Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams

被引:0
|
作者
Abdullah, R. [1 ,4 ]
Sapeei, M. S. [1 ]
Shukor, N. S. A. [1 ]
Yusof, M. F. M. [1 ]
Mohamed, M. [2 ,3 ]
Idris, N. R. N. [2 ,3 ]
Yusoff, A. L. [2 ,3 ]
Zakaria, A. [1 ]
Aziz, M. Z. A. [4 ]
机构
[1] Univ Sains Malaysia, Sch Hlth Sci, Hlth Campus USM, Kelantan, Malaysia
[2] Univ Sains Malaysia, Dept Nucl Med Radiotherapy & Oncol, Sch Med Sci, Hlth Campus USM, Kelantan, Malaysia
[3] Hosp Univ Sains Malaysia, Dept Nucl Med Radiotherapy & Oncol, Kelantan, Malaysia
[4] Univ Sains Malaysia, Adv Med & Dent Inst, Bertam, Pulau Pinang, Malaysia
关键词
RADIOCHROMIC FILM;
D O I
10.1088/1757-899X/298/1/012022
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Accurate dosimetry of small-field photon beams has always been difficult to achieve, due to the steep dose gradient and absence of lateral electronic equilibrium. The purpose of this study was to verify the measurement of relative output factor (ROF), which is one of the dosimetric parameter required for stereotactic radiosurgery (SRS) treatment planning. The ROFs for Radionics circular cone collimators with diameter in the range of 10.0 to 45.0 mm were measured using Gafchromic EBT2 and EBT3 films. The measurements were then compared with the ROFs obtained using a PinPoint ionisation chamber and Monte Carlo (MC) simulation. From the results, the ROFs measured by the ionisation chamber, EBT2, and EBT3 films were good agreement with the MC simulation, with deviations of less than 1.5, 2.6, and 5.0 % respectively. Based on the film dosimetry, the EBT2 film showed in a more reliable measurement for field size ranging from 15.0 to 45.0 mm, compared with EBT3. As a conclusion, based on the special characteristic of the small-field photon beams, ROF measurement using PinPoint ionisation chamber are being favoured, due to its accuracy. However, the EBT2 film can be used as an alternative, when high spatial resolution is required.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Impact of Flattening Filter On Output Correction Factor of MicroDiamond Detector in Small-Field Stereotactic Radiosurgery Beams
    Arai, Y.
    Sato, Y.
    Takizawa, T.
    Fujii, K.
    Yasui, K.
    Sakai, K.
    Sato, M.
    Nakajima, Y.
    Fujita, Y.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E948 - E948
  • [2] Evaluation of detectors response for small field output factor measurement using Gafchromic film
    Rucka, G.
    Patrice, B.
    Asquier, N.
    Mouttet, J. C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S686 - S687
  • [3] Characterization of small-field stereotactic radiosurgery beams with modern detectors
    Tyler, Madelaine
    Liu, Paul Z. Y.
    Chan, Kin Wa
    Ralston, Anna
    McKenzie, David R.
    Downes, Simon
    Suchowerska, Natalka
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (21): : 7595 - 7608
  • [4] Determination of the small-field output factor for 6 MV photon beam using EGSnrc Monte Carlo
    Chuah, K. W.
    Aziz, Abdul M. Z.
    Jayamani, J.
    [J]. JOURNAL OF MEDICAL PHYSICS, 2022, 47 (03) : 301 - 308
  • [5] Comparison of small-field output factor measurements
    Oliver, C.
    Takau, V.
    Butler, D.
    Williams, I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S765 - S765
  • [7] Measurement of output factors for small photon beams
    Sauer, Otto A.
    Wilbert, Juergen
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 1983 - 1988
  • [8] Verification of output factors for small photon beams using Monte Carlo methods
    Bezrukiy, NV
    DeMarco, JJ
    Chetty, I
    Smathers, JB
    Solberg, TD
    [J]. PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 1831 - 1833
  • [9] The effect of SSD, Field size, Energy and Detector type for Relative Output Factor measurement in small photon beams as compared with Monte Carlo simulation
    Setilo, Itumeleng
    Oderinde, Oluwaseyi Michael
    du Plessis, Freek C. P.
    [J]. POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2019, 25 (02): : 101 - 110
  • [10] Small-field radiotherapy photon beam output evaluation: Detectors reviewed
    Lam, S. E.
    Bradley, D. A.
    Khandaker, M. U.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2021, 178