Properties of IBA Razor Nano Chamber in small-field radiation therapy using 6MV FF, 6 MV FFF, and 10 MV FFF photon beams

被引:5
|
作者
Partanen, Mari [1 ,2 ]
Niemela, Jarkko [3 ,4 ,5 ]
Ojala, Jarkko [1 ,2 ]
Keyrilainen, Jani [3 ,4 ]
Kapanen, Mika [1 ,2 ]
机构
[1] Tampere Univ Hosp, Dept Oncol, Unit Radiotherapy, POB 2000, FI-33521 Tampere, Finland
[2] Tampere Univ Hosp, Med Imaging Ctr, Dept Med Phys, Tampere, Finland
[3] Turku Univ Hosp, Dept Med Phys, Turku, Finland
[4] Turku Univ Hosp, Dept Oncol & Radiotherapy, Turku, Finland
[5] Univ Turku, Dept Phys & Astron, Turku, Finland
关键词
Small-field dosimetry; relative dosimetry; ionization chamber; Monte Carlo simulations; EGSnrc; MONTE-CARLO; QUALITY;
D O I
10.1080/0284186X.2021.1979644
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Small megavoltage photon fields are increasingly used in modern radiotherapy techniques such as stereotactic radiotherapy. Therefore, it is important to study the reliability of dosimetry in the small-field conditions. The IBA Razor Nano Chamber (Nano chamber) ionization chamber is particularly intended for small-field measurements. In this work, properties of the Nano chamber were studied with both measurements and Monte Carlo (MC) simulations. Material and methods: The measurements and MC simulations were performed with 6MV, 6MV FFF and 10MV FFF photon beams from the Varian TrueBeam linear accelerator. The source-to-surface distance was fixed at 100cm. The measurements and MC simulations included profiles, percentage depth doses (PDD), and output factors (OF) in square jaw-collimated fields. The MC simulations were performed with the EGSnrc software system in a large water phantom. Results: The measured profiles and PDDs obtained with the Nano chamber were compared against IBA Razor Diode, PTW microDiamond and the PTW Semiflex ionization chamber. These results indicate that the Nano chamber is a high-resolution detector and thus suitable for small field profile measurements down to field sizes 2 x 2 cm(2) and appropriate for the PDD measurements. The field output correction factors K-Qclin,K- (f)(c)(l)(i)(n)(,)(fmsr)(Qmsr )( ) and field OFs Omega(Qclin,) (f)(c)(l)(i)(n)(,)(fmsr)(Qmsr )( ) were determined according to TRS-483 protocol In the 6MV FF and FFF beams, the determined correction factors K-Qclin,K- (f)(c)(l)(i)(n)(,)(fmsr)(Qmsr )( ) were within 1.2% for the field sizes of 1 x 1 cm(2)-3 x 3 cm(2) and the experimental and MC defined field output factors Omega(Qclin,) (f)(c)(l)(i)(n)(,)(fmsr)(Qmsr )( ) showed good agreement. Conclusion: The Nano chamber with its small cavity volume is a potential detector for the small-field dosimetry. In this study, the properties of this detector were characterized with measurements and MC simulations. The determined correction factors K-Qclin,K- (f)(c)(l)(i)(n)(,)(fmsr)(Qmsr )( ) are novel results for the NC in the TrueBeam fields.
引用
收藏
页码:1419 / 1424
页数:6
相关论文
共 50 条
  • [1] Surface Dose and Build-Up Dose Comparison of 6MV and 10MV Flattened with 6MV FFF and 10MV FFF Photon Beams with Different Detectors
    Nambiraj, Arunai
    Sigamani, Ashokkumar
    Kaviarasu, K.
    Rajesh, T.
    Karthikeyan, S.
    Subramanian, Narayanan
    Meerza, Kazem
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3502 - 3502
  • [2] Dosimetric Comparison of SBRT VMAT Treatment Plans Generated for 6 MV, 6 MV FFF, and 10 MV FFF Photon Beams
    Wilson, D.
    Wang, B.
    Dunlap, N.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3404 - 3404
  • [3] Polarity and ion recombination corrections for SemiFlex 3D 31021 IC in 6MV and 6MV FFF photon beams
    Castellet Garcia, M. C.
    Monasor Denia, P.
    Manjon Garcia, C.
    Lopez Tarjuelo, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S921 - S921
  • [4] Comparison of detector performance in small 6 MV and 6 MV FFF beams using a Versa HD accelerator
    Denia, Paula Monasor
    Castellet Garcia, Maria del Carmen
    Manjon Garcia, Carla
    Quiros Higueras, Juan David
    Blancas, Noelia de Marco
    Alandi, Jorge Bonaque
    Juan Senabre, Xavier Jordi
    Serra, Agustin Santos
    Lopez-Tarjuelo, Juan
    [J]. PLOS ONE, 2019, 14 (03):
  • [5] Treatment planning for a 6MV FFF Halcyon linac as a replacement for a 10MV-capable Clinac
    Withers, D.
    Kidane, G.
    Farmakidis, D.
    Crees, L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1073 - S1073
  • [6] Small field output factor measurements: a detector comparison for Versa HD 6 MV and 6 MV FFF beams.
    Monasor Denia, P.
    Castellet Garcia, M. C.
    Manjon Garcia, C.
    Lopez Tarjuelo, J.
    Juan Senabre, X. J.
    Quiros Higueras, J. D.
    De Marco Blancas, N.
    Bonaque Alandi, J.
    Montenegro Iglesias, N.
    Santos Serra, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S920 - S921
  • [7] Feasibility Study of Respiratory Gated RapidArc SBRT Using a 6MV FFF Photon Beam
    Dou, K.
    Jacobs, M.
    Laser, B.
    Safaraz, M.
    Rodgers, J.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3609 - 3610
  • [8] The impact of interplay effect in SBRT lung treatments for 6MV and 6MV-FFF beams using EBT3 film.
    Elezaj, D.
    Ottosson, W.
    Sjolin, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S829 - S829
  • [9] Applications of 6MV FFF Photon Beams in Optimizing Radiobiological Response for Respiratory-Gated Liver SBRT
    Karan, T.
    Moiseenko, V.
    Gill, B.
    Horwood, R.
    Minchinton, A.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3702 - 3702
  • [10] Dosimetrical properties of the new IBA RAZOR NanoChamber ionization chamber in MV photon beams
    Casar, B.
    Gershkevitsh, E.
    Mendez, I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S919 - S919