Evaluation of a hip joint simulator

被引:14
|
作者
Smith, SL [1 ]
Burgess, IC [1 ]
Unsworth, A [1 ]
机构
[1] Univ Durham, Sch Engn, Ctr Biomed Engn, Sci Labs, Durham DH1 3LE, England
关键词
wear; hip simulators; PTFE; UHMWPE;
D O I
10.1243/0954411991535077
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To evaluate the functioning of the Durham hip joint wear simulator, the wear rates of ultra high molecular weight polyethylene (UHMWPE) and polytetrafluoroethylene (PTFE) acetabular cups articulating against 22 mm diameter cobalt-chromium-molybdenum (CoCrMo) femoral heads were studied. A wear test was conducted in a lubricant of distilled water at 37 degrees C for a duration of 4.8 million cycles. The average penetration rate for the CoCrMo femoral heads against UHMWPE acetabular cups was 0.03 mm/10(6) cycles, while penetration rate for PTFE cups was some twenty times greater. These results are of a similar order of magnitude to other simulator studies in distilled water and are in a similar ratio to clinical data.
引用
收藏
页码:469 / 473
页数:5
相关论文
共 50 条
  • [1] Discussion on the design of a hip joint simulator
    Viceconti, M
    Cavallotti, G
    Andrisano, AO
    Toni, A
    MEDICAL ENGINEERING & PHYSICS, 1996, 18 (03) : 234 - 240
  • [2] NEW JOINT SIMULATOR FOR TOTAL HIP-JOINT PROSTHESIS
    SASADA, T
    MABUCHI, K
    MAEZAWA, H
    JOURNAL OF JAPAN SOCIETY OF LUBRICATION ENGINEERS, 1976, 21 (03): : 184 - 189
  • [3] A five-station hip joint simulator
    Smith, SL
    Unsworth, A
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2001, 215 (H1) : 61 - 64
  • [4] Wear characterization of clinically used hip joint prostheses by a hip simulator
    Takigawa, Y
    Jono, K
    Takadama, H
    Mizuno, M
    Nakamura, T
    27TH INTERNATIONAL COCOA BEACH CONFERENCE ON ADVANCED CERAMICS AND COMPOSITES: A, 2003, 24 (03): : 197 - 202
  • [5] Design of a Hip Joint Simulator According to the ISO 14242
    Galanis, Nikolaos I.
    Manolakos, Dimitrios E.
    WORLD CONGRESS ON ENGINEERING, WCE 2011, VOL III, 2011, : 2088 - 2093
  • [6] CHOICE OF A RATIONAL HIP JOINT WEAR SIMULATOR DESIGN
    Pakhaliuk, Volodymyr
    Pashkov, Yevgen
    Desyatov, Ivan
    Manchuk, Vitaly
    ANNALS OF DAAAM FOR 2009 & PROCEEDINGS OF THE 20TH INTERNATIONAL DAAAM SYMPOSIUM, 2009, 20 : 1559 - 1560
  • [7] Computational hip joint simulator for wear and heat generation
    Fialho, Jorge C.
    Fernandes, Paulo R.
    Eca, Luis
    Folgado, Joao
    JOURNAL OF BIOMECHANICS, 2007, 40 (11) : 2358 - 2366
  • [8] A 12-station anatomic hip joint simulator
    Saikko, V
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2005, 219 (H6) : 437 - 448
  • [9] A triaxial physiological hip joint simulator for the wear testing of hip joint prostheses - Drive and control principles
    Springstein, D
    Wendt, Z
    Boenick, U
    BIOMEDIZINISCHE TECHNIK, 2003, 48 (04): : 106 - 109
  • [10] Friction Measurement in the Biaxial Rocking Motion Hip Joint Simulator
    Saikko, Vesa
    JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 2009, 131 (01): : 1 - 8