Feasibility of operating a millimeter-scale graphite calorimeter for absolute dosimetry of small-field photon beams in the clinic

被引:2
|
作者
Cote, Benjamin [1 ]
Keszti, Federico [1 ]
Bancheri, Julien [1 ]
Sarfehnia, Arman [1 ,2 ]
Seuntjens, Jan [1 ]
Renaud, James [1 ,3 ]
机构
[1] McGill Univ, Gerald Bronfman Dept Oncol, Med Phys Unit, Montreal, PQ, Canada
[2] Univ Toronto, Dept Radiat Oncol, Toronto, ON, Canada
[3] Natl Res Council Canada, Metrol Res Ctr, Ottawa, ON, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
calorimetry; heat transfer modeling; small-field dosimetry; HIGH-ENERGY PHOTON; MICRODIAMOND; DETECTOR;
D O I
10.1002/mp.15244
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To characterize and build a cylindrically layered graphite calorimeter the size of a thimble ionization chamber for absolute dosimetry of small fields. This detector has been designed in a familiar probe format to facilitate integration into the clinical workflow. The feasibility of operating this absorbed dose calorimeter in quasi-adiabatic mode is assessed for high-energy accelerator-based photon beams. Methods This detector, herein referred to as Aerrow MK7, is a miniaturized version of a previously validated aerogel-insulated graphite calorimeter known as Aerrow. The new model was designed and developed using numerical methods. Medium conversion factors from graphite to water, small-field output correction factors, and layer perturbation factors for this dosimeter were calculated using the EGSnrc Monte Carlo code system. A range of commercially available aerogel densities were studied for the insulating layers, and an optimal density was selected by minimizing the small-field output correction factors. Heat exchange within the detector was simulated using a five-body compartmental heat transfer model. In quasi-adiabatic mode, the sensitive volume (a 3 mm diameter cylindrical graphite core) experiences a temperature rise during irradiation on the order of 1.3 mK center dot Gy(-1). The absorbed dose is obtained by calculating the product of this temperature rise with the specific heat capacity of the graphite. The detector was irradiated with 6 MV (%dd(10)x = 63.5%) and 10 MV (%dd(10)x = 71.1%) flattening filter-free (FFF) photon beams for two field sizes, characterized by Sclin dimensions of 2.16 and 11.0 cm. The dose readings were compared against a calibrated Exradin A1SL ionization chamber. All dose values are reported at dmax in water. Results The field output correction factors for this dosimeter design were computed for field sizes ranging from Sclin = 0.54 to 11.0 cm. For all aerogel densities studied, these correction factors did not exceed 1.5%. The relative dose difference between the two dosimeters ranged between 0.3% and 0.7% for all beams and field sizes. The smallest field size experimentally investigated, Sclin = 2.16 cm, which was irradiated with the 10 MV FFF beam, produced readings of 84.4 cGy (+/- 1.3%) in the calorimeter and 84.5 cGy (+/- 1.3%) in the ionization chamber. Conclusion The median relative difference in absorbed dose values between a calibrated A1SL ionization chamber and the proposed novel graphite calorimeter was 0.6%. This preliminary experimental validation demonstrates that Aerrow MK7 is capable of accurate and reproducible absorbed dose measurements in quasi-adiabatic mode.
引用
收藏
页码:7476 / 7492
页数:17
相关论文
共 4 条
  • [1] Aerrow: A probe-format graphite calorimeter for absolute dosimetry of high-energy photon beams in the clinical environment
    Renaud, James
    Sarfehnia, Arman
    Bancheri, Julien
    Seuntjens, Jan
    [J]. MEDICAL PHYSICS, 2018, 45 (01) : 414 - 428
  • [2] Development of a Quasi-Adiabatic Miniature Graphite Calorimeter for Absolute Clinical Dosimetry of High-Energy Photon Beams
    Keszti, F.
    Cote, B.
    Bancheri, J.
    Sarfehnia, A.
    Seuntjens, J.
    Renaud, J.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E287 - E287
  • [3] Small-field Dosimetry of 6-and 10-MV Flattening Filter-free and Flattening Filter Photon Beams for Therapeutic Use
    Shukla, Jooli
    Dwivedi, Shekhar
    Kansal, Sandeep
    Dangwal, Vinod Kumar
    Tiwari, K. P.
    Singla, Amit Kumar
    Rani, Supriya
    [J]. INDIAN JOURNAL OF PURE & APPLIED PHYSICS, 2023, 61 (10) : 846 - 850
  • [4] Feasibility study of using Stereotactic Field Diode for field output factors measurement and evaluating three new detectors for small field relative dosimetry of 6 and 10 MV photon beams
    Gul, Attia
    Fukuda, Shigekazu
    Mizuno, Hideyuki
    Taku, Nakaji
    Kakakhel, M. Basim
    Mirza, Sikander M.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (11): : 23 - 36