Dose monitoring and output correction for the effects of scanning field changes with uniform scanning proton beam

被引:1
|
作者
Zhao, Qingya [1 ,2 ]
Wu, Huanmei [3 ]
Cheng, Chee-Wai [1 ,4 ]
Das, Indra J. [1 ,4 ]
机构
[1] IUHPTC, Bloomington, IN 47408 USA
[2] Purdue Univ, Sch Hlth Sci, W Lafayette, IN 47907 USA
[3] IUPUI, Purdue Sch Engn & Technol, Indianapolis, IN 46202 USA
[4] Indiana Univ, Dept Radiat Oncol, Sch Med, Indianapolis, IN 46202 USA
关键词
uniform scanning proton beam; dose monitor; stem effect; proton output calculation; BRAGG PEAK FIELDS; UNIT CALCULATIONS; THERAPY; DOSIMETRY; RANGE;
D O I
10.1118/1.3609417
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The output of a proton beam is affected by proton energy, Spread-Out Bragg Peak (SOBP) width, aperture size, dose rate, and the point of measurement. In a uniform scanning proton beam (USPB), the scanning field size is adjusted (including the vertical length and the horizontal width) according to the treatment field size with appropriate margins to reduce secondary neutron production. Different scanning field settings result in beam output variations that are investigated in this study. Methods: The measurements are performed with a parallel plate Markus chamber at the center of SOBP under the reference condition with 16 cm range, 10 cm SOBP, and 5 cm air gap. The effect of dose rate on field output is studied by varying proton beam current from 0.5 to 7 nA. The effects of scanning field settings are studied by varying independently the field width and length from 12 x 12 to 30 x 30 cm(2). Results: The results demonstrate that scanning field variations can produce output variation up to 3.80%. In addition, larger output variation is observed with scanning field changes along the stem direction of the patient dose monitor (PDM). By investigating the underlying physics of incomplete charge collection and the stem effects of the PDM, an analytical model is proposed to calculate USPB output with consideration of the scanning field area and the PDM stem length that is irradiated. The average absolute difference between the measured output and calculated output using our new correction model are within 0.13 and 0.08% for the 20 and 30 cm snouts, respectively. Conclusions: This study proposes a correction model for accurate USPB output calculation, which takes account of scanning field settings and the PDM stem effects. This model may be used to extend the existing output calculation model from one snout size to other snout sizes with customized scanning field settings. The study is especially useful for calculating field output for treatment without individualized patient specific measurements. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3609417]
引用
收藏
页码:4655 / 4661
页数:7
相关论文
共 50 条
  • [1] The Effect of Scanning Pattern and Dose Rate on Output Factor for Uniform Scanning Proton Beam
    Zhao, Q.
    Wu, H.
    Pack, D.
    Cheng, C.
    Das, I.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [2] Output Dependence On Dose Rate for Uniform Scanning Proton Beams
    Ramirez, E.
    Zheng, Y.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3706 - 3706
  • [3] Machine Learning Based Output Prediction in Uniform Scanning Proton Beam Therapy
    Grewal, H.
    Twyford, T.
    Chacko, M.
    Ahmad, S.
    Jin, H.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E262 - E263
  • [4] Comparison of Full Intensity Modulated and Single Field Uniform Dose Proton Beam Scanning Lung Treatments
    Dowdell, S.
    Grassberger, C.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [5] MU Calculation for Uniform Scanning Proton Beam
    Zhao, Q.
    Wu, H.
    Wolanski, M.
    Hecksel, D.
    Pack, D.
    Das, I.
    [J]. MEDICAL PHYSICS, 2009, 36 (06) : 2795 - +
  • [6] Commissioning of output factors for uniform scanning proton beams
    Zheng, Yuanshui
    Ramirez, Eric
    Mascia, Anthony
    Ding, Xiaoning
    Okoth, Benny
    Zeidan, Omar
    Hsi, Wen
    Harris, Ben
    Schreuder, Andries N.
    Keole, Sameer
    [J]. MEDICAL PHYSICS, 2011, 38 (04) : 2299 - 2306
  • [7] Effects of beam irregularity on uniform scanning
    Chang Hyeuk Kim
    Sea duk Jang
    Tae-Keun Yang
    [J]. Journal of the Korean Physical Society, 2016, 69 : 1053 - 1056
  • [8] Effects of beam irregularity on uniform scanning
    Kim, Chang Hyeuk
    Jang, Sea Duk
    Yang, Tae-Keun
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2016, 69 (06) : 1053 - 1056
  • [9] A sector-integration method for dose/MU calculation in a uniform scanning proton beam
    Zhao, Qingya
    Wu, Huanmei
    Wolanski, Mark
    Pack, Daniel
    Johnstone, Peter A. S.
    Das, Indra J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (03): : N87 - N95
  • [10] Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking
    Farr, J. B.
    Mascia, A. E.
    Hsi, W. -C.
    Allgower, C. E.
    Jesseph, F.
    Schreuder, A. N.
    Wolanski, M.
    Nichiporov, D. F.
    Anferov, V.
    [J]. MEDICAL PHYSICS, 2008, 35 (11) : 4945 - 4954