The biomaterials challenge: A comparison of polyethylene wear using a hip joint simulator

被引:39
|
作者
Affatato, Sauerio [1 ]
Freccero, Nadia [1 ]
Taddei, Paola [2 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, Bologna, Italy
[2] Univ Bologna, Dipartimento Sci Biomed & Neuromotorie, I-40126 Bologna, Italy
关键词
Vitamin-E doped polyethylene; Cross-linked polyethylene; Ultra-high molecular weight polyethylene; Hip simulator; Raman spectroscopy; MOLECULAR-WEIGHT POLYETHYLENE; CROSS-LINKED UHMWPE; VITAMIN-E; IN-VITRO; MECHANICAL-PROPERTIES; RAMAN-SPECTROSCOPY; MICRO-RAMAN; CUPS; IRRADIATION; RESISTANCE;
D O I
10.1016/j.jmbbm.2015.08.001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Although hip arthroplasty is an established procedure that relieves pain and improves functions, problems remain with wear and osteolysis. Highly cross-linked polyethylene and Vitamin-E-stabilized polyethylene were introduced in the last years to solve these problems. In this study we compared the in vitro wear behaviour of cross-linked polyethylene (XLPE) versus Vitamin-E diffused XLPE (XLPE_VE) versus conventional ultra-high molecular weight polyethylene (UHMWPE) acetabular cups. The test was performed using a hip joint simulator run for two millions cycles under bovine calf serum as lubricant. Mass loss was found to decrease along the series UHMWPE>XLPE_VE>XLPE, although statistically significant differences were found only between the mass losses of XLPE and UHMWPE at 1.2 and 2 million cycles. The mass loss data were explained in relation to the crystalline morphology of the control unworn cups, as investigated by non-destructive micro-Raman spectroscopy. This technique allowed to disclose a different wear behaviour of the three sets of cups. Wear testing produced a stress-induced crystallisation in UHMWPE, with increases in both amorphous (alpha(a)) and orthorhombic (alpha(o)) phases at the expense of the third phase (alpha(b)), which decreased upon wear. Moreover, the all-trans content decreased, while the ortho-trans content increased, contrarily to the trend observed for XLPE and XLPE_VE, for which no statistically significant changes in alpha(o), alpha(a) and alpha(b) contents were detected. The XLPE_VE specimens underwent the least significant changes in the spectroscopic markers of micromorphology upon mechanical stress, probably due to their lower starting amorphous content. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:40 / 48
页数:9
相关论文
共 50 条
  • [41] Friction measurement in a hip wear simulator
    Saikko, Vesa
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (05) : 366 - 372
  • [42] Tribological behavior of ultra-high molecular weight polyethylene in a hip joint simulator
    Mohamad Raffi, N.
    Kanagarajan, D.
    Srinivasan, V.
    FRONTIERS OF MATERIALS SCIENCE, 2012, 6 (04) : 358 - 365
  • [43] Tribological behavior of ultra-high molecular weight polyethylene in a hip joint simulator
    N. Mohamad Raffi
    D. Kanagarajan
    V. Srinivasan
    Frontiers of Materials Science, 2012, 6 : 358 - 365
  • [44] Effects of the sterilisation method on the wear of UHMWPE acetabular cups tested in a hip joint simulator
    Affatato, S
    Bordini, B
    Fagnano, C
    Taddei, P
    Tinti, A
    Toni, A
    BIOMATERIALS, 2002, 23 (06) : 1439 - 1446
  • [45] A comparative tribological study of the wear of composite cushion cups in a physiological hip joint simulator
    Dept. Mechanical Engineering, University of Leeds, L52 9JT, Leeds, United Kingdom
    不详
    J. Biomech., 4 (363-369):
  • [46] Effect of random variation of input and various daily activities on wear in a hip joint simulator
    Viitala, Raine
    Saikko, Vesa
    JOURNAL OF BIOMECHANICS, 2020, 106
  • [47] Hip-simulator ranking of polyethylene wear - Comparisons between ceramic heads of different sizes
    Clarke, IC
    Gustafson, A
    Jung, H
    Fujisawa, A
    ACTA ORTHOPAEDICA SCANDINAVICA, 1996, 67 (02): : 128 - 132
  • [48] Evaluation of Two Total Hip Bearing Materials for Resistance to Wear Using a Hip Simulator
    St John, Kenneth R.
    LUBRICANTS, 2015, 3 (02): : 459 - 474
  • [49] A comparative tribological study of the wear of composite cushion cups in a physiological hip joint simulator
    Bigsby, RJA
    Auger, DD
    Jin, ZM
    Dowson, D
    Hardaker, CS
    Fisher, J
    JOURNAL OF BIOMECHANICS, 1998, 31 (04) : 363 - 369
  • [50] An easy technique to digest and isolate UHMWPE wear particles from a hip joint simulator
    Affatato, S
    Emiliani, D
    Bersaglia, G
    Traina, F
    Toni, A
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2004, 27 (05): : 424 - 432