Feasibility study of using Stereotactic Field Diode for field output factors measurement and evaluating three new detectors for small field relative dosimetry of 6 and 10 MV photon beams

被引:9
|
作者
Gul, Attia [1 ,2 ]
Fukuda, Shigekazu [1 ]
Mizuno, Hideyuki [1 ]
Taku, Nakaji [1 ]
Kakakhel, M. Basim [2 ]
Mirza, Sikander M. [2 ]
机构
[1] QST Hosp, Natl Inst Quantum & Radiol Sci & Technol, Chiba, Japan
[2] Pakistan Inst Engn & Appl Sci PIEAS, Dept Phys & Appl Math, Islamabad, Pakistan
来源
关键词
correction factors; field output factors; PTW31023; RAZOR diode; RAZOR nano chamber; small field dosimetry; stereotactic field diode; RADIOSURGERY BEAMS; SILICON DIODE; PROTOTYPE; CHAMBER;
D O I
10.1002/acm2.13007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study assesses the feasibility of using stereotactic field diode (SFD) as an alternate to gaf chromic films for field output factor (FF) measurement and further evaluating three new detectors for small field dosimetry. Varian 21EX linear accelerator was used to generate 6 and 10 MV beams of nominal square fields ranging from 0.5 x 0.5 cm(2)to 10 x 10 cm(2). One passive (EBT3 films) and five active detectors including IBA RAZOR diode(RD), SFD, RAZOR nanochamber (RNC), pinpoint chamber (PTW31023), and semiflex chamber (PTW31010) were employed. FFs were measured using films and SFD while beam profiles and percentage depth dose (PDD) distribution were acquired with active detectors. Polarity (k(pol)) and recombination (k(s)) effects of ion chambers were determined and corrected for output ratio measurement. Correction factors (CF) of RD, RNC, and PTW31023 in axial and radial orientation were also measured. Stereotactic field diode measured FFs have shown good agreement with films (with difference of <1%). RD and RNC measured beam profiles were within 3% deviation from the SFD values. Variation ink(pol)with field size for RNC and PTW31023 was up to 4% and 0.4% (for fields >= 1 x 1 cm(2)), respectively, while variation ink(s)of PTW31023 was <0.2 %. The maximum values of CF have been calculated to be 5.2%, 2.0%, 13.6%, and 25.5% for RD, RNC, PTW31023-axial, and PTW31023-radial respectively. This study concludes that SFD with appropriate CFs as given in TRS 483 may be used for measuring FFs as an alternate to EBT3 films. Whereas RD and RNC may be used for beam profile and PDD measurement in small fields. Considering the limit of usability of 2%, RNC may be used without CF for FF measurement in the smallfields investigated in this study.
引用
收藏
页码:23 / 36
页数:14
相关论文
共 33 条
  • [21] A novel extrapolation method using OSL detectors for very small field output factor measurement for stereotactic radiosurgery
    Huang, Te-An Polly
    Morales, Johnny E.
    Butson, Ethan
    Johnson, Annie
    Butson, Martin
    Hill, Robin
    [J]. PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (02) : 593 - 599
  • [22] A novel extrapolation method using OSL detectors for very small field output factor measurement for stereotactic radiosurgery
    Te-An Polly Huang
    Johnny E. Morales
    Ethan Butson
    Annie Johnson
    Martin Butson
    Robin Hill
    [J]. Physical and Engineering Sciences in Medicine, 2020, 43 : 593 - 599
  • [23] Properties of IBA Razor Nano Chamber in small-field radiation therapy using 6MV FF, 6 MV FFF, and 10 MV FFF photon beams
    Partanen, Mari
    Niemela, Jarkko
    Ojala, Jarkko
    Keyrilainen, Jani
    Kapanen, Mika
    [J]. ACTA ONCOLOGICA, 2021, 60 (11) : 1419 - 1424
  • [24] Determination of the small-field output factor for 6 MV photon beam using EGSnrc Monte Carlo
    Chuah, K. W.
    Aziz, Abdul M. Z.
    Jayamani, J.
    [J]. JOURNAL OF MEDICAL PHYSICS, 2022, 47 (03) : 301 - 308
  • [25] Investigation of field output factors using IAEA-AAPM TRS-483 code of practice recommendations and Monte Carlo simulation for 6 MV photon beams
    Yabsantia, Sumalee
    Suriyapee, Sivalee
    Phaisangittisakul, Nakorn
    Oonsiri, Sornjarod
    Sanghangthum, Taweap
    Seuntjens, Jan
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 22
  • [26] The effect of SSD, Field size, Energy and Detector type for Relative Output Factor measurement in small photon beams as compared with Monte Carlo simulation
    Setilo, Itumeleng
    Oderinde, Oluwaseyi Michael
    du Plessis, Freek C. P.
    [J]. POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2019, 25 (02): : 101 - 110
  • [27] 6-MV small field output factors: intra-/intermachine comparison and implementation of TRS-483 using various detectors and several linear accelerators
    Ghazal, Mohammed
    Westermark, Mathias
    Kaveckyte, Vaiva
    Carlsson-Tedgren, Asa
    Benmakhlouf, Hamza
    [J]. MEDICAL PHYSICS, 2019, 46 (11) : 5350 - 5359
  • [28] Small field detector correction factors kQclin,Qmsrfclin,fmsr for silicon-diode and diamond detectors with circular 6 MV fields derived using both empirical and numerical methods
    O'Brien, D. J.
    Leon-Vintro, L.
    McClean, B.
    [J]. MEDICAL PHYSICS, 2016, 43 (01) : 411 - 423
  • [29] Technical Note: Multicenter study of TrueBeam FFF beams with a new stereotactic diode: Can a common small field signal ratio curve be defined?
    Cagni, Elisabetta
    Russo, Serenella
    Reggiori, Giacomo
    Bresciani, Sara
    Fedele, David
    Iori, Mauro
    Marino, Carmelo
    Nardiello, Barbara
    Ruggieri, Ruggero
    Strigari, Lidia
    Mancosu, Pietro
    [J]. MEDICAL PHYSICS, 2016, 43 (10) : 5570 - 5576
  • [30] Small field output correction factors for 6-X and 6-X FFF beams: GAMOS Monte-Carlo study
    Akcay, D.
    Kandemir, R.
    Azaklioglu, O.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S694 - S695