Impact of machine log-files uncertainties on the quality assurance of proton pencil beam scanning treatment delivery

被引:17
|
作者
Toscano, S. [1 ,6 ]
Souris, K. [2 ]
Goma, C. [1 ,3 ]
Barragan-Montero, A. [2 ]
Puydupin, S. [4 ]
Vander Stappen, F. [4 ]
Janssens, G. [4 ]
Matic, A. [4 ]
Geets, X. [5 ]
Sterpin, E. [1 ,2 ]
机构
[1] Katholieke Univ Leuven, Dept Oncol, Lab Expt Radiotherapy, Leuven, Belgium
[2] Catholic Univ Louvain, Inst Rech Expt & Clin, Ctr Mol Imaging Radiotherapy & Oncol, Brussels, Belgium
[3] Univ Hosp Leuven, Dept Radiat Oncol, Leuven, Belgium
[4] Ion Beam Applicat, Louvain La Neuve, Belgium
[5] Clin Univ St Luc, Dept Radiat Oncol, Brussels, Belgium
[6] Univ Libre Bruxelles, Ave Franklin Roosevelt 50, B-1050 Brussels, Belgium
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2019年 / 64卷 / 09期
关键词
proton therapy; Monte Carlo simulation; machine log-files; quality assurance; RANGE UNCERTAINTIES; THERAPY; SENSITIVITY; DOSIMETRY;
D O I
10.1088/1361-6560/ab120c
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Irradiation log-files store useful information about the plan delivery, and together with independent Monte Carlo dose engine calculations can be used to reduce the time needed for patient-specific quality assurance (PSQA). Nonetheless, machine log-files carry an uncertainty associated to the measurement of the spot position and intensity that can influence the correct evaluation of the quality of the treatment delivery. This work addresses the problem of the inclusion of these uncertainties for the final verification of the treatment delivery. Dedicated measurements performed in an IBA Proteus Plus gantry with a pencil beam scanning (PBS) dedicated nozzle have been carried out to build a 'room-dependent' model of the spot position uncertainties. The model has been obtained through interpolation of the look-up tables describing the systematic and random uncertainties, and it has been tested for a clinical case of a brain cancer patient irradiated in a dry-run. The delivered dose has been compared with the planned dose with the inclusion of the errors obtained applying the model. Our results suggest that the accuracy of the treatment delivery is higher than the spot position uncertainties obtained from the log-file records. The comparison in terms of DVHs shows that the log-reconstructed dose is compatible with the planned dose within the 95% confidence interval obtained applying our model. The initial mean dose difference between the calculated dose to the patient based on the plan and recorded data is around 1%. The difference is essentially due to the log-file uncertainties and it can be removed with a correct treatment of these errors. In conclusion our new PSQA protocol allows for a fast verification of the dose delivered after every treatment fraction through the use of machine log-files and an independent Monte Carlo dose engine. Moreover, the inclusion of log-file uncertainties in the dose calculation allows for a correct evaluation of the quality of the treatment plan delivery.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] A machine learning-based framework for delivery error prediction in proton pencil beam scanning using irradiation log-files
    Maes, Dominic
    Bowen, Stephen R.
    Regmi, Rajesh
    Bloch, Charles
    Wong, Tony
    Rosenfeld, Anatoly
    Saini, Jatinder
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 78 : 179 - 186
  • [2] Machine Log File Analysis for Quality Assurance of Proton Pencil Beam Scanning System
    Zhang, R.
    Clasie, B.
    Depauw, N.
    Lin, Y.
    Shen, T.
    Flanz, J.
    Lu, H.
    Jee, K.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E264 - E264
  • [3] Machine Learning Based Log File Data Analysis for Quality Assurance of Proton Pencil Beam Scanning
    Dou, T.
    Clasie, B.
    Depauw, N.
    Shen, T.
    Flanz, J.
    Lu, H.
    Jee, K.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E603 - E603
  • [4] Using Machine Log-Files for Treatment Planning and Delivery QA
    Stanhope, C.
    Liang, J.
    Drake, D.
    Yan, D.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3852 - 3852
  • [5] Detectors for quality assurance of pencil beam scanning gantries for proton therapy
    Gagnon-Moisan, F.
    van der Meer, R.
    Chowdhuri, Z.
    Eichin, M.
    Koenig, S.
    Schmid, F. -J.
    Weber, D. C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S44 - S45
  • [6] FLASH dose rate calculation based on log files in proton pencil beam scanning therapy
    Jeon, Chanil
    Ahn, Sunghwan
    Amano, Daizo
    Kamiguchi, Nagaaki
    Cho, Sungkoo
    Sheen, Heesoon
    Park, Hee Chul
    Han, Youngyih
    [J]. MEDICAL PHYSICS, 2023, 50 (11) : 7154 - 7166
  • [7] First Quality Assurance Implementation of Treatment Planning System for Pencil Beam Scanning Proton Radiotherapy
    Xiong, W.
    Yu, G.
    Kang, M.
    Huang, S.
    Lin, H.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [8] Applications of a novel detector for pencil beam scanning proton therapy beam quality assurance
    Yang, Pei-Ying
    Hsieh, Yang-Wei
    Kang, Chen-Lin
    Tseng, Chin-Dar
    Lin, Chih-Hsueh
    Shieh, Chin-Shiuh
    Chao, Pei-Ju
    Lee, Tsair-Fwu
    [J]. INTERNATIONAL JOURNAL OF MODERN PHYSICS B, 2021, 35 (14N16):
  • [9] Plan Quality Assurance of VMAT Using Fluence Images Reconstituted From Log-Files
    Katsuta, Y.
    Shimizu, E.
    Matsunaga, K.
    Majima, K.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 283 - 284
  • [10] Improved 4D proton dosimetry via correlation with beam delivery details using log-files
    Kostiukhina, N.
    Palmans, H.
    Waid, S.
    Stock, M.
    Georg, D.
    Knaeusl, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S533 - S534