The effect of SSD, Field size, Energy and Detector type for Relative Output Factor measurement in small photon beams as compared with Monte Carlo simulation

被引:3
|
作者
Setilo, Itumeleng [1 ]
Oderinde, Oluwaseyi Michael [1 ]
du Plessis, Freek C. P. [1 ]
机构
[1] Univ Free State, Dept Med Phys, POB 339, ZA-9300 Bloemfontein, South Africa
来源
基金
英国医学研究理事会;
关键词
small-field dosimetry; relative output factor; electronic equilibrium; Monte Carlo; RADIATION-THERAPY; FILM DOSIMETRY; CONVOLUTION;
D O I
10.2478/pjmpe-2019-0014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Small fields photon dosimetry is associated with many problems. Using the right detector for measurement plays a fundamental role. This study investigated the measurement of relative output for small photon fields with different detectors. It was investigated for three-photon beam energies at SSDs of 90, 95, 100 and 110 cm. As a benchmark, the Monte Carlo simulation was done to calculate the relative output of these small photon beams for the dose in water. Materials and Methods: 6, 10 and 15 MV beams were delivered from a Synergy LINAC equipped with an Agility 160 multileaf collimator (MLC). A CC01 ion chamber, EFD-3G diode, PTW60019 microdiamond, EBT2 radiochromic film, and EDR2 radiographic film were used to measure the relative output of the linac. Measurements were taken in water for the CC01 ion chamber, EFD-3G diode, and the PTW60019. Films were measured in water equivalent RW3 phantom slabs. Measurements were made for 1 x 1, 2 x 2, 3 x 3, 4 x 4, 5 x 5 and a reference field of 10 x 10 cm(2). Field sizes were defined at 100cm SSD. Relative output factors were also compared with Monte Carlo (MC) simulation of the LINAC and a water phantom model. The influence of voxel size was also investigated for relative output measurement. Results and Discussion: The relative output factor (ROF) increased with energy for all fields large enough to have lateral electronic equilibrium (LEE). This relation broke down as the field sizes decreased due to the onset of lateral electronic disequilibrium (LED). The high-density detector, PTW60019 gave the highest ROF for the different energies, with the less dense CC01 giving the lowest ROFs. Conclusion: These are results compared to MC simulation, higher density detectors give higher ROF values. Relative to water, the ROF measured with the air-chamber remained virtually unchanged. The ROFs, as measured in this study showed little variation due to increased SSDs. The effect of voxel size for the Monte Carlo calculations in water does not lead to significant ROF variation over the small fields studied.
引用
收藏
页码:101 / 110
页数:10
相关论文
共 28 条
  • [1] Field Size Dependence of Output Factor for Uniform Scanning Proton Beams: A Comparison of TPS Calculation, Measurement and Monte Carlo Simulation
    Zheng, Y.
    Singh, H.
    Islam, M.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 362 - 362
  • [2] The Detector Characteristics for Output Factor Measurement of Small Field Electron Beams
    Wulandari, C.
    Wibowo, W. E.
    Pawiro, S. A.
    [J]. INTERNATIONAL SYMPOSIUM ON CURRENT PROGRESS IN MATHEMATICS AND SCIENCES 2016 (ISCPMS 2016), 2017, 1862
  • [3] ON THE POTENTIAL USE OF AN ALANINE-BASED PROTOCOL FOR SMALL FIELD OUTPUT FACTOR DETERMINATION IN HIGH ENERGY PHOTON BEAMS - A MONTE CARLO STUDY
    Ottosson, R.
    Behrens, C. F.
    Helt-Hansen, J.
    Andersen, C. E.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S24 - S24
  • [4] Comparison of Output Factor measurements with Monte Carlo Simulation for Small Field Dosimetry
    Mateus, D.
    Greco, C.
    Peralta, L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S698 - S698
  • [5] Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams
    Abdullah, R.
    Sapeei, M. S.
    Shukor, N. S. A.
    Yusof, M. F. M.
    Mohamed, M.
    Idris, N. R. N.
    Yusoff, A. L.
    Zakaria, A.
    Aziz, M. Z. A.
    [J]. INTERNATIONAL NUCLEAR SCIENCE, TECHNOLOGY AND ENGINEERING CONFERENCE 2017 (INUSTEC2017), 2018, 298
  • [6] Determination of output factor for 6 MV small photon beam: comparison between Monte Carlo simulation technique and microDiamond detector
    Krongkietlearts, K.
    Tangboonduangjit, P.
    Paisangittisakul, N.
    [J]. 13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [7] Output factor determination based on Monte Carlo simulation for small cone field in 10-MV photon beam
    Fukata K.
    Sugimoto S.
    Kurokawa C.
    Saito A.
    Inoue T.
    Sasai K.
    [J]. Radiological Physics and Technology, 2018, 11 (2) : 192 - 201
  • [8] Detector comparison for small field output factor measurements in flattening filter free photon beams
    Lechner, Wolfgang
    Palmans, Hugo
    Soelkner, Lukas
    Grochowska, Paulina
    Georg, Dietmar
    [J]. RADIOTHERAPY AND ONCOLOGY, 2013, 109 (03) : 356 - 360
  • [9] Energy response of an aluminium oxide detector in kilovoltage and megavoltage photon beams: An EGSnrc Monte Carlo simulation study
    Agyingi, EO
    Mobit, PN
    Sandison, GA
    [J]. RADIATION PROTECTION DOSIMETRY, 2006, 118 (01) : 28 - 31
  • [10] Effect of Detector Orientation and Influence of Jaw Position in the Determination of Small-field Output Factor with Various Detectors for High-energy Photon Beams
    Godson, Henry Finlay
    Manickam, Ravikumar
    Ponmalar, Y. Retna
    Ganesh, K. M.
    Saminathan, Sathiyan
    Chandraraj, Varatharaj
    Kumar, A. Sathish
    George, Seby
    Raman, Arun
    Singh, Rabi Raja
    [J]. JOURNAL OF MEDICAL PHYSICS, 2024, 49 (01) : 73 - 83