Computational hip joint simulator for wear and heat generation

被引:75
|
作者
Fialho, Jorge C.
Fernandes, Paulo R. [1 ]
Eca, Luis
Folgado, Joao
机构
[1] IDMEC IST, Lisbon, Portugal
[2] DEM IST, Lisbon, Portugal
关键词
hip joint; acetabular wear; heat generation; contact; finite elements;
D O I
10.1016/j.jbiomech.2006.12.005
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper presents a computational simulator for the hip to compute the wear and heat generation on artificial joints. The friction produced on artificial hip joints originates wear rates that can lead to failure of the implant. Furthermore, the frictional heating can increase the wear. The developed computational model calculates the wear in the joint and the temperature in the surrounding zone, allowing the use of different combinations of joint materials, daily activities and different individuals. The pressure distribution on the joint bearing surfaces is obtained with the solution of a contact model. The heat generation by friction and the volumetric wear is computed from the pressure distribution and the sliding distance. The temperature is obtained from the solution of a transient heat conduction problem that includes the time-dependent heat generated by friction. The contact and heat conduction problems are solved numerically with the Finite Element Method. The developed computational model performs a full simulation of the acetabular bearing surface behaviour, which is useful for acetabular cup design and material selection. The results obtained by the present model agree with experimental and clinical data, as well as other numerical studies. (c) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2358 / 2366
页数:9
相关论文
共 50 条
  • [1] CHOICE OF A RATIONAL HIP JOINT WEAR SIMULATOR DESIGN
    Pakhaliuk, Volodymyr
    Pashkov, Yevgen
    Desyatov, Ivan
    Manchuk, Vitaly
    ANNALS OF DAAAM FOR 2009 & PROCEEDINGS OF THE 20TH INTERNATIONAL DAAAM SYMPOSIUM, 2009, 20 : 1559 - 1560
  • [2] Wear characterization of clinically used hip joint prostheses by a hip simulator
    Takigawa, Y
    Jono, K
    Takadama, H
    Mizuno, M
    Nakamura, T
    27TH INTERNATIONAL COCOA BEACH CONFERENCE ON ADVANCED CERAMICS AND COMPOSITES: A, 2003, 24 (03): : 197 - 202
  • [3] A triaxial physiological hip joint simulator for the wear testing of hip joint prostheses - Drive and control principles
    Springstein, D
    Wendt, Z
    Boenick, U
    BIOMEDIZINISCHE TECHNIK, 2003, 48 (04): : 106 - 109
  • [4] A simulated synovial fluid for wear characterization of artificial hip joints by a hip joint simulator
    Takadama, H
    Mizuno, M
    BIOCERAMICS 18, PTS 1 AND 2, 2006, 309-311 : 1273 - 1276
  • [5] Frictional heating of bearing materials tested in a hip joint wear simulator
    Lu, Z
    McKellop, H
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (01) : 101 - 108
  • [6] Wear tests in a hip joint simulator of different CoCrMo counterfaces on UHMWPE
    Gonzalez-Mora, V. A.
    Hoffmann, M.
    Stroosnijder, R.
    Gil, F. J.
    MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC AND SUPRAMOLECULAR SYSTEMS, 2009, 29 (01): : 153 - 158
  • [7] The biomaterials challenge: A comparison of polyethylene wear using a hip joint simulator
    Affatato, Sauerio
    Freccero, Nadia
    Taddei, Paola
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 53 : 40 - 48
  • [8] A multi-station hip joint simulator study and wear characterization of commercial hip endoprostheses
    Jono, K
    Takigawa, Y
    Takadama, H
    Mizuno, M
    Nakamura, T
    27TH INTERNATIONAL COCOA BEACH CONFERENCE ON ADVANCED CERAMICS AND COMPOSITES: A, 2003, 24 (03): : 255 - 260
  • [9] Three-axis hip joint simulator for wear and friction studies on total hip prostheses
    Saikko, Vesa O.
    Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996, 210 (03) : 175 - 185
  • [10] Quantitative analysis of wear and wear debris from metal-on-metal hip prostheses tested In a physiological hip joint simulator
    Firkins, PJ
    Tipper, JL
    Saadatzadeh, MR
    Ingham, E
    Stone, MH
    Farrar, R
    Fisher, J
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2001, 11 (02) : 143 - 157