The biomaterials challenge: A comparison of polyethylene wear using a hip joint simulator

被引:39
|
作者
Affatato, Sauerio [1 ]
Freccero, Nadia [1 ]
Taddei, Paola [2 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, Bologna, Italy
[2] Univ Bologna, Dipartimento Sci Biomed & Neuromotorie, I-40126 Bologna, Italy
关键词
Vitamin-E doped polyethylene; Cross-linked polyethylene; Ultra-high molecular weight polyethylene; Hip simulator; Raman spectroscopy; MOLECULAR-WEIGHT POLYETHYLENE; CROSS-LINKED UHMWPE; VITAMIN-E; IN-VITRO; MECHANICAL-PROPERTIES; RAMAN-SPECTROSCOPY; MICRO-RAMAN; CUPS; IRRADIATION; RESISTANCE;
D O I
10.1016/j.jmbbm.2015.08.001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Although hip arthroplasty is an established procedure that relieves pain and improves functions, problems remain with wear and osteolysis. Highly cross-linked polyethylene and Vitamin-E-stabilized polyethylene were introduced in the last years to solve these problems. In this study we compared the in vitro wear behaviour of cross-linked polyethylene (XLPE) versus Vitamin-E diffused XLPE (XLPE_VE) versus conventional ultra-high molecular weight polyethylene (UHMWPE) acetabular cups. The test was performed using a hip joint simulator run for two millions cycles under bovine calf serum as lubricant. Mass loss was found to decrease along the series UHMWPE>XLPE_VE>XLPE, although statistically significant differences were found only between the mass losses of XLPE and UHMWPE at 1.2 and 2 million cycles. The mass loss data were explained in relation to the crystalline morphology of the control unworn cups, as investigated by non-destructive micro-Raman spectroscopy. This technique allowed to disclose a different wear behaviour of the three sets of cups. Wear testing produced a stress-induced crystallisation in UHMWPE, with increases in both amorphous (alpha(a)) and orthorhombic (alpha(o)) phases at the expense of the third phase (alpha(b)), which decreased upon wear. Moreover, the all-trans content decreased, while the ortho-trans content increased, contrarily to the trend observed for XLPE and XLPE_VE, for which no statistically significant changes in alpha(o), alpha(a) and alpha(b) contents were detected. The XLPE_VE specimens underwent the least significant changes in the spectroscopic markers of micromorphology upon mechanical stress, probably due to their lower starting amorphous content. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:40 / 48
页数:9
相关论文
共 50 条
  • [1] Wear comparison between a dual mobility total hip prosthesis and a typical modular design using a hip joint simulator
    Saikko, Vesa
    Shen, Ming
    WEAR, 2010, 268 (3-4) : 617 - 621
  • [2] Influence of hip joint simulator design and mechanics on the wear and creep of metal-on-polyethylene bearings
    Ali, Murat
    Al-Hajjar, Mazen
    Partridge, Susan
    Williams, Sophie
    Fisher, John
    Jennings, Louise M.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (05) : 389 - 397
  • [3] Wear behavior of γ-irradiated ultra high-molecular weight polyethylene in a hip joint simulator
    Raffi, N. Mohamad
    Kanagarajan, D.
    Srinivasan, V.
    JOURNAL OF THERMOPLASTIC COMPOSITE MATERIALS, 2016, 29 (02) : 249 - 259
  • [4] Aggressive 3rd-body wear challenge to highly crosslinked polyethylene: A hip simulator model
    Kubo, K.
    Clarke, I. C.
    Sorimachi, T.
    Williams, P. A.
    Donaldson, T. K.
    Yamamoto, K.
    WEAR, 2009, 267 (5-8) : 734 - 742
  • [5] Polyethylene cross-linking by two different methods reduces acetabular liner wear in a hip joint wear simulator
    D'Lima, DD
    Hermida, JC
    Chen, PC
    Colwell, CW
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2003, 21 (05) : 761 - 766
  • [6] CHOICE OF A RATIONAL HIP JOINT WEAR SIMULATOR DESIGN
    Pakhaliuk, Volodymyr
    Pashkov, Yevgen
    Desyatov, Ivan
    Manchuk, Vitaly
    ANNALS OF DAAAM FOR 2009 & PROCEEDINGS OF THE 20TH INTERNATIONAL DAAAM SYMPOSIUM, 2009, 20 : 1559 - 1560
  • [7] Computational hip joint simulator for wear and heat generation
    Fialho, Jorge C.
    Fernandes, Paulo R.
    Eca, Luis
    Folgado, Joao
    JOURNAL OF BIOMECHANICS, 2007, 40 (11) : 2358 - 2366
  • [8] Wear characterization of clinically used hip joint prostheses by a hip simulator
    Takigawa, Y
    Jono, K
    Takadama, H
    Mizuno, M
    Nakamura, T
    27TH INTERNATIONAL COCOA BEACH CONFERENCE ON ADVANCED CERAMICS AND COMPOSITES: A, 2003, 24 (03): : 197 - 202
  • [9] WEAR OF POLYETHYLENE ACETABULAR CUPS AGAINST ZIRCONIA FEMORAL HEADS STUDIED WITH A HIP-JOINT SIMULATOR
    SAIKKO, V
    WEAR, 1994, 176 (02) : 207 - 212
  • [10] Effect of head size on wear properties of metal-on-metal bearings of hip prostheses, and comparison with wear properties of metal-on-polyethylene bearings using hip simulator
    Okazaki, Yoshimitsu
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 31 : 152 - 163