摘要
To evaluate the functioning of the Durham hip joint wear simulator, the wear rates of ultra high molecular weight polyethylene (UHMWPE) and polytetrafluoroethylene (PTFE) acetabular cups articulating against 22 mm diameter cobalt-chromium-molybdenum (CoCrMo) femoral heads were studied. A wear test was conducted in a lubricant of distilled water at 37 degrees C for a duration of 4.8 million cycles. The average penetration rate for the CoCrMo femoral heads against UHMWPE acetabular cups was 0.03 mm/10(6) cycles, while penetration rate for PTFE cups was some twenty times greater. These results are of a similar order of magnitude to other simulator studies in distilled water and are in a similar ratio to clinical data.
引用
收藏
页码:469 / 473
页数:5
相关论文
共 50 条
  • [31] CLINICAL EVALUATION OF HIP-JOINT
    GERSHUNI, DH
    ISRAEL JOURNAL OF MEDICAL SCIENCES, 1974, 10 (06): : 659 - 665
  • [32] Study on Biotribological Behavior of the Combined Joint of CoCrMo and UHMWPE/BHA Composite in a Hip Joint Simulator
    Qingliang Wang
    Jinlong Liu
    Shirong Ge
    Journal of Bionic Engineering, 2009, 6 : 378 - 386
  • [34] Evaluation of a joint distractor to facilitate arthroscopy of the hip joint in dogs
    Devesa, V.
    Rovesti, G. L.
    Urrutia, P. G.
    San Roman, F.
    Rodriguez-Quiros, J.
    JOURNAL OF SMALL ANIMAL PRACTICE, 2014, 55 (12) : 603 - 608
  • [35] Evaluation of Two Total Hip Bearing Materials for Resistance to Wear Using a Hip Simulator
    St John, Kenneth R.
    LUBRICANTS, 2015, 3 (02): : 459 - 474
  • [36] Development of a ten-station, multi-axis hip joint simulator
    Goldsmith, AAJ
    Dowson, D
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H4) : 311 - 316
  • [37] Simplified motion and loading compared to physiological motion and loading in a hip joint simulator
    Smith, SL
    Unsworth, A
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H3) : 233 - 238
  • [38] Design and manufacturing of a hip joint motion simulator with a novel modular design approach
    Torabnia, Shams
    Mihcin, Senay
    Lazoglu, Ismail
    INTERNATIONAL JOURNAL OF INTERACTIVE DESIGN AND MANUFACTURING - IJIDEM, 2024, 18 (01): : 401 - 417
  • [39] Configuration and singularity analysis of a parallel hip joint simulator based on the forward kinematics
    Cheng, Gang
    Li, Yang
    Lodewijks, Gabriel
    Pang, Yusong
    Shan, Xianlei
    APPLIED MATHEMATICAL MODELLING, 2016, 40 (15-16) : 7281 - 7292
  • [40] Design and manufacturing of a hip joint motion simulator with a novel modular design approach
    Shams Torabnia
    Senay Mihcin
    Ismail Lazoglu
    International Journal on Interactive Design and Manufacturing (IJIDeM), 2024, 18 : 401 - 417