Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams

被引:0
|
作者
Abdullah, R. [1 ,4 ]
Sapeei, M. S. [1 ]
Shukor, N. S. A. [1 ]
Yusof, M. F. M. [1 ]
Mohamed, M. [2 ,3 ]
Idris, N. R. N. [2 ,3 ]
Yusoff, A. L. [2 ,3 ]
Zakaria, A. [1 ]
Aziz, M. Z. A. [4 ]
机构
[1] Univ Sains Malaysia, Sch Hlth Sci, Hlth Campus USM, Kelantan, Malaysia
[2] Univ Sains Malaysia, Dept Nucl Med Radiotherapy & Oncol, Sch Med Sci, Hlth Campus USM, Kelantan, Malaysia
[3] Hosp Univ Sains Malaysia, Dept Nucl Med Radiotherapy & Oncol, Kelantan, Malaysia
[4] Univ Sains Malaysia, Adv Med & Dent Inst, Bertam, Pulau Pinang, Malaysia
关键词
RADIOCHROMIC FILM;
D O I
10.1088/1757-899X/298/1/012022
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Accurate dosimetry of small-field photon beams has always been difficult to achieve, due to the steep dose gradient and absence of lateral electronic equilibrium. The purpose of this study was to verify the measurement of relative output factor (ROF), which is one of the dosimetric parameter required for stereotactic radiosurgery (SRS) treatment planning. The ROFs for Radionics circular cone collimators with diameter in the range of 10.0 to 45.0 mm were measured using Gafchromic EBT2 and EBT3 films. The measurements were then compared with the ROFs obtained using a PinPoint ionisation chamber and Monte Carlo (MC) simulation. From the results, the ROFs measured by the ionisation chamber, EBT2, and EBT3 films were good agreement with the MC simulation, with deviations of less than 1.5, 2.6, and 5.0 % respectively. Based on the film dosimetry, the EBT2 film showed in a more reliable measurement for field size ranging from 15.0 to 45.0 mm, compared with EBT3. As a conclusion, based on the special characteristic of the small-field photon beams, ROF measurement using PinPoint ionisation chamber are being favoured, due to its accuracy. However, the EBT2 film can be used as an alternative, when high spatial resolution is required.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Three dimensional film dosimetry of photon beam in small field sizes and beyond the heterogeneous regions using a GAFchromic films array
    Arab-Bafrani, Zahra
    Mahani, Leili
    Khoshbin-Khoshnazar, Alireza
    Kermani, Milad Zeinali
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2020, 166
  • [32] Feasibility of operating a millimeter-scale graphite calorimeter for absolute dosimetry of small-field photon beams in the clinic
    Cote, Benjamin
    Keszti, Federico
    Bancheri, Julien
    Sarfehnia, Arman
    Seuntjens, Jan
    Renaud, James
    [J]. MEDICAL PHYSICS, 2021, 48 (11) : 7476 - 7492
  • [33] Monte Carlo calculated output correction factors for Gafchromic EBT3 film for relative dosimetry in small stereotactic radiosurgery fields
    Johnny E. Morales
    Martin Butson
    Robin Hill
    Scott B. Crowe
    J. V. Trapp
    [J]. Physical and Engineering Sciences in Medicine, 2020, 43 : 609 - 616
  • [34] Monte Carlo calculated output correction factors for Gafchromic EBT3 film for relative dosimetry in small stereotactic radiosurgery fields
    Morales, Johnny E.
    Butson, Martin
    Hill, Robin
    Crowe, Scott B.
    Trapp, J., V
    [J]. PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (02) : 609 - 616
  • [35] Study of the response of plastic scintillation detectors in small-field 6 MV photon beams by Monte Carlo simulations
    Wang, Lilie L. W.
    Beddar, Sam
    [J]. MEDICAL PHYSICS, 2011, 38 (03) : 1596 - 1599
  • [36] Small-Field Dosimetry in Single-Isocenter Stereotactic Irradiation for Brain Metastases Using a Detector-Specific Field Output Correction Factor
    Ono, T.
    Nakamura, M.
    Uto, M.
    Mizowaki, T.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [37] Small Field Output Factor Measurement Using A-Si1200 DMI Panel
    Barbee, D.
    Das, I.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2956 - 2956
  • [38] Development of Simplified Measurement Technique for Small Field Output Factor
    Kim, M.
    Park, S.
    Ahn, S.
    Kim, J.
    Park, K.
    Lee, H.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E323 - E323
  • [39] An experimental extrapolation technique using the Gafchromic EBT3 film for relative output factor measurements in small x-ray fields
    Morales, Johnny E.
    Butson, Martin
    Crowe, Scott B.
    Hill, Robin
    Trapp, J. V.
    [J]. MEDICAL PHYSICS, 2016, 43 (08) : 4687 - 4692
  • [40] Deconvolution of detector size effect for output factor measurement for narrow Gamma Knife radiosurgery beams.
    Bednarz, G
    [J]. PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 1705 - 1709