Evaluation of a hip joint simulator

被引:14
|
作者
Smith, SL [1 ]
Burgess, IC [1 ]
Unsworth, A [1 ]
机构
[1] Univ Durham, Sch Engn, Ctr Biomed Engn, Sci Labs, Durham DH1 3LE, England
关键词
wear; hip simulators; PTFE; UHMWPE;
D O I
10.1243/0954411991535077
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To evaluate the functioning of the Durham hip joint wear simulator, the wear rates of ultra high molecular weight polyethylene (UHMWPE) and polytetrafluoroethylene (PTFE) acetabular cups articulating against 22 mm diameter cobalt-chromium-molybdenum (CoCrMo) femoral heads were studied. A wear test was conducted in a lubricant of distilled water at 37 degrees C for a duration of 4.8 million cycles. The average penetration rate for the CoCrMo femoral heads against UHMWPE acetabular cups was 0.03 mm/10(6) cycles, while penetration rate for PTFE cups was some twenty times greater. These results are of a similar order of magnitude to other simulator studies in distilled water and are in a similar ratio to clinical data.
引用
收藏
页码:469 / 473
页数:5
相关论文
共 50 条
  • [21] The clinical relevance of hip joint simulator testing: In vitro and in vivo comparisons
    Essner, A
    Schmidig, G
    Wang, A
    WEAR, 2005, 259 (259) : 882 - 886
  • [22] Multichannel Raman microspectrometry of UHMWPE hip cups run on joint simulator
    Bertoluzza, A
    Fagnano, C
    Rossi, M
    Tinti, A
    SPECTROSCOPY OF BIOLOGICAL MOLECULES: NEW DIRECTIONS, 1999, : 519 - 520
  • [23] 2ND GENERATION OF MUNICH HIP-JOINT-SIMULATOR
    UNGETHUM, M
    HINTERBERGER, J
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1978, 91 (03) : 233 - 237
  • [24] Frictional heating of bearing materials tested in a hip joint wear simulator
    Lu, Z
    McKellop, H
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (01) : 101 - 108
  • [25] A preliminary hip joint simulator study of the migration of a cemented femoral stem
    Liu, C
    Green, SM
    Watkins, ND
    Gregg, PJ
    McCaskie, AW
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2003, 217 (H2) : 127 - 135
  • [26] Wear tests in a hip joint simulator of different CoCrMo counterfaces on UHMWPE
    Gonzalez-Mora, V. A.
    Hoffmann, M.
    Stroosnijder, R.
    Gil, F. J.
    MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC AND SUPRAMOLECULAR SYSTEMS, 2009, 29 (01): : 153 - 158
  • [27] Friction Compensation in Trajectory Tracking Control for a Parallel Hip Joint Simulator
    Shan, Xianlei
    Cheng, Gang
    Chen, Xihui
    MECHANISM AND MACHINE SCIENCE, 2017, 408 : 143 - 153
  • [28] The biomaterials challenge: A comparison of polyethylene wear using a hip joint simulator
    Affatato, Sauerio
    Freccero, Nadia
    Taddei, Paola
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 53 : 40 - 48
  • [29] Biotribocorrosion: Some electrochemical observations from an instrumented hip joint simulator
    Hesketh, James
    Hu, Xinming
    Yan, Yu
    Dowson, Duncan
    Neville, Anne
    TRIBOLOGY INTERNATIONAL, 2013, 59 : 332 - 338