Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams

被引:0
|
作者
Abdullah, R. [1 ,4 ]
Sapeei, M. S. [1 ]
Shukor, N. S. A. [1 ]
Yusof, M. F. M. [1 ]
Mohamed, M. [2 ,3 ]
Idris, N. R. N. [2 ,3 ]
Yusoff, A. L. [2 ,3 ]
Zakaria, A. [1 ]
Aziz, M. Z. A. [4 ]
机构
[1] Univ Sains Malaysia, Sch Hlth Sci, Hlth Campus USM, Kelantan, Malaysia
[2] Univ Sains Malaysia, Dept Nucl Med Radiotherapy & Oncol, Sch Med Sci, Hlth Campus USM, Kelantan, Malaysia
[3] Hosp Univ Sains Malaysia, Dept Nucl Med Radiotherapy & Oncol, Kelantan, Malaysia
[4] Univ Sains Malaysia, Adv Med & Dent Inst, Bertam, Pulau Pinang, Malaysia
关键词
RADIOCHROMIC FILM;
D O I
10.1088/1757-899X/298/1/012022
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Accurate dosimetry of small-field photon beams has always been difficult to achieve, due to the steep dose gradient and absence of lateral electronic equilibrium. The purpose of this study was to verify the measurement of relative output factor (ROF), which is one of the dosimetric parameter required for stereotactic radiosurgery (SRS) treatment planning. The ROFs for Radionics circular cone collimators with diameter in the range of 10.0 to 45.0 mm were measured using Gafchromic EBT2 and EBT3 films. The measurements were then compared with the ROFs obtained using a PinPoint ionisation chamber and Monte Carlo (MC) simulation. From the results, the ROFs measured by the ionisation chamber, EBT2, and EBT3 films were good agreement with the MC simulation, with deviations of less than 1.5, 2.6, and 5.0 % respectively. Based on the film dosimetry, the EBT2 film showed in a more reliable measurement for field size ranging from 15.0 to 45.0 mm, compared with EBT3. As a conclusion, based on the special characteristic of the small-field photon beams, ROF measurement using PinPoint ionisation chamber are being favoured, due to its accuracy. However, the EBT2 film can be used as an alternative, when high spatial resolution is required.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Small field output factors measurement with EBT3 gafchromic film in water
    Obesso, A.
    Ferrer, C.
    Huertas, C.
    Corredoira, E.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S698 - S699
  • [22] The influence of detector size relative to field size in small-field photon-beam dosimetry using synthetic diamond crystals as sensors
    Ade, N.
    Nam, T. L.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2015, 113 : 6 - 13
  • [23] Extending the Eclipse™ AcurosXB output factor table for small field radiosurgery
    Wu, Ning Genevieve
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (01):
  • [24] Eclipse Output Factor Table: Should It Be Extended for Small Field Radiosurgery?
    Wu, G.
    Roberts, T.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E761 - E761
  • [25] Volume Effect Correction Factor KV for Small-Field Photon Dosimetry with Ionization Chambers
    Looe, H. K.
    Stelljes, T. S.
    Harder, D.
    Poppe, B.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3729 - 3729
  • [26] An Experimental Slope Method for a More Accurate Measurement of Relative Radiation Doses using Radiographic and Radiochromic Films and Its Application to Megavoltage Small-Field Dosimetry
    Holla, Raghavendra
    Khanna, D.
    Pillai, Bhaskaran K.
    Ali, K. V. Jafar
    Mon, P. S. Renil
    Clinto, C. O.
    Ganesh, Tharmarnadar
    [J]. JOURNAL OF MEDICAL PHYSICS, 2019, 44 (03) : 145 - 155
  • [27] A dosimetric study of photon small fields using polymer gel and Gafchromic EBT films
    Hassani, H.
    Nedaie, H. A.
    Zahmatkesh, M. H.
    Banaee, N.
    Shirani, K.
    Mirzaee, S.
    Samei, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S174 - S175
  • [28] A dosimetric study of small photon fields using polymer gel and Gafchromic EBT films
    Hassani, Hossein
    Nedaie, Hassan Ali
    Zahmatkesh, Mohammad Hassan
    Shirani, Kaveh
    [J]. MEDICAL DOSIMETRY, 2014, 39 (01) : 102 - 107
  • [29] Small field output factor measurement using various active detectors and OSLD
    Shamurailatpam, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S11 - S12
  • [30] Deconvolution of detector size effect for output factor measurement for narrow Gamma Knife radiosurgery beams
    Bednarz, G
    Huq, MS
    Rosenow, UF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (20): : 3643 - 3649