Lim
    [J]. Journal of the Korean Physical Society, 2019, 75 : 344 - 349
  • [24] 3D Gamma Test for Patient Specific Quality Assurance of Proton Pencil Beam Scanning Dose
    Pirelepesov, F.
    Yao, W.
    Ates, O.
    Lukose, R.
    Moskvin, V.
    Zhao, L.
    Farr, J.
    Merchant, T.
    [J]. MEDICAL PHYSICS, 2017, 44 (06)
  • [25] Impact of machine log-files uncertainties on the quality assurance of proton pencil beam scanning treatment delivery
    Toscano, S.
    Souris, K.
    Goma, C.
    Barragan-Montero, A.
    Puydupin, S.
    Vander Stappen, F.
    Janssens, G.
    Matic, A.
    Geets, X.
    Sterpin, E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (09):
  • [26] Use of a novel two-dimensional ionization chamber array for pencil beam scanning proton therapy beam quality assurance
    Lin, Liyong
    Kang, Minglei
    Solberg, Timothy D.
    Mertens, Thierry
    Baumer, Christian
    Ainsley, Christopher G.
    McDonough, James E.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 270 - 276
  • [27] A fast and reliable method for daily quality assurance in spot scanning proton therapy with a compact and inexpensive phantom
    Bizzocchi, Nicola
    Fracchiolla, Francesco
    Schwarz, Marco
    Algranati, Carlo
    [J]. MEDICAL DOSIMETRY, 2017, 42 (03) : 238 - 246
  • [28] A multi-layer strip ionization chamber (MLSIC) device for proton pencil beam scan quality assurance
    Zhou, Shuang
    Rao, Wei
    Chen, Qinghao
    Tan, Yuewen
    Smith, Winter
    Sun, Baozhou
    Zhou, Jun
    Chang, Chih-Wei
    Lin, Liyong
    Darafsheh, Arash
    Zhao, Tianyu
    Zhang, Tiezhi
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (17):
  • [29] Log File-Based Patient Specific Quality Assurance: 5 Years of Experience in Pencil Beam Scanning Pediatric Proton Therapy
    Ates, O.
    Pirlepesov, F.
    Hua, C.
    Merchant, T. E.
    Zhao, L.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [30] Characterization of penumbra sharpening and scattering by adaptive aperture for a compact pencil beam scanning proton therapy system
    Grewal, Hardev S.
    Ahmad, Salahuddin
    Jin, Hosang
    [J]. MEDICAL PHYSICS, 2021, 48 (04) : 1508 - 1519